NE-C30_main.book Page 2 Tuesday, December 13, 2011 12:14 PM Before using the unit Contents Before using the unit Introduction ................................................................................................... 3 Intended use .................................................................................................. 3 Important safety instructions....................................................................... 4 1. Know your unit ...............................................
NE-C30_main.book Page 3 Tuesday, December 13, 2011 12:14 PM Introduction Thank you for purchasing the OMRON Comp Air Elite. This product was developed in conjunction with respiratory therapists for the successful treatment of asthma, chronic bronchitis, allergies and other respiratory disorders. The compressor forces air to the nebuliser. When the air enters the nebuliser, it converts the prescribed medication into an aerosol of microscopic droplets that can easily be inhaled.
NE-C30_main.book Page 4 Tuesday, December 13, 2011 12:14 PM Important safety instructions Read all the information in the instruction manual and any other literature included in the box before using the device. When using an electrical product, especially when children are present, basic safety precautions should always be followed including the following: Warning: Indicates a potentially hazardous situation which, if not avoided, could result in serious injury.
NE-C30_main.book Page 5 Tuesday, December 13, 2011 12:14 PM Important safety instructions • Do not use the main unit, AC adapter, optional rechargeable battery, or optional battery charger while they are wet, or connect them to a power outlet or other devices with wet hands. • The compressor, AC adapter, and optional rechargeable battery and charger are not waterproof. Do not spill water, or other liquids, on these parts.
NE-C30_main.book Page 6 Tuesday, December 13, 2011 12:14 PM Important safety instructions • Do not leave the unit unattended with infants or persons who cannot express their consent. • Do not subject the device or the parts to any strong shocks such as dropping the device on the floor. • Do not distort the vaporiser head, or jab the nozzle of the medication tank with a pin etc. • Do not cover the compressor, optional rechargeable battery and charger with a blanket or towel etc., during use.
NE-C30_main.book Page 7 Tuesday, December 13, 2011 12:14 PM Important safety instructions General Safety Precautions: • Inspect the device and parts before using them each time, and check that there are no problems. In particular, be sure to check the following: - That the vaporiser head, nozzle or air tube are not damaged. - That the nozzle is not blocked. - That the compressor operates normally. • When using this device, there will be some noise and vibration caused by the pump in the compressor.
NE-C30_main.book Page 8 Tuesday, December 13, 2011 12:14 PM 1. Know your unit A D C A1 A2 A3 D1 D2 D3 5 3 A4 A5 E A6 E1 B1 E2 E3 B2 B E4 Rear View Bottom View E5 E6 5 3 A7 A8 A. Compressor (Main Unit) A1. Power Indicator A2. Power Switch A3. Charge Indicator A4. Air Connector A5. Filter Cover (* Air filter inside.) A6. Nebuliser Kit Holder A7. AC Adapter Jack A8. Battery Connector B. Air Tube (PVC, 207cm) B1. Air Plug B2. Air Plug C. Nebulising Parts 8 E7 D.
NE-C30_main.book Page 9 Tuesday, December 13, 2011 12:14 PM 1. Know your unit Accessories F G2 H G1 G I F. Spare Air Filters x 5 G. AC Adapter G1. Power Connector G2. Power Cord H. Carrying Bag J I. Instruction Manual J.
NE-C30_main.book Page 10 Tuesday, December 13, 2011 12:14 PM 2. Guide to using the Comp Air Elite Preparing the Nebuliser for Use (Section 3) Note: Clean and disinfect the nebuliser kit and applicable inhalation attachments before using the device for the first time, or if it has not been used for more than one week. Taking Medication (Section 4) Notes: • Do not tilt the nebuliser kit so that it is at an angle of greater than 45°. • Do not crease or fold the air tube.
NE-C30_main.book Page 11 Tuesday, December 13, 2011 12:14 PM Operating instructions 3. Preparing the nebuliser for use Warning: Clean and disinfect the nebuliser kit, mouthpiece and optional nosepiece or masks before using them for the first time after purchase, if the device has not been used for a long period of time, or if more than one person uses the same device. For directions on cleaning and disinfecting the parts of the device refer to Section 5 in this instruction manual.
NE-C30_main.book Page 12 Tuesday, December 13, 2011 12:14 PM 3. Preparing the nebuliser for use 3.2 Adding medication and assembling the parts 1. Remove the mouthpiece and cap from the nebuliser kit. 2. Remove the inhalation top from the medication 2 tank. 1) Rotate the inhalation top anti-clockwise. 2) Lift the inhalation top out of the medication 1 tank. 5 3 3. Remove the vaporiser head from the medication tank. 4. Add the correct amount of prescribed medication to the medication tank.
NE-C30_main.book Page 13 Tuesday, December 13, 2011 12:14 PM 3. Preparing the nebuliser for use 5. Reattach the vaporiser head to the medication tank. 6. Put the inhalation top back inside the medication tank. 1 1) Align the protrusion on the inhalation top with the indent on the medication tank, as shown. 2) Rotate the inhalation top clockwise and align max 5 3 2 with the medication tank until they click in place. 7. Attach the desired inhalation attachment, as described below.
NE-C30_main.book Page 14 Tuesday, December 13, 2011 12:14 PM 3. Preparing the nebuliser for use Attaching the child mask (PVC) Attach the mask to the inhalation top. 1) Attach the cap to the inhalation air inlet. 2) Attach the child mask to the inhalation top. 2 1 Attaching the nosepiece Attach the cap when using the nosepiece. 1) Attach the cap to the inhalation air inlet. 2) Attach the nosepiece to the inhalation top.
NE-C30_main.book Page 15 Tuesday, December 13, 2011 12:14 PM 3. Preparing the nebuliser for use 8. Attach the air tube. 1) Attach the air tube to the air connector on the compressor. 2) Attach the other end of the air tube to the air tube connector on the medication tank. 2 1 Notes: • Make sure that the air tube is securely attached to the compressor (main unit) and nebuliser kit, and does not come loose.
NE-C30_main.book Page 16 Tuesday, December 13, 2011 12:14 PM 4. Taking medication 1. Hold the nebuliser kit as indicated on the right. max 5 3 Caution: Do not tilt the nebuliser kit so that it is at an angle of greater than 45°. Medication may flow into the mouth. 3 5 ma x 45° 2. Press the power switch. The compressor starts and nebulisation starts. Notes: • The green light shows that the power is on. • Observe whether aerosol is generated. To stop nebulisation Press the power switch again.
NE-C30_main.book Page 17 Tuesday, December 13, 2011 12:14 PM 4. Taking medication 3. Inhale medication according to your doctor’s and/or health care adviser’s instructions. Using the mouthpiece Prepare your mouthpiece according to Section 3.2.7. Insert the mouthpiece into your mouth and inhale the medication breathing normally. Exhale as normal through the mouthpiece. max 5 3 Using the nosepiece Prepare your nosepiece according to Section 3.2.7.
NE-C30_main.book Page 18 Tuesday, December 13, 2011 12:14 PM 4. Taking medication 4. When treatment is complete, turn the power off. Check that there is no condensation or moisture in the air tube and unplug the device from the power outlet. Power Outlet Warning: Condensation may occur in the air tube. Do not store the air tube while it contains condensation or moisture. This could result in infection via bacteria.
NE-C30_main.book Page 19 Tuesday, December 13, 2011 12:14 PM Care and maintenance 5. Cleaning and daily disinfecting 5.1 Cleaning Following the cleaning instructions after each use will prevent any remaining medication in the tank from drying, resulting in the device not nebulising effectively and also for preventing infections.
NE-C30_main.book Page 20 Tuesday, December 13, 2011 12:14 PM 5. Cleaning and daily disinfecting 6. After cleaning and disinfecting the parts, rinse them thoroughly with clean water and allow the parts to air dry in a clean environment. 5 3 7. If the compressor is dirty, wipe it clean with a soft Caution: The compressor is not waterproof. Mild Detergent cloth moistened with water or mild detergent. 8.
NE-C30_main.book Page 21 Tuesday, December 13, 2011 12:14 PM 5. Cleaning and daily disinfecting Handling the Vaporiser Head The vaporiser head is an important part that is used to vaporise medication. Be sure to observe the following precautions when handling it. Caution: • Always wash the vaporiser head after each Baffle use. Failure to do so could cause blockages and a deterioration in your condition due to a failure to vaporise medication effectively. • Do not use a brush or pin etc.
NE-C30_main.book Page 22 Tuesday, December 13, 2011 12:14 PM 6. Maintenance and storage To keep your device in the best condition and protect the device from damage follow these directions: Caution: • Do not leave the unit unattended with infants or persons who cannot express their consent. • Do not subject the device or the parts to any strong shocks such as dropping the device on the floor. Notes: • Do not store the device in extreme hot or cold temperatures, high humidity or under direct sunlight.
NE-C30_main.book Page 23 Tuesday, December 13, 2011 12:14 PM 7. Changing the air filter On average, the air filter should be changed every 60 days or so, or when the filter has changed colour from white. 1. Pull the air filter cover off the compressor. 2. Change the air filter. Remove the old air filter with a sharp object, such as a toothpick, and insert the new air filter. 3. Put the air filter cover back in place. Notes: • Use only OMRON air filters designed for this unit.
NE-C30_main.book Page 24 Tuesday, December 13, 2011 12:14 PM 8. Troubleshooting Check the following if your unit should fail during operation. You can also refer to the pages of this manual for complete instructions. Symptom Cause Remedy Check that the plug is Is the AC adapter connected inserted in an electrical correctly to the an electrical outlet. Unplug then reinsert outlet and the compressor? the plug if necessary. Nothing happens when the power switch is pressed.
NE-C30_main.book Page 25 Tuesday, December 13, 2011 12:14 PM 8. Troubleshooting Symptom The device is very hot. Cause The compressor is covered by a cloth. Remove the cloth. Is the optional rechargeable battery charged sufficiently? Charge the battery. The battery has expired and The device stops soon when cannot be recharged enough. using the optional rechargeable battery, or operation time is short. Remedy Has the battery been left unused for a long time? Replace with a new battery.
NE-C30_main.book Page 26 Tuesday, December 13, 2011 12:14 PM 9. Technical data Name: Model: Type: Power Supply: Power Consumption: Operating Temperature/ Humidity: Storage Temperature/ Humidity/Air Pressure: Weight: Dimensions: Contents: Comp Air Elite OMRON NE-C30-E Non-heating compressor nebuliser 12VDC 1A (Compressor) 100 - 240VAC, 50/60Hz (AC adapter) 12 watts +10°C to +40°C 30% to 85% RH -20°C to +60°C 10% to 95% RH 700 - 1060 hPa 440 g (compressor only) Approx.
NE-C30_main.book Page 27 Tuesday, December 13, 2011 12:14 PM 9. Technical data Technical data for the OMRON Comp Air compressor with the OMRON V.V.T. (Virtual Valve Technology) Nebuliser kit: Particle Size: Medication Tank Capacity: Appropriate Medication Quantities: Sound: Nebulisation Rate: Aerosol Output: Aerosol Output Rate: *MMAD **approximately 3µm 7 ml maximum 2 ml minimum - 7 ml maximum Noise level (at 1 m distance) 53 db 0.35 ml/min (by weight loss) **0.3 ml (2ml, 1%NaF) **0.
NE-C30_main.book Page 28 Tuesday, December 13, 2011 12:14 PM 9. Technical data Notes: TypeBB == Type • Do not use the device where it may be exposed to flammable gas. • This device fulfils the provisions of the EC directive 93/42/EEC (Medical Device Directive) and the European Standard EN13544-1:2007, Respiratory therapy equipment - Part1: Nebulizing systems and their components.
NE-C30_main.book Page 29 Tuesday, December 13, 2011 12:14 PM 9. Technical data Correct Disposal of This Product (Waste Electrical & Electronic Equipment) This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life.
NE-C30_main.book Page 30 Tuesday, December 13, 2011 12:14 PM 10.Optional parts and accessories OMRON V.V.T. Nebuliser Kit Set Air Tube (PVC, 207cm) Mouthpiece Order No. 9956270-0 Order No. 9956273-5 Air Tube (Silicon, 100cm) Nosepiece Order No. 9956269-7 Order No. 9956274-3 Child Mask (SEBS) Set Adult Mask (SEBS) Adult Mask (PVC) Order No. 9956281-6 Order No. 9956312-0 Order No. 9956275-1 Child Mask (PVC) Air Filters (x 5) Order No. 9956276-0 Order No. 9956636-6 Order No.
NE-C30_main.book Page 31 Tuesday, December 13, 2011 12:14 PM 10. Optional parts and accessories Filter Cover Instruction Manual Carrying Bag Order No.9956638-2 Order No.9956633-1 Order No.9956631-5 Battery Set Rechargeable Battery Battery Charger Includes the rechargeable battery, battery charger. Order No.9986453-7 Order No.9986457-0 Order No.9986461-8 AC Adapter Order No.
NE-C30_main.book Page 32 Tuesday, December 13, 2011 12:14 PM 11.Using optional accessories 11.1 Using the optional rechargeable battery Use the rechargeable battery as shown in the chart below. For details, refer to the following pages. When the battery has not been used for a while. Using the battery for the first time. Fully charge the battery. Attach the battery to the main unit and use the device. Attach the battery to the main unit. Notes: • Limit use to 20 minutes at a time.
NE-C30_main.book Page 33 Tuesday, December 13, 2011 12:14 PM 11. Using optional accessories Important notes on using the optional rechargeable battery • Charge the battery before using it for the first time. The battery is not charged at the time of shipment. • Use the optional charger to charge the battery. The battery takes about 3 hours to charge fully. • A fully charged battery can be used to operate the device for approximately 35 minutes, when used within the normal operating temperature range.
NE-C30_main.book Page 34 Tuesday, December 13, 2011 12:14 PM 11. Using optional accessories 3) Attach the optional battery to the Charging Indicator blinks charger. Align the connectors on the battery and charger, and press the battery until it clicks into place. Charging starts and the green charging indicator blinks. 3 2. After the battery is fully charged, remove the battery from the charger.
NE-C30_main.book Page 35 Tuesday, December 13, 2011 12:14 PM 11. Using optional accessories When battery power is low The charge indicator on the main unit will light when battery power is low. Please note the following points about using the device after the charge indicator lights: • Stop using the device when the charge indicator comes on. • Battery power is low and the device will cease to Charge Indicator operate shortly after the charge indicator comes on.
NE-C30_main.book Page 36 Tuesday, December 13, 2011 12:14 PM 11. Using optional accessories Manufacturer OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS www.omron-healthcare.com Production facility OMRON (DALIAN) CO., LTD. Dalian, CHINA OMRON HEALTHCARE UK LTD. Opal Drive Fox Milne, Milton Keynes, MK15 0DG, U.K.