INSTRUCTION MANUAL 3 series Automatic Wrist Blood Pressure Monitor ENGLISH ESPAÑOL Model BP629
TABLE OF CONTENTS Before using the Monitor Introduction ......................................................................................................... 3 Safety Information ................................................................................................ 4 Operating the Device .......................................................................... 4 Care and Maintenance .........................................................................
INTRODUCTION Thank you for purchasing the OMRON® BP629 IntelliSense® Automatic Wrist Blood Pressure Monitor. Fill in for future reference. DATE PURCHASED:__________________ SERIAL NUMBER:____________________ s 3TAPLE YOUR PURCHASE RECEIPT HERE s 2EGISTER YOUR PRODUCT ON LINE AT WWW REGISTER OMRON COM The monitor is compact and easy-to-use at home, at work, and during travel. Perfect for people who frequently monitor their own blood pressure.
SAFETY INFORMATION To assure the correct use of the product, basic safety measures should always be followed including the warnings and cautions listed in this instruction manual. SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
SAFETY INFORMATION Do not use a cellular phone near the device. It may result in an operational failure. Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit. 8VH RQO\ 9 ³$$$´ DONDOLQH EDWWHULHV ZLWK WKLV GHYLFH 'R QRW XVH RWKHU W\SHV RI batteries. This may damage the unit.
BEFORE TAKING A MEASUREMENT To ensure a reliable reading follow these recommendations: 1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement. 2. Stress raises blood pressure. Avoid taking measurements during stressful times. 3. The cuff can be applied to your left or right wrist. 4. Measurements should be taken in a quiet place. 5. Position the unit at heart level throughout the measurement.
KNOW YOUR UNIT Main Unit: Display Wrist cuff Memory (MEM) button START/STOP button Battery compartment Date/Time Setting ( Date/Time Setting ( ) button ) button Components: INSTRUCTION MANUAL 3 series Automatic Wrist Blood Pressure Monitor Model BP629 5329416-7A INTRODUCTION Thank you for purchasing the OMRON® BP629 IntelliSense® Automatic Wrist Blood Pressure Monitor . Fill in for future reference.
UNIT DISPLAY Systolic blood pressure Diastolic blood pressure Memory symbol Average value symbol Irregular heartbeat symbol Blood pressure level indicator Deflation symbol 8 Pulse display Movement error symbol Battery low symbol Date/Time display Heartbeat symbol (Flashes during measurement)
DISPLAY SYMBOLS IRREGULAR HEARTBEAT SYMBOL ( ) When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the measurement values. $Q LUUHJXODU KHDUWEHDW UK\WKP LV GH¿QHG DV D .ORMAL (EARTBEAT rhythm that varies by less than 25% from the Pulse average rhythm or 25% more than the average Blood rhythm detected while the monitor is measuring pressure the systolic and diastolic blood pressure.
DISPLAY SYMBOLS This is a general guideline because blood pressure depends on age and morbidity. A lower home blood pressure goal may be recommended for certain patients, including diabetic patients, pregnant women, and patients with renal failure. WARNING &RQWDFW \RXU SK\VLFLDQ IRU VSHFL¿F LQIRUPDWLRQ DERXW \RXU EORRG SUHVVXUH Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider.
BATTERY INSTALLATION CAUTION 8VH RQO\ 9 ³$$$´ DONDOLQH EDWWHULHV ZLWK WKLV GHYLFH 1. Press the hook on the bottom of the battery cover and lift the cover off in the direction of the arrow. ,QVWDOO ³$$$´ VL]H EDWWHULHV VR WKH SRVLWLYH DQG (negative) polarities match the polarities of the battery compartment as indicated. 3. Set the date and time. 5HIHU WR ³6HWWLQJ WKH 'DWH DQG 7LPH´ 4. Replace the battery cover. NOTE: Make sure that the battery cover is securely in position.
BATTERY INSTALLATION BATTERY REPLACEMENT Low Battery Indicator When the Low Battery Indicator appears on the display, turn the monitor off and remove all the batteries. Replace with 2 new batteries at the same time. Long-life alkaline batteries are recommended. WARNING ,I EDWWHU\ ÀXLG VKRXOG JHW LQ \RXU H\HV LPPHGLDWHO\ ULQVH ZLWK plenty of clean water. Contact a physician immediately.
SETTING THE DATE AND TIME 6HW WKH PRQLWRU WR WKH FXUUHQW GDWH DQG WLPH EHIRUH WDNLQJ D PHDVXUHPHQW IRU WKH ¿UVW time. If the date and time has not been set, the measurement values are stored without any date or time information. 1. TO START THE SETTING Press the Date/Time Setting button ( ) to adjust the date and time. 7KH \HDU ÀDVKHV RQ WKH GLVSOD\ 2. SETTING THE YEAR Press the Date/Time Setting button ( ) to advance by increments of one year.
SETTING THE DATE AND TIME 3. SETTING THE MONTH Press the Date/Time Setting button ( ) to advance by increments of one month. Press the Date/Time Setting button ( ) to set the current month. 7KH GD\ ÀDVKHV RQ WKH GLVSOD\ 4. SETTING THE DAY Press the Date/Time Setting button ( ) to advance by increments of one day. Press the Date/Time Setting button ( ) to set the current day. 7KH KRXU ÀDVKHV RQ WKH GLVSOD\ 5. SETTING THE HOUR The time is set using AM or PM.
APPLYING THE WRIST CUFF CAUTION Read all of the information in the instruction manual and any other literature in the box before operating the unit. CAUTION This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions. APPLYING THE CUFF ON THE LEFT WRIST 1. Roll up sleeve. Make sure your sleeve is not rolled up too WLJKWO\ RQ \RXU DUP 7KLV PD\ FRQVWULFW WKH ÀRZ RI EORRG LQ your arm. 2.
APPLYING THE WRIST CUFF 5. Fold the remaining part of the wrist cuff back out of the way. APPLYING THE CUFF ON THE RIGHT WRIST When taking a measurement using the right wrist position the cuff as shown in the illustration.
TAKING A MEASUREMENT 6LW FRPIRUWDEO\ LQ D FKDLU ZLWK \RXU IHHW ÀDW RQ WKH ÀRRU 2. Hold your arm across your chest at heart level. +ROG \RXU HOERZ ¿UPO\ WR DYRLG ERG\ PRYHPHQW Sit still and do not talk or move during the measurement. 3. Press the START/STOP button. All symbols appear on the display.
TAKING A MEASUREMENT 7KH FXII VWDUWV WR LQÀDWH DXWRPDWLFDOO\ $V WKH FXII LQÀDWHV WKH PRQLWRU DXWRPDWLFDOO\ GHWHUPLQHV \RXU LGHDO LQÀDWLRQ OHYHO 7KLV PRQLWRU GHWHFWV \RXU EORRG SUHVVXUH DQG SXOVH UDWH GXULQJ LQÀDWLRQ The Heartbeat Symbol ( ÀDVKHV DW HYHU\ KHDUWEHDW Remain still and do not move until the entire measurement process is completed. NOTE: 7R VWRS WKH LQÀDWLRQ RU PHDVXUHPHQW SXVK WKH 67$57 6723 EXWWRQ 7KH PRQLWRU ZLOO VWRS LQÀDWLQJ VWDUW GHÀDWLQJ DQG ZLOO WXUQ RII 4.
USING THE MEMORY FUNCTION Every time you complete a measurement the monitor stores the blood pressure and pulse rate in the memory. The monitor automatically stores up to 60 sets of measurement values (blood pressure and pulse rate). When 60 sets of measurement values are stored, the oldest record is deleted to save the most recent values.
USING THE MEMORY FUNCTION NOTES: 3UHVV WKH 0(0 EXWWRQ UHSHDWHGO\ WR GLVSOD\ WKH QH[W YDOXHV 3UHVV DQG KROG WKH 0(0 EXWWRQ WR GLVSOD\ WKH YDOXHV IDVWHU The date and time are alternately displayed with the measurement values. 3. Press the START/STOP button to turn the monitor off. TO DELETE ALL VALUES STORED IN THE MEMORY You cannot partially delete values stored in the memory. All values will be deleted. 1. Press the START/STOP button to turn the monitor on. 2.
CARE AND MAINTENANCE To keep your digital blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed below: Do not forcefully bend the cuff. Do not fold tightly. Clean the monitor with a soft dry cloth. Do not use any abrasive or volatile cleaners. Do not attempt to clean the cuff. CAUTION Do not submerge the device or any of the components in water. Do not subject the monitor to extreme hot or cold temperatures, humidity or direct sunlight.
ERROR INDICATORS SYMBOL CAUSE CORRECTION Take another measurement and remain still until the measurement is complete. Monitor could not detect pulse wave. &XII XQGHU LQÀDWHG Cuff not applied correctly. Remove the monitor from your wrist. 5HDG ³7DNLQJ D 0HDVXUHPHQW´ &XII RYHU LQÀDWHG Wait 2-3 minutes. Take another measurement. Device error. Contact Customer Service. Movement during measurement. Keep still and take another measurement. Irregular or weak pulses are detected.
TROUBLESHOOTING TIPS PROBLEM CAUSES AND SOLUTIONS No power. No display appears on the unit. Replace both batteries with new ones at the same time. Check the battery installation for proper placement of the battery polarities. Measurement values appear too high or too low. Blood pressure varies constantly. Many factors including stress, time of day, how you wrap the cuff, may affect your blood pressure.
FCC STATEMENT Note: POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
LIMITED WARRANTY Your BP629 IntelliSense® Automatic Wrist Blood Pressure Monitor, excluding the wrist cuff and batteries, is warranted to be free from defects in materials and workmanship appearing within 2 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The above warranties extend only to the original retail purchaser. We will, at our option, repair or replace without charge any monitor or wrist cuff covered by the above warranties.
SPECIFICATIONS BP629 (HEM-6200-Z) LCD Digital Display Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min. Pressure: ±3mmHg or 2% of reading Pulse: ±5% of reading Automatic by electric pump Inflation: Automatic pressure release valve Deflation: Automatic exhaust valve Rapid Air Release: Measurement Method: Oscillometric method 9 ³$$$´ EDWWHULHV Power Source: Approx.
Omron Healthcare, Inc. 1200 Lakeside Drive Bannockburn, IL 60015 www.omronhealthcare.com Copyright © 2010 Omron Healthcare, Inc.