Instruction Manual Automatic Digital Blood Pressure Monitor Model HEM-703CP HEM-703CP-xxxx 0692581-7 A
Important 3. This unit is a precision instrument − avoid rough handling, direct sunlight, high temperature/humidity and magnetic field. Fold-ing or bending cuff may result in damage to cuff. 4. Do not inflate cuff above 280 mmHg. 5. Do not take unit apart. You cannot calibrate unit. If you have any difficulties or problems with unit, contact our Service Center. 1-800-634-4350. Thank you for purchasing our Digital Blood Pressure & Pulse Monitoring Kit.
Name and Function of Each Part Main Unit Systolic Bood Pressure Display Diastolic Blood Pressure and Pulse Display Printing Paper Air Jack Paper Cutter MODE Selector Button SPHYG/CLOCK Button SET Button START Button PRINT / PAPER FEED Button Cuff Tab Air Plug 2 Air Tube
Preparations for Use 1 How to insert the batteries Replacement 1. Take off the battery cover at the back of the main unit by sliding it in the direction indicated by the arrow symbol. • When the “Replace Batteries” symbol ( ) flashes during measurement, please replace the batteries with the new ones. • When the batteries have run down, random numbers may be displayed.
Preparations for Use 2 How to Set Time and Date 3 Plug in and Turn on 1. When batteries are installed, the display shows PM 12:00. 1. Insert the air plug into the air jack on the left side of the main unit correctly. 2. When the MODE button is pressed down, the “month” sign is displayed. When the SET button is pressed down, the figure for the month increases. Adjust the “month”. 2. When the SPHYG/CLOCK button is changed to the sphygmomanometer mode, all the LCD symbols are momentarily displayed.
Preparations for Use 4 How to Set the Printing Paper Please be sure not to print without inserting printing paper as it will shorten the life span of the printer head. 4. Insert the printing paper in the installation section, and then attach the cover back to the main unit in such a way that the edge of the paper protrudes from the cover. 1. Slide the plastic printer cover in the direction indicated by the arrow symbol and remove it from the main unit. 5.
Preparations for Use 5 Changing the Printer Ink Though not originally equipped with an ink roller, the HEM-703CP digital blood pressure monitor contains enough printer ink to last about two years when used once a day at normal room temperature. 3. Insert the supplement ink roller into the allotted space next to the printer. Supplement ink roller When the ink on the print out becomes light and hard to read, purchase a replacement ink roller from your retailer and install as fol-lows. 1.
Preparations for Use 6 Attach the Cuff This cuff has eliminated the need for a microphone thereby lessening possible interference from vibrations and noise. Fitting also has been made much easier through the widening of the artery oscillation detection area. 3. Place your arm on a table or a support with the palm facing upward so that the cuff is placed at the same level as the heart. 1. Place the cuff on the left arm so that the air tube lies across the center of the arm. Artery no space 2.
Preparations for Use 7 Apply Pressure to the Cuff Depress the START button continuously untill the cuff is inflated up to the targeted pressure. 1. When you press the start button, the pump operates and pressure is applied. When the pressure reaches 170mmHg, the pump stops automatically. After the cuff has been inflated, release the START button. Pressure being applied 2.
Preparations for Use 8 Blood Pressure and Pulse are displayed * When you wish to stop measuring during operation, press the SPHYGICLOCK button. The pump will stop and all the cuff pressure will be released rapidly. 1. When the unit has continued to deflate to a point when the artery oscillations are no longer detected, the unit will display the systolic and diastolic blood pressure values, along with a long beep, and causes the deflation indication symbol ( ) to start flashing.
When Error Indication Symbols Appear The following special error symbols will appear in the display if the unit is being used incorrectly: ERROR SYMBOL CAUSE 1. Cuff underinflated 2. Movement during measurement Residual air pressure in cuff. Cuff overinflated (more than 300 mmHg). No pulse or blood pressure reading due to: 1) Motion 2) Very weak artery oscillation 10 CORRECTION 1) Release air pressure in cuff 2) Wait 10 minutes before taking another reading.
Specifications Power Source: Battery (four type “AA” dry cell batteries) 6V 4W Measurement Method: Oscillometric Method Display: Liquid Crystal Digital Display Measurement Range: Pressure: 0 ~ 280mmHg Pulse: 40 ~ 200/minute Accuracy: Pressure: within ± 3 mmHg Pulse: within ± 5% reading Pressure Detection: Capacitive Pressure Sensor Inflation: Automatic Pressure Application by a Pump Deflation and Rapid Air Release: Automatic pressure release valve Automatic-shut-off System When the internal
This equipment generates and used radio frequency energy and if not installed and used properly, that is, in strict accordance with the manufacturer’s instructions, may cause interference to radio and television reception. It has been type tested and found to comply with the limits for a Class B computing device in accordance with the specifications in Subpart J of Part 15 of FCC Rules, which are designed to provide reasonable protection against such interference in a residential installation.
Limited One Year Warranty YOUR OMRON MODEL HEM-713C Digital Blood Pressure Monitor is warranted to be free from manufacturing defects for a period of one year under normal use. This warranty extends only to the original retail purchaser. Should repair be needed, within the warranty period, ship the unit and receipt as proof of purchase prepaid to OMRON HEALTHCARE, INC., 300 Lakeview Parkway, Vernon Hills, Illinois 60061, together with $5.00 for return shipping and insurance.