Datasheet

Temperature Meter K3MA-L K-47
Panel
indicators
Characteristics
Note: The indication accuracy of the K thermocouple at a temperature of 200 to 1300°C is ±2°C ±1 digit maximum.
The indication accuracy of the T and N thermocouples at a temperature of 100°C or less is ±2°C ±1 digit maximum.
The indicator accuracy of the U and L thermocouples at any temperature is ±2°C ±1 digit maximum.
The indication accuracy of the B thermocouple at a temperature of 400°C or less is unrestricted.
The indication accuracy of the R and S thermocouples at a temperature of 200°C or less is ±3°C ±1 digit maximum.
Measuring Ranges
Platinum-resistance Thermometer
Thermocouple
Indication accuracy (at 23±5°C)
(See note.)
Thermocouple:
(±0.5% of indication value or ±1°C, whichever greater) ±1 digit max.
Platinum-resistance thermometer:
(±0.5% of indication value or ±1°C, whichever greater) ±1 digit max.
Input Thermocouple: K, J, T, E, L, U, N, R, S, B
Platinum-resistance thermometer: JPt100, Pt100
Measurement method Double integral method
Sampling period 500 ms
Display refresh period Sampling period (sampling times multiplied by number of averaging times if average processing is se-
lected.)
Max. displayed digits 4 digits (1999 to 9999)
Display 7-segment digital display, Character height: 14.2 mm
Polarity display ” is displayed automatically with a negative input signal.
Zero display Leading zeros are not displayed.
Input shift Input shift equivalent to the setting value supported for all points within the sensor measurement range.
Hold function Max hold (maximum value), Min hold (minimum value)
Hysteresis setting Programmable with front-panel key inputs (0001 to 9999).
Other functions Display color change (green (red), green, red (green), red)
Average processing (simple average OFF/2/4/8 operations)
Setting change lockout
Parameter initialization
Output Relay contact (SPDT)
Delay in comparative outputs 1 s max.
Degree of protection Front panel: NEMA4X for indoor use (equivalent to IP66)
Rear case: IEC standard IP20
Terminals: IEC standard IP00 + finger protection (VDE0106/100)
Memory protection Non-volatile memory (EEPROM) (possible to rewrite 100,000 times)
Input Pt100 JPt100
Range °C 200 to 850 199.9 to 500.0 0.0 to 100.0 199.9 to 500.0 0.0 to 100.0
°F 300 to 1500 199.9 to 900.0 0.0 to 210.0 199.9 to 900.0 0.0 to 210.0
Parameter 01234
Input K J T E L U N R S B
Range °C 200
to
1300
20.0
to
500.0
100
to
850
20.0
to
400.0
200
to
400
199.9
to
400.0
0
to
600
100
to
850
200
to
400
199.9
to
400.0
200
to
1300
0
to
1700
0
to
1700
100
to
1800
°F 300
to
2300
0.0
to
900.0
100
to
1500
0.0
to
750
300
to
700
199.9
to
700.0
0
to
1100
100
to
1500
300
to
700
199.9
to
700.0
300
to
2300
0
to
3000
0
to
3000
300
to
3200
Parameter56789101112131415161718