The LDX-3A requires three connections.
1. Transducer
2. Mains power supply.
3. Output signal, voltage or current.
A small hole should be made in the grommet prior to passing the wires through. If a screened cable is to be used, an
EMC cable gland is recommended.
For best performance in electrically noisy environments, the case of the LDX-3A should be connected to a local earth.
This can be achieved via the mounting bracket. This earth connection is not a safety earth, but is part of the overall
electrical screening scheme.
The wiring layout arrangements are similar for LDX-3A fitted with EMC glands and screened cable.
0V +VE
V OUT 0V I OUT
COARSE GAIN
FINE GAIN
COARSE
OFFSET
-VE
+VE
5V
10V
FINE OFFSET
10K
2K
- -
¦
R
¦
|F|
- -
DIV 2
X2
X4
G BK W R B
OUTPUT
POWER
1 2 3 4 5 6 (7)
FREQ
LO
(ON)
HI
(OFF)
NULL
4.3: Electrical Connections
4.0: Installation (continued)
M-1633/0905
LDX-3A