Full Product Manual

32
Voltage dips, short
interruptions and
voltage variations on
power supply inputlines
IEC 61000-4-11
<5 % U
T (>95 % dip
in U
T)
for 0.5 cycle
Not Applicable. Not Applicable.
40 % U
T (60 % dip
in U
T)
for 5 cycles
70 % U
T (30 % dip
in U
T)
for 25 cycles
<5 % U
T (95 % dip
in U
T)
for 5 sec.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Note: U
T is the A.C. mains voltage prior to application of the test level.