Product includes: El producto incluye: English Battery Set Juego de Pilas Español ComFit™ Cuff Brazalete ComFit™ Monitor Monitor or English Español 7 series Blood Pres sure Mon Model BP7 60N Instruction Manual 2298836-5C ii itor Instruction Manual Manual de Instrucciones Quick Start Guide Guía resumida
Contents Thank you for purchasing the OMRON® BP760N Blood Pressure Monitor. Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
Important Safety Information Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure. The monitor is not intended to be a diagnostic device.
Important Safety Information Caution: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. (General Usage) Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous. Consult your physician before using the device for any of the following conditions: •If you have had a mastectomy. •Do not take measurements more than necessary.
Important Safety Information When handling the power cord, take care not to do the following: Do not damage. Do not break it. Do not tamper with it. Do not forcibly bend or pull. Do not twist. Do not bundle during use. Do not pinch. Do not place under heavy objects. Wipe the dust off from the power plug. Unplug monitor when not in use. Disconnect the power plug before cleaning. Use only an OMRON AC adapter designed for this device.
1. Know Your Device Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page.
1. Know Your Device 1.1 Display Symbols Irregular Heartbeat Symbol ( ) When the monitor detects an irregular rhythm two or more times during the Pulse measurement, the irregular heartbeat symbol will appear on the display with Blood pressure the measurement values. An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure.
1. Know Your Device Blood Pressure Color Indicator If your Systolic Blood Pressure is 135 mmHg or above and/or the Diastolic Blood Pressure is 85 mmHg or above, the blood pressure color indicator will light in “orange” when the measurement result is displayed. If the results are within the standard range, the blood pressure color indicator will light in “green”.
2. Preparation 2.1 Battery Installation 1. Remove the battery cover. 2. Insert 4 “AA” batteries as EN indicated in the battery compartment. 3. Replace the battery cover. Notes: • When the low battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same time. Long life alkaline batteries are recommended. • The measurement values continue to be stored in memory even after the batteries are replaced.
2. Preparation Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page. 2.2 Setting the Date and Time Set the monitor to the correct date and time before taking a measurement for the first time. 1. Press the C Date/Time setting ( ) button. The year flashes on the T Date/Time display. 2. Push or button to change the year. Push button to confirm the year and then the month flashes.
3. Using the Device Open both the front and rear covers to read the following: The letter identifiers on the cover page correspond to those in the body of this page. 3.1 Applying the Arm Cuff Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm. Do not place the arm cuff over thick clothes. EN 1. Insert L the air plug into A the air jack securely. 2. Apply the arm cuff to your left upper arm. The bottom edge of the arm cuff should be a 1/2 inch (1 to 2 cm) above the elbow.
3. Using the Device 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement. • • • • 10 Sit on a chair with your feet flat on the floor. Sit upright with your back straight. --- c Sit with your back and arm being supported. The arm cuff should be placed on your arm at the same level as your heart.
3. Using the Device 3.3 Taking a Measurement Notes: • To stop the measurement, press the START/STOP button once to deflate the arm cuff. • Remain still and quiet while taking a measurement. The monitor is designed to take measurements and store the measurement values in the memory for 2 people using USER ID 1 and USER ID 2. EN 1. Select your USER ID (1 or 2). 2. Press the START/STOP button. The arm cuff starts to inflate automatically.
3. Using the Device 3. Remove the arm cuff. 4. Press the START/STOP button to turn the monitor off. The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes. Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement. Using the Guest Mode The monitor stores measurement values for 2 users in the memory.
3. Using the Device DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure. This device is not intended to be a diagnostic device. Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous. Inflating to a higher pressure than necessary may result in bruising where the arm cuff is applied.
3. Using the Device 3.4 Using the Memory Function The monitor automatically stores the results up to 60 sets for each user (1 and 2). It can also calculate an average value based on the last 3 measurement values taken within 10 minutes. Notes: • If there are only 2 measurement values in the memory for that period, the average will be based on these 2 values. • If there is 1 measurement value in the memory for that period, this is displayed as the average.
3. Using the Device 3. Press the or button to view the values stored in the memory. : To view the older values : To view the more recent values Note: The value with the TruReadTM symbol is the average for the consecutive 3 measurements. To display the individual measurement values, press the EN button while the average is displayed. To View the Average Value 1. Select your USER ID (1 or 2). 2. Press and hold the button for more than 3 seconds.
3. Using the Device To Delete All the Values Stored in Memory The values stored in the memory are deleted by USER ID. 1. Select your USER ID (1 or 2). 2. Press the Memory button, while the memory symbol ( appears. 3. While holding the button down, press the START/STOP button for more than 3 seconds. Note: You cannot partially delete the values stored in the memory. All values for the user you select will be deleted.
4. Error Messages and Troubleshooting 4.1 Error Messages Display Cause Irregular heartbeats are detected. Solution Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Repeat the steps in section 3.3. If this error continues to appear, contact your physician. Movement during measurement. Carefully read and repeat the steps in section 3.3. The batteries are low. You should replace the batteries with new ones ahead of time. Refer to section 2.1. The batteries are exhausted.
4. Error Messages and Troubleshooting Display Cause Air plug disconnected. Insert the air plug securely. Refer to section 3.1. Arm cuff is applied too loosely. Apply the arm cuff tighter. Refer to section 3.1. Air is leaking from the arm cuff. Replace the arm cuff with a new one. Refer to section 5.3. Movement during measurement and the arm cuff has not been inflated sufficiently. 18 Solution Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3.
4. Error Messages and Troubleshooting 4.2 Troubleshooting Problem Cause and Solution No power. No display appears on the monitor. Replace all batteries with new ones. Check the battery installation for proper placement of the battery polarities. Refer to section 2.1. Measurement values appear too high or too low. Blood pressure varies constantly. Many factors including stress, time of day, and how you wrap the cuff, may affect your blood pressure. Review the section 1.2 and section 3.3.
5. Maintenance and Storage 5.1 Maintenance To protect your device from damage, please observe the following: • • • • Store the device and the components in a clean, safe location. Do not use any abrasive or volatile cleaners. Do not wash the device and any components or immerse them in water. Do not use gasoline, thinners or similar solvents to clean the device. • Use a soft and dry cloth, or a soft and moistened cloth and neutral soap to clean on the monitor and the arm cuff.
5. Maintenance and Storage 5.2 Storage 1. Unplug the air plug from the air jack. 2. Gently fold the air tube into the EN arm cuff. Note: Do not bend or crease the air tube excessively. Do not store the device in the following situations: • If the device is wet. • Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapors such as bleach. • Locations exposed to vibrations, shocks or where it will be at an angle.
5. Maintenance and Storage 5.
5. Maintenance and Storage Using the Optional AC Adapter 1. Insert the AC adapter plug into the AC adapter jack on the rear side of the monitor. EN 2. Plug the AC adapter into an electrical outlet. To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the monitor.
6. Specifications Model Display Measurement range Accuracy Inflation Deflation Measurement method IP classification Power source Battery life Operating temperature / humidity Storage temperature / humidity / air pressure Weight Dimensions Cuff circumference Memory Contents Applied part Protection against electric shock BP760N HEM-7320-Z LCD digital display Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats / min.
7. FCC Statement FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
8. Limited Warranty Your BP760N Automatic Blood Pressure Monitor, excluding the arm cuff and batteries, is warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The arm cuff is warranted to be free from defects in materials and workmanship appearing within 1 year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor.
9. Guidance and Manufacturer’s Declaration OMRON Automatic Blood Pressure Monitor Information for accompanying documents in the scope of IEC60601-1-2:2007 Model: BP760N including AC-adapter Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices.
9. Guidance and Manufacturer’s Declaration Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP760N including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP760N including AC-adapter should assure that it is used in such environment.
9. Guidance and Manufacturer’s Declaration Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP760N including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP760N including AC-adapter should assure that it is used in such environment.
9. Guidance and Manufacturer’s Declaration Recommended separation distance between portable and mobile RF communications equipment and the OMRON BP760N including AC-adapter OMRON BP760N including AC-adapter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
31
9.
E G I F H J A G H B C N D T O P K L Q U V R W S X Y Z M
1 2 b A L c d a M
Manufactured for : OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN Distributed by : OMRON HEALTHCARE, INC. 1925 West Field Court Lake Forest, IL 60045 U.S.A. www.omronhealthcare.com © 2014 OMRON HEALTHCARE, INC. Made in Vietnam Fabricado para : OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPÓN Distribuido por : OMRON HEALTHCARE, INC. 1925 West Field Court Lake Forest, IL 60045 U.S.A. www.omronhealthcare.com © 2014 OMRON HEALTHCARE, INC.