7 series Upper Arm Blood Pressure Monitor Model BP7350 Instruction Manual 3608691-7G
Table of Contents Introduction ................................................2 9. Optional Accessories ......................39 Safety Instructions ...............................2 Intended Use .......................................2 Receiving and Inspection .....................2 Symbols Glossary ................................2 Using the AC Adapter (optional accessory) .........................................40 Important Safety Information ....................3 1. Know Your Monitor ..............
Introduction Thank you for purchasing the OMRON BP7350 Blood Pressure Monitor. This monitor can be used by multiple users. This blood pressure monitor uses the oscillometric method of blood pressure measurement. This means this monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. Safety Instructions This instruction manual provides you with important information about the OMRON BP7350 Blood Pressure Monitor.
Important Safety Information Read the Important Safety Information in this instruction manual before using this monitor. Follow this instruction manual thoroughly for your safety. Keep for future reference. For specific information about your own blood pressure, CONSULT WITH YOUR PHYSICIAN. Indicates a potentially hazardous situation which, if not Warning avoided, could result in death or serious injury. • DO NOT use this monitor on infants, toddlers, children or persons who cannot express themselves.
Important Safety Information AC Adapter (optional accessory) Handling and Usage • DO NOT use the AC adapter if this monitor or the AC adapter cable is damaged. If this monitor or the cable is damaged, turn off the power and unplug the AC adapter immediately. • Plug the AC adapter into the appropriate voltage outlet. DO NOT use in a multi-outlet plug. • NEVER plug in or unplug the AC adapter from the electric outlet with wet hands. • DO NOT disassemble or attempt to repair the AC adapter.
Important Safety Information • DO NOT use this monitor in places with high or low humidity or high or low temperatures. Refer to section 10. • During measurement, observe the arm to ensure that the monitor is not causing prolonged impairment to blood circulation. • DO NOT use this monitor in high-use environments such as medical clinics or physician offices. • DO NOT use this monitor with other medical electrical (ME) equipment simultaneously.
Important Safety Information Data Transmission • DO NOT replace batteries or unplug the AC adapter while your readings are being transferred to your smart device. This may result in incorrect operation of this monitor and failure to transfer your blood pressure data. AC Adapter (optional accessory) Handling and Usage • Fully insert the AC adapter into the outlet. • When unplugging the AC adapter from the outlet, be sure to safely pull from the AC adapter. DO NOT pull from the AC adapter cable.
1. Know Your Monitor 1.1 Contents/Product Includes Monitor, Easy Wrap ComFit Cuff, 4 “AA” batteries, instruction manual, quick start guide 1.
1.
1. Know Your Monitor 1.
1. Know Your Monitor A Average value symbol Flashes when viewing the average of the latest 2 or 3 readings taken within the most recent 10 minute timeframe. B Memory symbol Appears when viewing readings stored in the memory. C Systolic blood pressure reading Comparison display D Displays past readings. Refer to section 5 for more information. Pulse display / Memory number When pressing the button, the memory number appears for E approximately 1.5 seconds before the pulse rate appears on the display.
1. Know Your Monitor K Sync symbol Flashes/appears when your data needs to be transferred because the stored memory is either almost, or completely full. Once you pair your monitor with your smart device, transfer your readings immediately before the monitor deletes the oldest reading. Up to 60 readings can be stored in the internal memory of your monitor per user. Low battery symbol Flashes when batteries are low. L Depleted battery symbol Appears when batteries are depleted.
1. Know Your Monitor Irregular heartbeat symbol Appears when irregular rhythm is detected 2 or more times during a measurement. The symbol with the number appears as below: Irregular heartbeat is detected 2 times: Irregular heartbeat is detected 3 times or more: Q or An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while your monitor is measuring blood pressure.
1. Know Your Monitor 1.4 Classification of BP (Blood Pressure) ACC/AHA 2017 Hypertension guidelines define categories of BP in adults as below. Categories of BP in Adults* BP Category Normal Systolic BP < 120 mmHg Elevated Diastolic BP and < 80 mmHg 120 - 129 mmHg and < 80 mmHg Hypertension Stage 1 Stage 2 130 - 139 mmHg or 80 - 89 mmHg ≥ 140 mmHg or ≥ 90 mmHg * Individuals with Systolic BP and Diastolic BP in 2 categories should be designated to the higher BP category.
2. Preparation 2.1 Installing Batteries 1. Push down the hook of the battery cover and pull downward. 2. Insert 4 “AA” batteries as indicated in the battery compartment. 3. Close the battery cover. As soon as batteries are correctly installed, the year flashes on the display. Refer to sub-section 2.2 to set the date and time manually, or refer to sub-section 4.1 to pair your monitor to set the date and time automatically.
2. Preparation 2.2 Setting the Date and Time Note •Set your monitor to the correct date and time before taking a measurement for the first time. •The date and time will automatically be set when the monitor is paired with “OMRON connect US/CAN” app. Refer to sub-section 4.1. •As soon as batteries are correctly installed, the year flashes on the display of your monitor. 1. If nothing is displayed, press button (rear surface of the monitor) to display the year.
2. Preparation 3. Press the button to confirm the year, then the month will flash. Repeat the same steps to adjust the month, day, hour, and minute. Year Month Day Hour Minute 4. Press the button to store the setting. Your monitor automatically turns off 3 seconds after the minutes have been set in this sequence. Note •If you need to reset the date and time, replace batteries or press the button, then set the date and time as shown in above steps.
2. Preparation 2.3 Blood Pressure Measurement Tips To help ensure an accurate reading, follow these directions: •Stress raises blood pressure. Avoid taking measurements during stressful times. •Measurements should be taken in a quiet place. •It is important to take measurements at the same times each day. Taking measurements in the morning and in the evening is recommended. •Remember to have a record of your blood pressure and pulse readings for your physician.
2. Preparation 2.4 Applying the Arm Cuff Note •The following steps are for applying the arm cuff to your left arm. When you take a measurement on your right arm, follow the cuff wrap instructions for a right arm use mentioned under “To take a measurement on your right arm” located at the end of this subsection. •The blood pressure can differ between the right arm and the left arm, and the measured blood pressure readings can be different. OMRON recommends to always use the same arm for measurement.
2. Preparation 2. Apply the arm cuff to your left upper arm. 0.5 inch (1 to 2 cm) Bottom edge Note • The bottom edge of the arm cuff should be 0.5 inch (1 to 2 cm) above the inside elbow. The air tube is on the inside of your arm and aligned with your middle finger. 3. Make sure that the air tube is positioned on the inside of your arm and wrap the cuff securely, so it can not move around your arm.
2. Preparation 2.5 Sitting Correctly To take a measurement, you need to be relaxed and comfortably seated in a room at a comfortable temperature. Place your arm on the table. •Sit in a comfortable chair with your back and arm supported. •Keep your feet flat and your legs uncrossed. •The arm cuff should be placed on your arm at the same level as your heart, with the arm resting comfortably on a table.
2. Preparation 2.6 Selecting a User ID Your monitor is designed to take measurements and store the readings in the memory for 2 people using the [user ID selection] switch. 1. Select your user ID (1 or 2).
3. Using Your Monitor without a Smart Device 3.1 Taking a Measurement You can use your monitor without pairing to a smart device. To pair your monitor with a smart device, refer to sub-section 4.1. Note •To stop the measurement, press the [START/STOP] button once to deflate the arm cuff. Caution • DO NOT use this monitor with other medical electrical (ME) equipment simultaneously. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
3. Using Your Monitor without a Smart Device 3. Press the [START/STOP] button to turn your monitor off. Note • Your monitor automatically turns off after 3 minutes. • Wait 2 - 3 minutes between measurements. The wait time allows the arteries to decompress and return to their premeasurement state. You may need to increase the wait time depending on your individual physiological characteristics. Using the Guest Mode Your monitor stores readings for 2 users in the memory.
3. Using Your Monitor without a Smart Device If your systolic pressure is more than 210 mmHg If your systolic pressure is more than 210 mmHg, take a measurement manually. After the arm cuff starts to inflate, press and hold the [START/STOP] button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure. Note •Do not inflate above 299 mmHg. Caution • Inflating to a higher pressure than necessary may result in bruising where the arm cuff is applied.
4. Using Your Monitor with a Smart Device 4.1 Pairing Your Monitor with a Smart Device The date and time on your monitor will automatically be set when you pair your monitor with your smart device. Review the list of compatible smart devices at OmronHealthcare.com 1. Enable Bluetooth on your smart device. 2. Download and install the free “OMRON connect US/CAN” app onto your smart device.
4. Using Your Monitor with a Smart Device 5. Press [START/STOP] button to turn your monitor off. Note • If the “Err” symbol appears, follow the instructions in the “OMRON connect US/CAN” app. • Your monitor automatically turns off after 2 minutes. • Be aware that OMRON will not be responsible for the loss of data and/or information in the app. • “OMRON connect US/CAN” is the only app that we recommend to use with your monitor to transfer data correctly. 4.2 Taking a Measurement Refer to sub-section 3.
4. Using Your Monitor with a Smart Device 4.3 Disabling/Enabling Bluetooth Warning • This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this product in locations where RF is restricted, such as on an aircraft or in hospitals. Turn off the Bluetooth feature in this monitor and remove batteries when in RF restricted areas. For further information on potential restrictions refer to documentation on the Bluetooth usage by the FCC. Bluetooth is enabled as default. 1.
5. Using Memory Function Your monitor automatically stores up to 60 readings per user. Note •This monitor can be used by multiple users. Refer to sub-section 2.6 to switch user. Make sure which user ID stores your own readings. 5.1 Viewing the Readings Stored in Memory 1. Select your user ID. 2. Press the button. Alternating date/time display 1 sec. Memory number Past reading Latest reading After a memory number appears, the latest reading will appear with a past reading.
5. Using Memory Function 5.2 Viewing the Average Reading Your monitor can calculate an average reading based on the latest 2 or 3 readings taken within the most recent 10 minute timeframe. 1. Select your user ID. 2. When your monitor is off, press and hold the button for more than 3 seconds. The “ ” symbol flashes. Note • If there are only 2 readings in the memory for that period, the average will be based on these 2 readings. • If the average reading is high (refer to sub-section 1.
6. Other Settings 6.1 Deleting All Readings Stored in Memory 1. Select your user ID. 2. Press the button. After a memory number appears, the latest reading will appear with a past reading. 3. While holding the button down, press and hold the [START/STOP] button for more than 3 seconds. The display to the right flashes. 4. While holding the button down, press and hold the [START/STOP] button for more than 3 seconds. The display stops flashing, then all readings are deleted.
6. Other Settings 5. Press [START/STOP] button to turn your monitor off. Note • All readings will be deleted. You cannot partially delete the readings stored in the memory. • Your monitor automatically turns off after 2 minutes. 6.2 Restoring Your Monitor to the Default Settings To delete all the information stored in your monitor, follow the instructions below. Make sure that your monitor is turned off. 1.
6. Other Settings 3. Press [START/STOP] button to turn your monitor off. Note • Reverting to the default setting of your monitor does not delete the information in the app. • Your monitor automatically turns off after 3 minutes. • You will need to re-pair your monitor when using the monitor again. Without re-pairing, your readings will not be transferred to your app.
7. Error Messages and Troubleshooting If any of the below problems occur during measurement, check to make sure that no other electrical device is within 12 inches (30 cm). If the problem persists, refer to the table below. Display/ Problem Possible Cause Solution Insert the air plug Air plug is not completely plugged into securely. Refer to subsection 2.4. the monitor. The arm cuff is not appears or the applied correctly. arm cuff does not inflate. Air is leaking from the arm cuff.
7. Error Messages and Troubleshooting Display/ Problem appears appears Possible Cause Do not touch the arm cuff and/or bend the air tube The arm cuff is inflated while taking a exceeding the maximum measurement. If inflating the arm cuff allowable pressure. manually, refer to the end of sub-section 3.1. You move or talk during Remain still and do not a measurement. talk during a Vibrations disrupt a measurement. measurement. The pulse rate is not detected correctly.
7. Error Messages and Troubleshooting Display/ Problem appears flashes flashes flashes appears flashes Possible Cause The monitor cannot connect to a smart device or transmit data correctly. Solution Follow the instructions shown in the “OMRON connect US/CAN” app. If the “Err” symbol still appears after checking the app, contact customer service. Refer to section 12. Refer to sub-section 4.
7. Error Messages and Troubleshooting Display/ Problem Possible Cause appears or the monitor is Batteries are depleted. turned off unexpectedly during a measurement. Solution Immediately replace all 4 batteries with new ones. Refer to sub-section 2.1. Batteries are depleted. Nothing Check the battery appears on the display of the Battery polarities are not installation for proper placement. Refer to subproperly aligned. monitor. section 2.1. Blood pressure varies constantly.
8. Maintenance 8.1 Maintenance To protect your monitor from damage, follow the directions below: •Changes or modifications not approved by the manufacturer will void the user warranty. Caution • DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading. 8.2 Storage Store your monitor and other components in a clean, safe location. 1. Remove the arm cuff from the monitor.
8. Maintenance 8.3 Cleaning •Do not use any abrasive or volatile cleaners. •Use a soft dry cloth or a soft cloth moistened with mild (neutral) detergent to clean your monitor and arm cuff and then wipe them with a dry cloth. •Do not wash or immerse your monitor and arm cuff or other components in water. •Do not use gasoline, thinners or similar solvents to clean your monitor and arm cuff or other components. 8.
9. Optional Accessories Easy Wrap ComFit Cuff Arm circumference 9" - 17" (22 - 42 cm) Model: HEM-FL31-B AC Adapter HEM-FL31-BZ Model: HEM-ADPTW5 Protective LCD Cover For further information, visit our website at OmronHealthcare.com Model: HEM-CACO-734 Note •Do not throw the air plug away. The air plug can be applicable to the optional cuff.
9. Optional Accessories Using the AC Adapter (optional accessory) Note •Make sure not to place your monitor in a location where it is difficult to plug and unplug the AC adapter. •We recommend keeping batteries in your monitor at all times, even if you choose to use the AC adapter. If only the AC adapter is used without keeping batteries in your monitor, you may need to reset the date and time each time you unplug and plug back the AC adapter. The readings will not be deleted. 1.
10. Specifications Model Display Cuff pressure range Blood pressure measurement range Pulse measurement range BP7350 HEM-7340T-Z LCD digital display 0 to 299 mmHg SYS: 60 to 260 mmHg DIA: 40 to 215 mmHg 40 to 180 beats / min. Pressure: ±3 mmHg Pulse: ±5% of display reading Inflation Automatic by electric pump Deflation Automatic pressure release valve Measurement method Oscillometric method Accuracy Transmission method Bluetooth® low energy technology Frequency range: 2.4 GHz (2400 - 2483.
10. Specifications Weight Dimensions Cuff circumference applicable to the monitor Memory Contents Protection against electric shock Applied part Monitor: Approximately 16.2 oz. (460 g) not including batteries Arm cuff: Approximately 5.8 oz. (163 g) Monitor: Approximately 7.5" (w) × 3.1" (h) × 4.1" (l) (191 mm × 79 mm × 105 mm) Arm cuff: Approximately 5.7" × 20.9" (air tube: 29.
10. Specifications About a wireless communication interference This product operates in an unlicensed ISM band at 2.4 GHz. In the event this product is used near other wireless devices such as microwave and wireless LAN, which operate on the same frequency band as this product, there is a possibility that interference may occur. If interference occurs, stop the operation of the other devices or relocate this product away from other wireless devices before attempting to use it.
10.
11. FCC Statement FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
12. Limited Warranty Your BP7350 Blood Pressure Monitor, excluding the arm cuff and batteries is warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The arm cuff is warranted to be free from defects in materials and workmanship appearing within 1 year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor.
12. Limited Warranty FOR CUSTOMER SERVICE Visit our web site at: Call toll free: OmronHealthcare.
13. Guidance and Manufacturer’s Declaration OMRON Battery-operated / AC Adapter-operated Blood Pressure Monitor Information for Accompanying Documents in the Scope of IEC60601-1-2:2014 Important information regarding Electromagnetic Compatibility (EMC) BP7350 manufactured by OMRON HEALTHCARE Co., Ltd. conforms to IEC60601-12:2014 Electromagnetic Compatibility (EMC) standard. Further documentation in accordance with this EMC standard is available at OmronHealthcare.com/emc.
OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN Distributed by: OMRON HEALTHCARE, INC. 1925 West Field Court Lake Forest, IL 60045 U.S.A. OmronHealthcare.com © 2019 OMRON HEALTHCARE, INC.