Programmable Recorder User's Guide
7-21
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
7
7.13 Enabling the Partial Expanded Recording
Function
This section explains the details of enabling/disabling the partial expanded recording
function.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1. Hold down the 
MENU
 key for 3 seconds to enter Setting mode.
2. Hold down the 
 and   keys simultaneously for 3 seconds to enter Basic
Setting mode.
3. Press the   key or SHIFT +   key to select Partial and then press the 
key.
4. Set each item and press the   key.
Use the   key or SHIFT +   key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5. When the Setting complete screen appears, do either of the following:
To correct the setting, press the 
 key.
If you are done, press the ESC key.
Enable/Disable partial expanded recording.
The new setting takes effect.
ESC/?
Partial
Setting complete
Basic=Partial
Partial=Not
Applying the Changes and Returning to Operation Mode
Press the 
 key or SHIFT +   key to select End and then press the   key. Press
the   key to select Store and then press the   key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the   key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Partial
Use: Enables the setting of the boundary position and boundary value of the partial
expanded recording function in Setting mode.
Not: The Partial item does not appear in Setting mode.
<Related Topics>
Setting the partial expanded recording function: Section 6.5