Technical Specification
82
Service and calibration frequency
Calibration - At 12 monthly intervals
Country Specific Approvals
FCC
OmiScan Gas Handset OM4720/2:
FCC CFR47, Part 15 Subpart B
This device complies with part 15 of the FCC Rules. Operation is subject to the
following 2 conditions: (1) This device may not cause harmful interference, and (2)
This device must accept any interference received, including interference that may
cause undesired operation.
Grantee and Equipment Product Code:
FCC ID: SV4-OM4720
‘We hereby declare that product OmiScan Gas Handset OM4720/2 is a class 2 BT
device (max 2.7 mW EIRP). Due to the low level of transmit output power the device
is not subject to the routine RF exposure valuation as per Section 2.1093 of the FCC
rules’.
OmiScan Gas Base Station OM4720/1:
FCC CFR47, Part 15 Subpart B
This device complies with part 15 of the FCC Rules. Operation is subject to the
following 2 conditions: (1) This device may not cause harmful interference, and (2)
This device must accept any interference received, including interference that may
cause undesired operation.
Note: This Product contains FCC ID: PI403B
Service
Replace disposable filter element (0.1 µm) –
Part No. OM 4700/11/10 (Pack of 10)
Weekly – if in continuous use,
else as necessary
Single disposable coalescing filter (0.1 µm) –
Part No. FL5720A
Weekly – if in continuous use,
else as necessary
In Line hydrophobic pancake filter –
Part No. FL5718A
As necessary
Cleaning of sample probe and hose Weekly – if in continuous use,
else as necessary
Replace oxygen cell - OM4000/12 Two years, or as needed