- Omega Digital Panel Meter User's Guide

1. DISPLAY: 5 digit LCD 0.48" (12.2 mm) high digits
DP63000A-I: Reflective LCD with full viewing angle
DP63000B-I: Transmissive LCD with selectable red or green LED backlight,
viewing angle optimized. Display color change capability with output state
when using an output module.
2. POWER: Input voltage range is +9 to +28 VDC with short circuit and input
polarity protection. Must use a DP6-MLPS1 or a Class 2 or SELV rated power
supply.
3. INPUT RANGES: Jumper Selectable
D.C. Currents: 200 µA, 2 mA, 20 mA, or 200 mA
4. SIGNAL INPUTS:
5. OVERRANGE RATINGS, PROTECTION & INDICATION:
9 to 28 VDC power circuit is not isolated from the signal circuit.
Input Overrange Indication: “OLOL”.
Input Underrange Indication: “ULUL”.
Display Overrange/Underrange Indication: “
.....”/“-.....
6. DISPLAY RESPONSE TIME: 500 msec min.
7. NORMAL MODE REJECTION: 60 dB 50/60 Hz
8. USER INPUT (USR): Programmable input. Connect terminal to common
(USR COMM) to activate function. Internal 10K pull-up resistor to +9 to
28 VDC.
Threshold Levels: V
IL
= 1.0 V max; V
IH
= 2.4 V min; V
MAX
= 28 VDC
Response Time: 5 msec typ.; 50 msec debounce (activation and release)
9. MEMORY: Nonvolatile E
2
PROM memory retains all programming
parameters and max/min values when power is removed.
10. ENVIRONMENTAL CONDITIONS:
Operating Temperature Range for DP63000A-I: -35 to 75°C
Operating Temperature Range for DP63000B-I depends on display color
and intensity level as per below:
Storage Temperature: -35 to 85°C
Operating and Storage Humidity: 0 to 85% max. relative humidity (non-
condensing)
Vibration According to IEC 68-2-6: 5 to 500 Hz, in X, Y, Z direction for 1.5
hours, 5 g’s.
Shock According to IEC 68-2-27: Operational 30 g, 11 msec in 3 directions
Altitude: Up to 2000 meters
11. CONNECTIONS: Wire clamping screw terminals
Wire Strip Length: 0.3" (7.5 mm)
Wire Gage: 30-14 AWG copper wire
Torque: 5 inch-lbs (0.565 N-m) max.
12. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 requirements for
indoor use. Installation Category I, Pollution Degree 2. High impact plastic
case with clear viewing window. Panel gasket and mounting clip included.
13. CERTIFICATIONS AND COMPLIANCES:
SAFETY
UL Recognized Component, File #E313607, UL61010A-1, CSA 22.2 No. 61010-1
Recognized to U.S. and Canadian requirements under the Component
Recognition Program of Underwriters Laboratories, Inc.
UL Listed, File # E313547, UL508, CSA C22.2 No. 14-M95
LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards
IEC 61010-1, EN 61010-1: Safety requirements for electrical equipment
for measurement, control, and laboratory use, Part 1.
IP65 Enclosure rating (Face only), IEC 529
Type 4X Indoor Enclosure rating (Face only), UL50
ELECTROMAGNETIC COMPATIBILITY
Emissions and Immunity to EN 61326: Electrical Equipment for Measurement,
Control and Laboratory use.
Notes:
1. Criterion A: Normal operation within specified limits.
14. WEIGHT: 3.2 oz (100 g)
4
GENERAL METER SPECIFICATIONS
MODEL
NO.
DISPLAY COLOR
INPUT CURRENT
@ 9 VDC WITH
DP6-RLY0
DP63000A-I --- 10 mA 40 mA
DP63000B-I Red (max intensity) 85 mA 115 mA
DP63000B-I Green (max intensity) 95 mA 125 mA
TEMPERATURE
1 & 2 -35 to 75°C
3 -35 to 70°C
4 -35 to 60°C
5 -35 to 50°C
1 & 2 -35 to 75°C
3 -35 to 65°C
5
4
-35 to 35°C
-35 to 50°C
INTENSITY LEVEL
Class AEN 55011Emissions
Emissions:
30 A/m
3 V/rms
Criterion A
Criterion A
EN 61000-4-8
EN 61000-4-6RF conducted interference
1 kV L-L,
Criterion AEN 61000-4-5Surge
1 kV signal
2 kV power
Criterion AEN 61000-4-4Fast transients (burst)
2 kV L&N-E power
10 V/m
Criterion AEN 61000-4-3Electromagnetic RF fields
8 kV air discharge
4 kV contact discharge
Criterion AEN 61000-4-2Electrostatic discharge
Immunity to Industrial Locations:
Power frequency magnetic fields
Green Display
Red Display
INPUT CURRENT
@ 9 VDC WITHOUT
DP6-RLY0
INPUT
RANGE
ACCURACY
@23 °C, less
than 85% RH
INPUT
IMPEDANCE
MAX
INPUT
SIGNAL
RESOLUTION
TEMP.
COEFFICIENT
200 µA
0.1% of span
1.111 K
15 mA 10 nA 70 ppm / °C
2 mA 0.1% of span 50 mA .1 µA 70 ppm / °C
20 mA 0.1% of span
11
150 mA 1 µA 70 ppm / °C
200 mA 0.1% of span
1
500 mA 10 µA 70 ppm / °C
111
PART NUMBER INFORMATION
DP6-SNK0Dual Sinking Open Collector Output card
DP6-SOFT
DP6-485-CABLERS485 Programming Cable (DB9-RJ11)
DP63000B-I
PC Configuration Software for Windows 98, ME, 2000, XP
*
DC Current Meter with backlight display
DP6-232-CABLE
DP6-RLY0
DP6-MLPS1Micro-Line Power Supply, 85 to 250 VAC
Single Relay Output Card
RS232 Programming Cable (DB9-RJ11)
DP6-COM2RS232 Serial Communications Card
DP6-COM1RS485 Serial Communications Card
DP63000A-IDC Current Meter with reflective display
PART NUMBERDESCRIPTION