User’s Guide Shop on line at ® ® www.omega.com e-mail: info@omega.com iSeries info: www.omega.
® ® OMEGAnet® On-Line Service www.omega.com Internet e-mail info@omega.com Servicing North America: USA: ISO 9001 Certified Canada: One Omega Drive, P.O. Box 4047 Stamford CT 06907-0047 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com 976 Bergar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.
TABLE OF CONTENTS Part 1: Introduction............................................................................................2 1.1 Description .................................................................................2 1.2 Safety Considerations ...............................................................3 1.3 Before You Begin .......................................................................4 Part 2: Setup...................................................................................
LIST OF FIGURES: Figure 2.1 Figure 2.2 Figure 2.3 Figure 2.4 Figure 2.5 Figure 2.6 Figure 2.7 Figure 2.8 Figure 2.9 Figure 2.10 Figure 2.11 Figure 2.12 Figure 2.13 Figure 3.1 Figure 3.2 Figure 3.3 Figure 3.4 Figure 3.5 Figure 3.6 Figure 3.7 Figure 3.8 Figure 3.9 Figure 3.10 Figure 3.11 Figure 3.12 Figure 3.13 Figure 3.14 Mounting ...............................................................................................5 Rear Panel Power and Output Connector Labels..............................
NOTES, WARNINGS and CAUTIONS Information that is especially important to note is identified by following labels: • NOTE • WARNING or CAUTION • IMPORTANT • TIP NOTE: Provides you with information that is important to successfully setup and use the Programmable Digital Meter. CAUTION or WARNING: Tells you about the risk of electrical shock. CAUTION, WARNING or IMPORTANT: Tells you of circumstances or practices that can effect the instrument’s functionality and must refer to accompanying documents.
PART 1 INTRODUCTION 1.1 Description This device can be purchased as monitor (read process value only) or as a controller. • The iLD Big Display controller offers unparalleled flexibility in process measurement. Each unit allows the user to select the input type, from 10 thermocouple types (J, K, T, E, R, S, B, C, N and J DIN), Pt RTDs (100, 500 or 1000 Ω, with either 385 or 392 curve), DC voltage, or DC current.
1.2 Safety Considerations This device is marked with the international caution symbol. It is important to read this manual before installing or commissioning this device as it contains important information relating to Safety and EMC (Electromagnetic Compatibility). This instrument is protected in accordance with Class I of EN 61010 (110/240 AC power connections). Installation of this instrument should be done by qualified personnel.
1.3 Before You Begin Inspecting Your Shipment: Remove the packing slip and verify that you have received everything listed. Inspect the container and equipment for signs of damage as soon as you receive the shipment. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent. The carrier will not honor damage claims unless all shipping material is saved for inspection.
PART 2 SETUP 2.1 Mounting Figure 2.
Mounting iLD Big Display Through Panel: 1. Using the panel cutout diagram shown in your Quick Start Manual, cut an opening in the panel. 2. Remove six (or eight) screws at the back of iLD Big Display to remove back cover. 3. Insert the unit into the opening from the front of the panel so the gasket seals between the bezel and the front of the panel. 4. Pass all wiring through customer drilled holes in back cover and connect wiring to terminal blocks. 5.
2.2 Rear Panel Connections The rear panel connections are shown in Figures 2.2 and 2.3. 1 2 1 2 3 1 3 2 3 4 1 2 3 4 5 5 6 4 5 6 4 5 6 6 Figure 2.2 Rear Panel Power and Output Connections Figure 2.3 Rear Panel Input Connections Table 2.
2.3 Electrical Installation 2.3.1 Power Connections Caution: Do not connect power to your device until you have completed all input and output connections. Failure to do so may result in injury! Connect the main power connections as shown in Figure 2.4. 2 1 3 Figure 2.
2.3.2 Thermocouple The figure below shows the wiring hookup for any thermocouple type. For example, for Type K hookup, connect the yellow wire to the TB7(+) terminal and the red wire to the TB8(-) terminal. When configuring your controller, select Thermocouple and Thermocouple Type in the Input Type menu (see Part 3). Figure 2.6 Thermocouple Wiring Hookup Table 2.
2.3.3 Two/Three/Four-Wire RTD The figures below show the input connections and input connector jumpers (shown in bold lines) required to hookup a 2-, 3- or 4-wire RTD. Figure 2.7 a) RTD-1000 ohm and 500 ohm Wiring Hookup b) RTD-100 ohm Wiring Hookup The two-wire connection is simplest method, but does not compensate for lead-wire temperature change and often requires calibration to cancel lead-wire resistance offset. The three-wire connection works best with RTD leads closely equal in resistance.
2.3.4 Process Current The figure below shows the wiring hookup for Process Current 0 – 20 mA. Figure 2.8 Process Current Wiring Hookup (Internal and External Excitation) When configuring your instrument, select Process Type in the Input Type Menu (see Part 3). 2.3.5 Process Voltage The figure below shows the wiring hookup for Process Voltage 0 – 100 mV, 0 – 1 V, 0 – 10 V. Figure 2.
2.3.6 Wiring Outputs This meter has two, factory installed, outputs. The SPDT Mechanical Relay, SPST Solid State Relay, Pulse and Analog Output Connection are shown below. 1 2 3 4 5 6 1 Figure 2.10 a) Mechanical Relay and SSR Outputs Wiring Hookup 2 3 4 5 6 b) Pulse and Analog Outputs Wiring Hookup This device may have a programmable communication output. The RS-232 and RS-485 Output Connection are shown below. 1 1 Figure 2.
This device may also have an excitation output. Figure 2.12 Excitation Output Excitation is not available if communication option is installed. This device has snubber circuits designed to protect the contacts of the mechanical relays when it switches to inductive loads (i.e. solenoids, relays). These snubbers are internally connected between the Common (C) and Normally Open (NO) relay contacts of Output 1 and Output 2.
PART 3 OPERATION: Configuration Mode 3.1 Introduction The instrument has two different modes of operation. The first, Run Mode, is used to display values for the Process Variable, and to display or clear Peak and Valley values. The other mode, Menu Configuration Mode, is used to navigate through the menu options and configure the controller. Part 3 of this manual will explain the Menu Configuration Mode. For your instrument to operate properly, the user must first "program" or configure the menu options.
3.2 Menu Configuration It is recommended that you put the controller in the Standby Mode for any configuration change other than Setpoints & Alarms. Figure 3.
3.2.1 ID Number SEE ID MENU SELECTION IN CONFIGURATION SECTION FOR ENABLE/DISABLE OR CHANGE ID CODE. If ID Code is Disabled or set as Default (0000) the menu will skip ID step to Setpoint Menu. If ID Code is set to Full Security Level and user attempts to enter the Main Menu, they will be prompted for an ID Code. If ID Code is set to Setpoint/ID Security Level and user attempts to enter the Configuration Menu, they will be prompted for an ID Code. ENTERING YOUR NON-DEFAULT FULL SECURITY ID NUMBER.
3.2.2 Set Points SETPOINT 1: Press a 1) Press a, if necessary until SP1 prompt appears. Press d 2) Display shows previous value of “Setpoint 1”. Press b & c 3) Press b and c to increase or decrease Setpoint 1 respectively. Holding b & c buttons down for approximately 3 seconds will speed up the rate at which the Setpoint value increments or decrements. Press b & c 4) Continue to use b and c to enter your 4-digit Setpoint 1 value.
3.2.3 Configuration Menu Figure 3.2 Flow Chart for Configuration Menu Enter Configuration Menu: Press a Press d Press a 1) Press a, if necessary, until CNFG prompt appear. 2) Display advances to INPT Input Menu. 3) Pressing and releasing a to scroll through all available menus of Configuration section. 3.2.4 Input Type Menu Figure 3.
Input Type (Thermocouple) ENTER INPUT TYPE MENU: Press a Press d Press d 1) Press a, if necessary, until CNFG prompt appears. 2) Display advances to INPT Input Menu. 3) Display flashes T.ç, RTD or PROC (Thermocouple, RTD or Process). If the displayed input type is T.ç, press a to skip to step 6 (T.ç stops flashing). THERMOCOUPLE SUBMENU: Press b Press d Press d Press b Press d 4) Scroll through the available selection to T.ç (flashing). 5) Display shows STRD stored message momentarily and then T.
Input Type (RTD) ENTER INPUT TYPE MENU: Press a Press d Press d 1) Press a, if necessary, until CNFG prompt appears. 2) Display advances to INPT Input Menu. 3) Display flashes T.ç, RTD or PROC (Thermocouple, RTD or Process). If the displayed input type is RTD, press a to skip to step 6 (RTD stops flashing). RTD SUBMENU: Press b Press d Press d Press b Press d 4) Scroll through the available selection to RTD (flashing). 5) Display shows STRD stored message momentarily and then RTD (not flashing).
Input Type (Process) ENTER INPUT TYPE MENU: Press a 1) Press a, if necessary, until CNFG prompt appears. Press d 2) Display advances to INPT Input Menu. Press d 3) Display flashes T.ç, RTD or PROC (Thermocouple, RTD or Process). If the displayed input type is PROC, press a to skip to step 6 (PROC stops flashing). PROCESS SUBMENU: Press b 4) Scroll through the available selection to PROC (flashing). Press d 5) Display shows STRD stored message momentarily and then PROC (not flashing).
ENTER READING CONFIGURATION MENU: Press a Press d Press a Press d 1) Press a, if necessary, until CNFG prompt appears. 2) Display advances to INPT Input Menu. 3) Display advances to RDG Reading Configuration Menu. 4) Display advances to DEC Decimal Point. DECIMAL POINT SUBMENU: Press d Press b Press d 5) Display flashes previous selection for Decimal location. 6) Scroll though the available selections and choose Decimal location: FFFF or FFF.F (also FF.FF and F.
Reading Configuration (If Process was selected) INPUT/READING (SCALE AND OFFSET) SUBMENU: Input Voltage or Current can be converted or scaled into values appropriate for the process or signal being measured. So, a reading may be displayed, for example, in units of weight or velocity instead of in amperes or volts. The instrument determines Scale and Offset values based on two user-provided input values entered with the corresponding readings.
Conversion number is a coefficient of conversion between input values and real full display range (10000 counts, shown as 9999). See Table 3.2 below for proper conversion number. Table 3.2 Conversion Table RANGE 100 mV 1V 10 V 0 -20 mA Example 0-1V In 1 Rd 1 Inp 2 Rd 2 CONVERSION NUMBER 10000 / (100 x 1) = 100 10000 / (1000 x 1) = 10 10000 / (1000 x 10) = 1 10000 / (20 x 1) = 500 = = = = = = 0 - 100.0 0 0 9999 100.
3.2.6 Alarm 1 This unit is equipped with two physical outputs that can only be configured as follows: Alarm 1 & Alarm 2, Alarm 1 & Output 2, Output 1 & Alarm 2, Output 1 & Output 2, Analog Out 1 & Alarm 2, Analog Out 1 & Output 2. Analog Out available only if Analog Output Option board is factory installed. If Analog Output Option is installed, the controller will skip Alarm 1 Menu item to Analog Output. Alarm must be DISABLED if Ramp is ENABLED. Figure 3.
ALARM 1 ENABLE/DISABLE SUBMENU: Press b Press d 5) Scroll though the available selection until ENBL displays to use Alarm 1. 6) Display shows STRD stored message momentarily and then advances to ABSo only if it was changed, otherwise press a to advance to ABSo Alarm 1 Absolute/Deviation Submenu. If DSBL Alarm 1 Disabled was selected, all submenus of Alarm 1 Menu will be skipped and meter advances to ALR2 Alarm 2 Menu.
CONTACT CLOSURE SUBMENU: Press d Press d 11) Display flashes previous selection. Press b to N.ç. Normally Closed or N.o. Normally Open. 12) Display shows STRD stored message momentarily and then advances to AçTV only if it was changed, otherwise press a to advance to AçTV Active Submenu. Normally Open: If this feature is selected, then the relay is "energized" only when an alarm condition occurs. Normally Closed: "Fail Safe" Mode.
ALARM ENABLE/DISABLE AT POWER ON: Press d Press d 15) Display flashes previous selection. Press b to ENBL enable or DSBL disable. 16) Display shows STRD stored message. momentarily and then advances to ALR.L only if it was changed, otherwise press a to advance to the ALR.L Alarm 1 Low Value Submenu. If the alarm is enabled at Power On, the alarm will be active right after reset. If the alarm is disabled at Power On, the alarm will become enabled when the process value enters the non alarm area.
3.2.7 Analog Output (Retransmission) Analog Output can be configured as Retransmission or Control outputs. In this section we will discuss Retransmission Output. This unit is equipped with two physical outputs that can only be configured as follows: Alarm 1 & Alarm 2, Alarm 1 & Output 2, Output 1 & Alarm 2, Output 1 & Output 2, Analog Out 1 & Alarm 2, Analog Out 1 & Output 2. Analog Output is available only, if Analog Output Option board is factory installed.
ANALOG OUTPUT ENABLE/DISABLE SUBMENU: Press b Press d 5) Scroll though the available selection until ENBL displays to use Analog Output Retransmission (output proportional to the input signal). 6) Display shows STRD stored message momentarily and then advances to CURR or VoLT Submenu only if it was changed, otherwise press a to advance to CURR or VoLT Current/Voltage Submenu.
Accuracy of Analog Output board is +/-1% of FS (Full Scale) when following conditions are satisfied: 1. The input is not scaled below 1% of Input FS (10 mV @ 1 V or 0.2 mA @ 20 mA input ranges). 2. Analog Output is not scaled below 3% of Output FS (300 mV @ 10 V or 0.6 mA @ 20 mA output ranges). Otherwise certain corrections need to be applied. For example: For entire range of process input, the Analog Output on 10 V FS scaled for 300 mV output range: Rd1 = 0000, Out1 = 00.00 RD2 = 9999, Out2 = 00.
3.2.8 Alarm 2 This unit is equipped with two physical outputs that can only be configured as follows: Alarm 1 & Alarm 2, Alarm 1 & Output 2, Output 1 & Alarm 2, Output 1 & Output 2, Analog Out 1 & Alarm 2, Analog Out 1 & Output 2. Analog Out available only if Analog Output Option board is factory installed. Alarm must be DISABLED if Ramp is ENABLED. Figure 3.7 Flow Chart for Alarm 2 ENTER ALARM 2 MENU: Press a Press d Press a Press d 1) Press a, if necessary, until CNFG prompt appears.
ALARM 2 ENABLE/DISABLE SUBMENU: Press d Press d 5) Display flashes previous selection. Press b until ENBL displays to use Alarm 2. 6) Display shows STRD stored message momentarily and then advances to ABSo only if it was changed, otherwise press a to advance to ABSo Absolute/Deviation Submenu. If DSBL Alarm 2 Disabled was selected, all submenus of Alarm 2 will be skipped and meter advances to LOOP Loop Break Time Menu.
LOOP BREAK TIME VALUE SUBMENU: Press d 7) Display flashes 1st digit of previous Loop Value. Press b & c 8) Press b and c buttons to enter a new Loop Value (0 to 99.59). Press d 9) Display shows STRD stored message momentarily and then advances to R.ADJ Reading Adjust Submenu.
THERMOCOUPLE FIELD CALIBRATION SUBMENU: CAUTION: Do not perform the following steps until you fully understand this entire section. RTD and Process are perfectly calibrated. This section is applicable to Thermocouple (TC) calibration only. Be sure that the TC being used to calibrate the meter is of the type selected in the TC submenu. Place the TC in an ice-bath (or other 0°C / 32°F environment). In ambient temperature conditions: connect the TC to the meter, apply power to the meter.
3.2.10 Output 1 Alarm 1 and Output 1 or Analog Output (Retransmission) share the same contacts on the rear panel connector. If Alarm 1 or Analog Output (Retransmission) is Enabled, Output 1 is automatically Disabled. It is recommended that you put the controller in the Standby Mode for any configuration change other than Setpoints & Alarms. Figure 3.
ENTER OUTPUT 1 MENU: Press a Press d Press a Press d 1) Press a, if necessary, until CNFG prompt appears. 2) Display advances to INPT Input Menu. 3) Press a, if necessary, until Display advances to OUT1 Output 1 Menu. 4) Display advances to SELF Self Submenu. SELF SUBMENU: The Self Option allows the output of the instrument to be controlled manually from the front panel. Press d Press b Press d 5) Display flashes the current setting of Self, ENBL Enabled or DSBL Disabled.
MAXIMUM/PERCENT HIGH SUBMENU: Specify in percent, the maximum value (99) for control output. If the output is analog proportional (Current or Voltage), then the maximum voltage or current, in percent, is specified. If the output is time proportional (Relay, SSR, or Pulse), then the maximum duty-cycle, in percent, is specified. Press d 12) Display flashes 1st digit of previous “Percent High” setting. Press b & c 13) Use b and c buttons to enter a new value for “Percent High”.
ACTION TYPE SUBMENU: The error that results from the measurement of the Process Variable may be positive or negative since it may be greater or smaller than the Setpoint. If a positive error should cause the instrument output to increase (i.e. cooling), it would be called Direct Acting. If a negative error should cause the output to increase (i.e. heating), it would be called Reverse Acting. Press d Press b Press d 18) Display flashes DRçT Direct or RVRS Reverse.
If Anti Integral (Anti Windup) Submenu “Enabled”, this feature allows the error term outside the proportional band to be calculated and accumulated for integration. This may be an important feature in applications where fast response time is desirable. START AUTO TUNE PID: Press d Press b Press d 27) Display flashes ENBL or DSBL. 28) Scroll through the available selections: “Enable” or “Disable”.
RESET SETUP SUBMENU: Press d 33) Display flashes 1st digit of the previous I REST Reset value. Press b & c 34) Press b and c buttons to enter a new “Reset” value. Press d 35) Display shows STRD stored message momentarily and then advances to RATE only, if it was changed, otherwise press a to advance to RATE Rate Setup Submenu. Reset unit is in seconds 0-3999. RATE SETUP SUBMENU: Press d 36) Display flashes 1st digit of previous D RATE Rate value.
DAMPING FACTOR SUBMENU: Press d Press b Press d 42) Display flashes the previous “Damping Factor” selection. 43) Scroll through the available selections: 0000, 0001, 0002, 0003, 0004, 0005, 0006, 0007. 44) Display flashes STRD stored message and then advances to OUT2 only, if it was changed, otherwise press a to advance to OUT2 Output 2 Menu.
3.2.11 Output 2 Output 2 and Alarm 2 share the same contacts on the rear panel connector. If Alarm 2 is Enabled, Output 2 is automatically Disabled. It is recommended that you put the controller in the Standby Mode for any configuration change other than Setpoints & Alarms. Figure 3.10 Flow Chart for Output 2 ENTER OUTPUT 2 MENU: Press a Press d Press a Press d 1) Press a, if necessary, until CNFG prompt appears. 2) Display advances to INPT Input Menu.
ACTION TYPE SUBMENU: The error that results from the measurement of the Process Variable may be positive or negative since it may be greater or smaller than the Setpoint. If a positive error should cause the instrument output to increase (i.e. cooling), it would be called Direct Acting. If a negative error should cause the output to decrease (i.e. heating), it would be called Reverse Acting. Press d Press b Press d 6) Display flashes DRçT Direct or RVRS Reverse.
CYCLE TIME SUBMENU: Press d 15) Display flashes 1st digit of the previous “Cycle Time” value. Press b & c 16) Press b and c buttons to enter a new “Cycle Time” value (1 to 199 seconds). Press d 17) Display shows STRD stored message momentarily and then advances to RAMP only, if it was changed, otherwise press a to advance to RAMP Ramp Value Submenu. A cycle time selected between 1 to 199 seconds indicates the total On/Off time of each proportional cycle.
It is recommended that you put the controller in the Standby Mode for any configuration change other than Setpoints & Alarms. 3.2.12 Ramp & Soak Alarm must be DISABLED if Ramp is ENABLED. Figure 3.11 Flow Chart for Ramp and Soak ENTER RAMP AND SOAK MENU: Press a Press d Press a 1) Press a, if necessary, until CNFG prompt appears. 2) Display advances to INPT Input Menu. 3) Press a, if necessary, until Display advances to RAMP Ramp and SOAK Soak Menu.
Ramp & Soak provides users with the flexibility to slowly bring the Process Variable (PV) to the desired setpoint. Ramp & Soak values are specified in HH.MM format. The Ramp value indicates the time specified to bring the process variable to Setpoint 1 (SP1). Once the set point is reached, the PID takes over and the Process Variable will be controlled at the desired set point indefinitely.
3.2.13 ID CODE Figure 3.12 Flow Chart for ID Code ENTER ID CODE MENU: Press a Press d Press a 1) Press a, if necessary, until CNFG prompt appears. 2) Display advances to INPT Input Menu. 3) Press a, if necessary, until Display advances to ID ID Code Menu. ENTERING OR CHANGING YOUR (NON-DEFAULT) ID CODE: Press d 4) Display advances to ____ with 1st under score flashing. Press b & c 5) Press b and c to enter your 4-digit “ID Code” number. Press d 6) Display advances to CH.ID Change ID Code Submenu.
ENTERING OR CHANGING YOUR (DEFAULT) ID CODE: Enter ID menu (Repeat steps from 1 to 3). Press d Press d 10) Display advances to CH.ID Change ID Code Submenu. 11) Display shows 0000 message with flashing 1st digit. If you want to change your default “ID Code” you can do it now, otherwise press a and menu will skip to FULL Full Security Submenu. Press b & c 12) Press b and c buttons to enter your new “ID Code” number.
3.2.14 COMMUNICATION OPTION Purchasing the controller with Serial Communications permits an instrument to be configured or monitored from an IBM PC compatible computer using software available from the website or on the CD-ROM enclosed with your shipment. For complete instructions on the use of the Communications Option, refer to the Serial Communications Reference Manual. Figure 3.
ENTER COMMUNICATION OPTION MENU: Press a Press d Press a Press d 1) Press a, if necessary, until CNFG prompt appears. 2) Display advances to INPT Input Menu. 3) Press a, if necessary, until Display advances to COMM Communication Options Menu. 4) Display advances to C.PAR Communication Parameters Submenu. If Communication Option is not installed, the display shows NONE and skips to the Color Display Menu.
STOP BIT SUBMENU: Press d Press b Press d 15) Display flashes previous selection for “Stop Bit”. 16) Scroll through the available selections: 1-BIT, 2-BIT. 17) Display shows STRD stored message momentarily and then advances to BUS.F only, if it was changed, otherwise press a to advance to BUS.F Bus Format Submenu. BUS FORMAT SUBMENU: Determines Communications Standards and Command/Data Formats for transferring information into and out of the controller via the Serial Communications Bus.
Press d Press b Press d 25) Display flashes previous selection for “Echo”. 26) Scroll through the available selections: NO, YES. 27) Display flashes STRD stored message momentarily and then advances to STND only if it was changed, otherwise press a to advance to STND Communication Standard Submenu.
DATA FORMAT SUBMENU: Preformatted data can be sent automatically or upon request from the controller. Use the Data Format Submenus to determine what data will be sent in this preformatted data string. Refer to the iSeries Communications Manual for more information about the data format. At least one of the following suboptions must be enabled and hence output data to the Serial Bus. This menu is applicable for Continuous Mode of RS-232 communication.
TEMPERATURE UNIT SUBMENU: Includes a byte in the data string to indicate whether reading is in Celsius or Fahrenheit. Press d Press b Press d 50) Display flashes previous selection for UNIT. 51) Scroll through the available selections: NO, YES. 52) Display shows STRD stored message momentarily and then advances to ADDR only, if it was changed, otherwise press a to advance to ADDR Address Setup Submenu. ADDRESS SETUP SUBMENU: This menu is applicable to the RS-485 Option only.
3.2.15 DISPLAY COLOR SELECTION This submenu allows the user to select the color of the display. Figure 3.14 Flow Chart for Display Color Selection ENTER DISPLAY COLOR SELECTION MENU: Press a Press d Press a Press d 1) Press a, if necessary, until CNFG prompt appears. 2) Display advances to INPT Input Menu. 3) Press a, if necessary, until Display advances to COLR Display Color Selection Menu. 4) Display advances to N.CLR Normal Color Submenu.
ALARM 2 DISPLAY COLOR SUBMENU: Press d Press b Press d 11) Display flashes previous selection for “Alarm 2 Color Display”. 12) Scroll through the available selections: GRN, RED or AMBR. 13) Display shows STRD stored message momentarily and then momentarily shows the software version number, followed by RST Reset, and then proceeds to the Run Mode. IN ORDER TO DISPLAY ONE COLOR, SET THE SAME DISPLAY COLOR ON ALL THREE SUBMENUS ABOVE.
Example 3: Output 1 = Analog Output (Alarm 1 disabled), Setpoint 1 = 300, Output 2 = Relay, Setpoint 2 = 200 Alarm 1 & 2 Setup: Deviation, Band, “ALR.H” = 10 Color Display Setup: “N.CLR” = Green, “1.CLR” = Amber, “2.CLR” = Red Display Colors change sequences: RED RED RED GREEN RED •➤---------------•------•------•--------------------------------•------•-------•----------------➤ 0 190 200 210 290 300 310 Alarm 1 is designed to monitor the Process Value around the Setpoint 1.
PART 4 SPECIFICATIONS Configuration Single-ended Polarity Unipolar Step Response 0.7 sec for 99.9% Decimal Selection None, 0.1 for temperature None, 0.1, 0.01 or 0.001 for process Setpoint Adjustment -1999 to 9999 counts Span Adjustment 0.001 to 9999 counts Offset Adjustment -1999 to +9999 Accuracy ±0.5°C temp; 0.03% reading process Resolution 1°/0.1°; 10 µV process Temperature Stability 1) RTD: 0.04°C/°C 2) TC @ 25°C (77°F): 0.
CONTROL OUTPUT 1 & 2 Relay 250 Vac or 30 Vdc @ 3 A (Resistive Load); configurable for on/off, PID and Ramp and Soak Output 1: SPDT type, can be configured as Alarm 1 output Output 2: SPDT type, can be configured as Alarm 2 output SSR 20-265 Vac @ 0.05-0.
Table 4.1 Input Properties TC J K Input Type Iron-Constantan CHROMEGA®ALOMEGA® T Copper-Constantan E CHROMEGAConstantan R Pt/13%Rh-Pt S Pt/10%Rh-Pt B 30%Rh-Pt/ 6%Rh-Pt C N 5%Re-W/ 26%Re-W L J DIN Pt, 0.00385, 100 Ω, 500 Ω, 1000 Ω Pt, 0.
PART 5 FACTORY PRESET VALUES Table 5.1 Factory preset value MENU ITEMS FACTORY PRESET VALUES NOTES Set Point 1 (SP1) 000.0 Set Point 2 (SP2) 000.0 Input: Input Type (INPT) TC, type K Reading Configuration (RDG): Decimal Point (DEC.P) FFF.F Temperature unit (TEMP) °F Filter value (FLTR) 0004 Alarm 1 & 2: Alarm 1 (ALR1), Alarm 2 (ALR2) Disable (DSBL) Absolute/Deviation (ABSO/DEV) Absolute (ABSO) Latch/Unlatch (LTCH/UNLT) Unlatch (UNLT) Contact Closure (CT.CL) Normally Open (N.O.
MENU ITEMS Ramp & Soak (RAMP): Ramp (RAMP) Soak (SOAK) Ramp Value (RAMP) Soak Value (SOAK) ID: ID Value Full ID (FULL) Set Point ID (ID.SP) Communication Parameters: Baud Rate (BAUD) Parity (PRTY) Data bit (DATA) Stop Bit Modbus Protocol (M.BUS) Line Feed (LF) Echo (ECHO) Standard Interface (STND) Command Mode (MODE) Separation (SEPR) Alarm Status (STAT) Reading (RDNG) Peak Valley (VALY) Units (UNIT) Multipoint Address (ADDR) Transmit Time (TR.TM) Display Color (COLR): Normal Color (N.CLR) Alarm 1 Color (1.
CE APPROVAL INFORMATION 1. Electromagnetic Compatibility (EMC) This device conforms with requirements of EMC Directive 89/336/EEC, amended by 93/68/EEC. This instrument complies with the following EMC Immunity Standards as tested per EN 61326-2:1998 (Industrial Locations) Phenomena Test Specification Basic Standard Electrostatic Discharge +/- 4 kV contact discharge +/- 8 kV air discharge IEC 1000-4-2 Performance Criteria B Radio Frequency electromagnetic field.
NOTES 65
WARRANTY/DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of one (1) year from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned to OMEGA. If the unit should malfunction, it must be returned to the factory for evaluation.
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA…Of Course! Shop on line at www.omega.