Owner's manual
LDX-3A
2.0: Safety Information
4
M-1633/0905
CAUTION:
Electrostatic Discharge
This equipment is susceptible to ESD (Electrostatic Discharge) when being installed or adjusted, or whenever the case
cover is removed. To prevent ESD related damage, handle the conditioning electronics by its case and do not touch the
connector pins. During installation, follow the guidelines below.
- Ensure all power supplies are turned off.
- If possible, wear an ESD strap connected to ground. If this is not possible, discharge yourself by touching a
metal part of the equipment into which the conditioning electronics is being installed.
- Connect the transducer and power supplies with the power switched off.
- Ensure any tools used are discharged by contacting them against a metal part of the equipment into which the
conditioning electronics is being installed.
- During setting up of the conditioning electronics, make link configuration changes with the power supply turned off.
Avoid touching any other components.
- Make the final gain and offset potentiometer adjustments, with power applied, using an appropriate potentiometer
adjustment tool or a small insulated screwdriver.
CAUTION:
Fuse Rating
The LDX-3A power supply unit does not need a fuse for safe operation. However, wiring to the LDX-3A power supply
unit may need to be fused. If the mains lead supplied with your LDX-3A has a fuse incorporated into the plug, then a
fuse of the same type and value should replace it. Otherwise, fusing and wiring to the LDX-3A must be performed in
accordance with local safety regulations. Refer wiring and fuse replacement to qualified personnel.