Programmable Recorder User's Guide

5-6
1-5V Input
Select 1-5V.
Set the channel range.
Set the left span value.
Set the right span value.
Set the left scale value.
Set the right scale value.
Turn On or Off the low-cut function.
The new setting takes effect.
First channel
(Select the right scale value.)
Last channel
ESC/?
Set=Range
CH=01-01
Mode=1-5V
Span_L= 1,000
Span_R= 5,000
Scale_L= 0.00
Scale_R= 200.00
Low-cut=Off
01-01 Channel
Setting complete
Description
Span Left and Span Right
Selectable span left range: 0.8 to 1.2 V
Selectable span right range: 4.8 to 5.2 V
Scale Left and Scale Right
Selectable range (mantissa): –20000 to 30000
Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
The decimal position is the position specified by scale left.
Make sure that scale left is less than scale right.
Low-cut
On: Cuts values below 0% of the recording span (set to 0% (scale left value)).
Off: Does not apply the low-cut function.
Note
The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.
If the leftmost value of the scale is set to –20000 (excluding the decimal), the value when the
low-cut function is applied is displayed as “–Over.”
5.1 Setting the Input range