R INSTRUCTION MANUAL Automatic Blood Pressure Monitor and Print-Out with IntelliSense™ Model HEM-705CP
TITLE TABLE OF CONTENTS Introduction...................................................................................................4 Notes on Safety.............................................................................................5 Know Your Unit ............................................................................................6 Quick Reference Guide ................................................................................7 Initial Set-Up .....................................
INTRODUCTION TITLE Thank you for purchasing the Omron Automatic Blood Pressure Monitor and Print-Out with IntelliSense™, HEM-705CP. Although your blood pressure fluctuates throughout the day, no self-adjustment is necessary as the monitor intelligently adjusts to your changing conditions. The IntelliSense™ monitor uses fuzzy logic intelligence to sense both your systolic and diastolic blood pressure values, and the result is personalized inflation with every reading.
NOTESTITLE ON SAFETY • The warning signs and the sample icons shown here are listed for you to use the product safely and correctly as well as to prevent the risk and the damage to you and others from happening. • The icons and meanings are as follow. Warning sign Caution Contents Indicates matters in which bodily harm or material damage* may arise as a result of incorrect handling. * Material damage refers to a wide range of damage involving your house, household goods, domestic animals, and pets.
KNOWTITLE YOUR UNIT Main Unit: Heart Symbol AC Adapter Jack Adjust Date/Time Set Date/Time Systolic Blood Pressure Printer Jack Cuff Jack ON/OFF Button Diastolic Blood Pressure START Button Pulse Rate Memory Call Button Date/Time Display Printer-Unit: Printer Cover Printer Cable Plug Power/Operation Light Paper Feed Print Current Data/Stop Printing Print Graph Print All Data Accessories: Arm Cuff (Applicable arm cuff for Adults 9 to 13 in.
TITLE QUICK REFERENCE GUIDE 1. Avoid eating, smoking, and exercising for at least 30 minutes before taking a measurement. 2. Remove tight-fitting clothing from your upper arm. 3. Sit in a chair with your feet flat on the floor and place your arm on a table so that the cuff is at the same level as your heart. 4. Put your arm through the cuff loop making sure that the bottom edge of the cuff is approximately one-half inch above the elbow and that the Green Marker on the cuff is above brachial artery.
BATTERY INSTALLATION/REPLACEMENT Main Unit: 1. Slide the battery cover off in the direction of the arrow. 2. Install or replace 4 "AA" size batteries so the + (positive) and – (negative) polarities match the polarities of the battery compartment as indicated. 3. Replace the battery cover. Notes: If the Low Battery Indicator ( ) appears on the display, replace all four batteries. Long-life alkaline batteries are recommended.
HOW TO SET TIME AND DATE 1. When the batteries are installed, the display will show 12:00a.m. 2. Press and hold the “SET” button. The year digits (2001) will flash. 3. Press the “ADJUST” button to advance the digit(s) one at a time. If you hold down the “ADJUST” button, the digits will advance rapidly. 4. Press “SET” button when desired number is on the display to lock setting. 5. Repeat Step 3-4 for month and date. 6. Repeat Step 3-4 for hour and minutes.
HOW TO CONNECT PRINTER UNIT 1. Remove printer cable plug from around printer unit. 2. Connect printer cable plug cord to the main unit. Arrow ( ) on printer cable plug should face down to main unit.
HOW TO LOAD PRINTER PAPER 1. Open printer cover in the direction indicated by the arrow symbol and remove paper roll from the main unit. 2. Peel of the glued edge of the printing paper included with the unit, and cut the first 4 inches with a pair of scissors. 3. Press the PAPER FEED button while inserting the edge of the printing paper into the paper feeding slot. When the edge of the paper protrudes sufficiently, release the PAPER FEED button. 4. Close the plastic printer cover.
HOW TO USE THE OPTIONAL AC ADAPTER To avoid potential damage to the monitor, use only the exclusive AC adapter (Model HEM-ADPT1). To connect the AC adapter 1. Insert the AC adapter plug into the AC Adapter jack on the back of the main unit. 2. Plug in the AC adapter into a 120V AC outlet (50 - 60 cycles). To remove the AC adapter To remove the AC adapter, disconnect the adapter plug from the AC outlet first and then disconnect the cord from the monitor’s jack.
TITLE HOW TO APPLY THE ARM CUFF AU ODEL H M 1. If the cuff is assembled correctly, the hook material will be on the outside of the cuff loop and the metal D-ring will not touch your skin. 2. If the cuff is not assembled, pass the end of the cuff furthest from the tubing through the metal D-ring to form a loop. The smooth cloth should be on the inside of the cuff loop. IC TOL SYS LIC STO DIA Pile Side Material Sewn Hook Material 3.
TITLE HOW TO APPLY THE ARM CUFF 5. Pull the cuff so that the top and bottom edges are tightened evenly around your arm. 6. When the cuff is positioned correctly, press the sewn hook material FIRMLY against the pile side of the cuff. 7. Make certain the cuff fits snugly around your arm. The cuff should make good contact with your skin. You should be able to fit your index finger between the cuff and your arm easily, so you can pull the cuff off and on. 8. Relax your arm and turn your palm upward. 9.
TITLEA READING HOW TO TAKE 1. Press the ON/OFF button. a) All display symbols appear for approximately one second. b) When the monitor becomes ready to measure, the Heart Symbol appears on the display. 2. Press the START button and remain still. As the cuff begins to inflate, the monitor automatically determines your ideal inflation level. Because this monitor detects the pulse even during inflation, do not move your arm and remain still until the entire measurement completes. 3.
TITLEA READING HOW TO TAKE Instructions for special conditions: • If your systolic pressure is known to be more than 220 mmHg, push and hold the START button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure. NOTE: Do not apply more pressure than necessary. The monitor will not inflate above 300 mmHg. • If you want to stop the measurement, push the ON/OFF button. The monitor will stop inflating and start deflating rapidly, then the monitor will turn off.
TITLE HOW TO USE THE MEMORY FUNCTION This monitor has a memory capable of storing twenty-eight (28) sets of readings. Every time you complete the measurement, the monitor automatically stores blood pressure and pulse rate. • Push the Symbol ( button to recall stored readings while the Heart ) is displayed. • When twenty-eight (28) sets of readings are stored in memory, the oldest set will be deleted to store a new set. • Date and Time of stored readings will be alternately displayed.
HOW TO USE THE PRINTER-UNIT The HEM-705CP has the ability to print all the readings stored in memory. NOTE: • Make sure printer-unit is connected to main unit (see How to Connect Printer-Unit) • Do not use printer-unit without loading printer paper (see How to Load Printer Paper). Improper use will decrease operation life cycle of printer-unit. • Main unit does NOT have to be ON to operate printer-unit. • Printer-unit will automatically shut OFF when printing is completed.
HOW TO USE THE PRINTER-UNIT The numerical print-out will list all stored readings (maximum of 28 readings) starting with the most recent reading to last stored reading: Systolic Reading Column Month(s) Date Time/Hour Diastolic Reading Column 3-6 mmHg D. T. 20 15 23 11 1 15 3 14 5 7 15 7 20 12 3 12 10 12 12 14 14 7 21 15 22 12 4 6 *AV.
HOW TO USE THE PRINTER-UNIT 3. Press “GRAPH” button to print a graph of all stored readings (stores a maximum of 28 readings). Month(s) 3-6 50 100 mmHg 150 200 Last reading stored If reading is lower than 50 mmHg or higher than 200 mmHg, ( ) will appear.
TITLE TROUBLESHOOTING Main Unit: Error Indicator Cause Cuff under-inflated. Movement during measurement. Correction Turn monitor off, wait 2-3 minutes and take another measurement. Repeat steps listed under "How to Take a Reading." If "EE" is displayed again, take another measurement by pressing and holding the START button as shown on instruction page 12 under "How to Take a Reading", instructions for special conditions. Your cuff maybe wrapped too tightly around your arm.
TITLE TROUBLESHOOTING Printer Unit: Error Indicator Nothing is printed Power/Operation light is blinking, stopped printing No printer paper comes out Cause Correction Printer cable plug is not connected properly or damaged Connect correctly reference p.
TITLE CAUTION CAUTION: Changes or modifications not expressly approved by Omron Healthcare, Inc. could void the user’s warranty. NOTE: POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
CARE AND MAINTENANCE To protect your monitor from damage, please AVOID the following: • Subjecting your monitor and cuff to extreme temperatures, humidity, moisture, and direct sunlight. • Folding the cuff and tubing tightly. • Inflating the monitor over 280 mmHg. • Disassembling the monitor. • Subjecting the monitor to strong shocks (for example, dropping the monitor on the floor.) • Do not clean the monitor with volatile liquids. THE MONITOR SHOULD BE CLEANED WITH A SOFT, DRY CLOTH. • Do not clean cuff.
WARRANTY Your HEM-705CP IntelliSense™ Automatic Blood Pressure Monitor is warranted to be free from manufacturing defects for a period of five years under normal use. The five year warranty excludes the monitor cuff. The cuff is warranted for a one year period. This warranty extends only to the original retail purchaser. Should repair be needed within the warranty period, ship the unit prepaid with the proof of purchase date and $5.00 for return shipping and insurance to: Omron Healthcare, Inc.
SPECIFICATIONS MAIN UNIT: Model: HEM-705CP Display: LCD Digital Display Measurement Range: Pressure: 0 to 299 mmHg Pulse: 40 to 180/minute Accuracy/Calibration: Pressure: ±4 mmHg Pulse: ±5% of reading Inflation: Automatic by electric pump Deflation: Active electronic control valve Rapid Pressure Release: Active electronic control valve Pressure Detection: Capacitive pressure sensor Measurement Method: Oscillometric method Pulse Wave Detection: Capacitive pressure sensor Power Source: 4
SPECIFICATIONS PRINTER UNIT: Power Source: 4 “AA” batteries Battery Life: Approximately 300 uses (current data) Paper Usage: Approximately 200 uses for 1 roll (current data) (Printer) Weight: Approximately 6 oz. (170 g) (not including batteries) (Printer) Dimensions: Approximately 4 4⁄5'' (l) x 2 4⁄5'' (w) x 2'' (h) (123 mm x 72 mm x 50 mm) NOTE: These specifications are subject to change without notice.
For Customer Service Call Toll Free 1-800-634-4350 Manufactured by OMRON Distributed by: OMRON HEALTHCARE, INC. 300 Lakeview Parkway Vernon Hills, Illinois 60061 www.omronhealthcare.com Copyright © 2001 OMRON HEALTHCARE, INC.