This section explains the details of registering display types to screens 01 to 15. As an
example, the procedure of setting 1-channel digital display is explained below.
Procedure
1-Channel Digital Display
1. Hold the 
MENU
1
 key for 3 seconds to show the data display setup screen.
2. Press the   key or SHIFT +   key to select the screen number and then
press the 
 key.
A sample screen of the display type appears. A section of the display blinks to
indicate that this is a data display setup screen.
3. Press the 
 or SHIFT +   key to switch the display. Select the 1-channel
digital display and then press the   key.
4. Press the   key or SHIFT +   key to select the channel switching interval
and then press the 
 key.
5. When the Setting complete screen appears, press the   key.
Hold the 
MENU
1
 key down for 3 seconds to exit from the data display setup
screen.
Select the channel switching interval.
Select the display type.
Select the screen No.
Current display type name.
The new setting takes effect.
Screen No.01
Setting complete
Screen No.=01
Interval=Auto2s
1 200.0 mV
Explanation
Interval
Sets the channel switching interval.
Select the interval from 1 s, 2 s, 3 s, 4 s, 5 s, and manual.
1s, 2s, 3s, 4s, 5s: Switches the channels at the specified interval.
Manual: Switch the channel manually.
Assigning Other Display Types (See the menu on the previous page)
• When assigning the flag display, specify whether to display the flag for each channel.
• If the channel switching interval is set to manual when assigning the display in which
different screens can be assigned to the top and bottom sections, set the channel
number.
<Related Topics>
Display types: Section 12.4