FAQ

OMRON Battery-operated / AC Adapter-operated (or rechargeable)
Blood Pressure Monitor
Information for Accompanying Documents
in the Scope of IEC60601-1-2:2014
Important information regarding Electro Magnetic Compatibility (EMC)
This blood pressure monitor manufactured by OMRON HEALTHCARE Co., Ltd. conforms to IEC60601-1-2:2014 Electro
Magnetic Compatibility (EMC) standard. Nevertheless, special precautions need to be observed:
The use of accessories and cables other than those specied or provided by OMRON could result in increased
electromagnetic emission or decreased electromagnetic immunity of the monitor and result in improper operation.
During measurement, the use of the monitor adjacent to or stacked with other device should be avoided because it could
result in improper operation. In case such use is necessary, the monitor and other device should be observed to verify that
they are operating normally.
During measurement, portable RF communications device (including peripherals such as antenna cables and external
antennas) should be used no closer than 12 inches (30 cm) to any part of the monitor, including cables specied by OMRON.
Otherwise, degradation of the performance of the monitor could result.
Refer to further guidance below regarding the EMC environment in which the monitor should be used.
Table 1 - EMISSION Limits and Compliance
Phenomenon EMISSION Limits Compliance
Conducted and radiated RF EMISSIONS CISPR 11 Group1, Class B
Voltage uctuations and icker See IEC 61000-3-3 Complies
Table 2 - IMMUNITY TEST LEVELS
Phenomenon Basic EMC standard IMMUNITY TEST LEVELS
Electrostatic discharge IEC 61000-4-2 ±8 kV contact
±2 kV,±4 kV,±8 kV,±15 kV air
for enclosure port
Radiated RF electromagnetic elds IEC 61000-4-3 10 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
for enclosure port
Proximity elds from RF wireless communications equipment IEC 61000-4-3 See table 3
Electrical fast transients / bursts IEC 61000-4-4 ±2 kV for Input a.c. power port
100 kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5 ±0.5kV, ±1 kV
for Input a.c. power port
Conducted disturbances induced by RF elds IEC 61000-4-6 3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM and amateur radio bands
between 150 kHz and 80 MHz
80 % AM at 1 kHz
for Input a.c. power port
Rated power frequency magnetic elds IEC 61000-4-8 30 A/m
50 Hz and 60 Hz
for enclosure port
Voltage dips IEC 61000-4-11 0 % U
T
; 0.5 cycle
At 0°,45°, 90°, 135°, 180°, 225°, 270° and 315°
for Input a.c. power port
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles single phase: at 0°
for Input a.c. power port
Voltage interruptions IEC 61000-4-11 0 % U
T
; 250/300 cycle
for Input a.c. power port
NOTE: U
T
is the A.C. mains voltage prior to application of the test level.
Table 3 - Test specications for ENCLOSURE PORT IMMUNITY to RF wireless communications device
Test frequency
(MHz)
Band (MHz) Service Modulation Maximum
power (W)
Distance
(m)
IMMUNITY TEST
LEVEL (V/m)
385 380 to 390 TETRA 400 Pulse modulation 18 Hz 1.8 0.3 27
450 430 to 470 GMRS 460,
FRS 460
FM
± 5 kHz deviation
1 kHz sine
2 0.3 28
710 704 to 787 LTE Band 13, 17 Pulse modulation 217 Hz 0.2 0.3 9
745
780
810 800 to 960 GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation 18 Hz 2 0.3 28
870
930
1720 1700 to 1990 GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3, 4, 25; UMTS
Pulse modulation 217 Hz 2 0.3 28
1845
1970
2450 2400 to 2570 Bluetooth, WLAN,
802.11 b/g/n , RFID 2450,
LTE Band 7
Pulse modulation 217 Hz 2 0.3 28
5240 5100 to 5800 WLAN 802.11
a/n
Pulse modulation 217 Hz 0.2 0.3 9
5500
5785
EMC tests have included the AC adapter.