HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 1 HEM-907-Z2_E_M03_090828.pdf INSTRUCTION MANUAL IntelliSense® Blood Pressure Monitor Model HEM-907XL Table of Contents Be Sure to Read This Section XL SYS ® DIA mmH g Exemptions.........................................2 Notes on Safety..................................3 Know Your Unit PUL SE mm Hg ON/O BPM AC FF HID E STA RT P-SE T 140 100 180 MO D 220 AU TO E AVG SINGL E . Features of the Product .....................
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 2 Be Sure to Read This Section EXEMPTIONS OMRON does not accept liability and warranty becomes void under the following circumstances: 1. When persons, not authorized by OMRON, perform repairs or modifications of this product. 2. When use and/or operation of this device is adversely effected by a device not manufactured by OMRON. 3.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 3 NOTES ON SAFETY • The warning signs and the sample icons shown here are listed to insure safe and accurate use. • The icons and meanings are as follows. Warning sign Contents Indicates matters in which death or severe bodily damage may arise as a result of incorrect handling. Indicates matters in which bodily harm or material damage* may arise as a result of incorrect handling.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 4 NOTES ON SAFETY Unplug the AC adapter from the electric outlet if this unit is unused for an extended period of time. Unplug the AC adapter from the electric outlet when installing, removing, or cleaning the unit. Confirm readings with a stethoscope when an irregular pulse wave is displayed or when the measured value is questionable or erratic. Use an AC adapter indicated for use with a power supply of 110 VAC.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 5 NOTES ON SAFETY Do not use a cellular phone near this unit. Do not use this unit in a vehicle. Do not install the parts and/or instruments not specified for this unit. Do not use a broken power cord or AC adapter. Do not install or store this unit where it may come in contact with water or liquid medication. This is a Class II device with double insulation. General advice Do not place or put anything on this unit. Do not drop this unit.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 6 NOTES ON SAFETY Maintenance and inspection 1. Check the unit operation on regular basis. 2. If this unit has not been used for more than three months, be sure to check that this unit operates normally and safely before use. Troubleshooting If device error 9 (Er9) occurs, take the following procedure promptly: (1) Remove the cuff from the patient’s arm. (2) Turn off the power of the unit and unplug the AC adapter from the electric outlet.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 7 Know Your Unit FEATURES OF THE PRODUCT OMRON IntelliSense® Blood Pressure unit, Model HEM-907XL is developed to measure blood pressure and pulse rate accurately and simply in a doctor’s office, examination room, or patient bedside. • One-button operation Simply wrap the cuff and push the START Button. Blood pressure and pulse rate are automatically measured by the oscillometric method.
HEM-907-Z2_E_M.qxd 09.8.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 9 FEATURES AND FUNCTIONS Names of the Parts Main unit 1. Display (See Page 10 for detail) XL SYS ® 2. HIDE (non-display) Button 6. ON/OFF (power) Button mm Hg DIA 7. START Button PUL SE mm Hg BPM DC 8V ON/O FF AC HID E 3. DC jack STA RT P-SE T 140 100 180 MO 220 4. P-SET (pressure setting) Knob AU TO D E AVG SINGL E . 260 280 8. DEFLATION (deflation control) /Measurement Result Display Switch Button 9. Air Connector MA NU .
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 10 FEATURES AND FUNCTIONS Display Systolic blood pressure Ready to Measure / Pulse synchronization Diastolic blood pressure Pulse level / Number of irregular pulses Number of irregular pulse waves: The pulse wave having a difference of more than 25% from the average interval of pulse waves is determined irregular. Number of irregular pulse waves is displayed up to three times. SYS mmHg Example of display DIA 1 No.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 11 FEATURES AND FUNCTIONS Functions setting (1) Inflation level setting P-SET AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected. The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure (approximately 30-40 mmHg above the patient’s systolic pressure). Manual level setting: Inflation level can be set manually between 100 and 280 mmHg.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 12 FEATURES AND FUNCTIONS List of Functions for each Mode Measurement Mode Single (SINGLE) Mode Function Average (AVG) Mode Auscultation (MANU) Mode Check (CHECK) Mode Inflation level setting Non-display function Manual deflation control (4) AVG Function setting You can set the number of measurements, the waiting time until the 1st measurement, and the measurement interval for the AVG Mode.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 13 Preparations before Measurement HOW TO APPLY THE CUFF The cuff of OMRON IntelliSense® Blood Pressure Monitor HEM-907XL plays an important role of collecting the information on blood vessels. Please wrap the cuff according to the procedure, below. Do not wrap the cuff over an arm to which intravenous injection or transfusion is being conducted, or when otherwise contraindicated.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 14 HOW TO APPLY THE CUFF 3. Place the right or left hand of the patient with the palm of hand facing upward. 4. Align the Artery Position Mark with the brachial artery. 5. Wrap the cuff snugly using both hands and securely fasten it with the Velcro™ tape. At this time, the lower edge of the cuff must be placed 1/2" to 1" above the inner side of elbow joint.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 15 HOW TO USE THE POWER SOURCE - DANGER TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK, CAREFULLY FOLLOW THESE INSTRUCTIONS. For connection to a power supply not in the U.S.A., use an attachment plug adapter of the proper configuration for the power outlet. The power unit (AC Adapter) is intended to be correctly oriented in a vertical or floor mount position. NOTE: Use only the authorized AC Adapter that came with this monitor.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 16 HOW TO USE THE POWER SOURCE How to use the AC adapter Be sure to use the AC adapter from the power supply of 110 VAC. Do not use this unit in the presence of flammable gas, or anesthetics, or in a high pressure oxygen room or oxygen tent. Do not install or store this unit where it may come in contact with water or liquid medication. Do not touch the AC adapter with wet hands. This is a Class II device with double isolation.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 17 HOW TO USE THE POWER SOURCE Installation and Replacement of Battery Pack If battery fluid gets into your eye or comes in contact with skin, wash the affected area with water repeatedly. Immediately consult a doctor for treatment. Do not use the battery pack for devices other than for this unit. Do not disassemble the battery pack. 1. Remove both screws on the upper portion of the battery cover of this unit, and remove the cover. 2.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 18 How to Measure Blood Pressure HOW TO MEASURE BLOOD PRESSURE Confirm readings with the stethoscope when an irregular pulse wave is displayed or when the measured value is questionable or erratic. If cuff inflation does not stop, remove the cuff or pull out the air tube from the main unit. General advice Do not use a cellular phone near this unit. • Patients should not touch the unit. Do not use this unit in a vehicle.
HEM-907-Z2_E_M.qxd E 09.8.28 11:45 AM Page 19 HOW TO MEASURE BLOOD PRESSURE (IN SINGLE MODE) 1. Push the ON/OFF (power) Button to turn on the power. 2. Set the MODE Selector to “SINGLE”. 3. Set the P-SET (inflation level) Knob to “AUTO” or to the target pressure value. 4. Measure the patient’s arm size, and wrap appropriate cuff over the patient’s arm. (Refer to Pages 13 and 14.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 20 HOW TO MEASURE BLOOD PRESSURE (IN SINGLE MODE) 5. Push the START Button to start the measurement. • Do not push the START Button without wrapping the cuff. • If you want to stop measurement, push the STOP Button. The cuff will rapidly deflate. 6. The measurement results are displayed. • While the battery pack is in use, the monitor will turn off automatically after five minutes of inactivity and the display (measurement results) will disappear.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 21 HOW TO MEASURE BLOOD PRESSURE (IN AVERAGE MODE) 1. Push the ON/OFF (power) Button to turn on the power. 2. Set the MODE Selector to “AVG”. The factory-set values are set as follows: • Number of measurements: 2 • Waiting time until the 1st measurement: 0 sec. • Interval: 1 min. To change these factory-set values, refer to Page 12. 3. Set the P-SET (inflation level setting) Knob to “AUTO” or the target pressure value. 4.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 22 HOW TO MEASURE BLOOD PRESSURE (IN AVERAGE MODE) 5. Push the START Button to start the measurement. After the pre-select waiting time, the unit takes the 1st measurement. After displaying the results of 1st measurement, subsequent measures occur automatically at the specified intervals. • For setting the number of measurements, the waiting time until the 1st measurement, and the interval, refer to Page 12.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 23 HOW TO MEASURE BLOOD PRESSURE (IN MANUAL MODE) 1. Push the ON/OFF (power) Button to turn on the power. 2. Set the MODE Selector to “MANU”. 3. Set the P-SET (inflation level setting) Knob to “AUTO” or the target pressure value. 4. Measure the patient’s arm size and wrap appropriate cuff over the patient’s arm. (Refer to pages 13 and 14.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 24 HOW TO MEASURE BLOOD PRESSURE (IN MANUAL MODE) 5. Place the stethoscope on the patient’s arm. 6. Push the START Button to start the measurement. • Do not push the START Button without wrapping the cuff. • Do not squeeze or press the cuff during the measurement. • If you want to inflate again after the start of deflation, push the START Button.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 25 HOW TO CHECK PRESSURE ACCURACY (IN CHECK MODE) Accuracy of pressure display can be checked in the CHECK Mode. What you need to prepare (1) Calibrated mercury manometer (including inflation bulb), (2) T-tube, (3) two air tubes, and (4) a sturdy cylindrical shaped object on which the cuff is wrapped. How to check ® 1. Connect the manometer, inflation bulb, cuff, and the monitor with the T-tube as shown in the figure on the right. 2.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 26 HOW TO CLEAN THE UNIT AFTER USE General advice When cleaning this unit, unplug the AC adapter from the electric outlet. • Do not clean this unit with gasoline, paint thinner, or high concentration alcohol. After cleaning this unit, dry it well, before plugging the AC adapter in the electric outlet. • Do not disinfect this unit by autoclave or gas sterilization (EtO, gluteraldehyde, or high concentration ozone.) HOW TO CLEAN THE UNIT AFTER USE 1.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 27 LIST OF ERROR CODES Error code Explanation How to correct Inflation error • When the pressure does not exceed 12 mmHg within the set time after the start of inflation • When the inflation does not reach the set cuff pressure within the specified time after the start of inflation • Confirm that the air tube connecting the cuff and the main unit is connected securely.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 28 TROUBLESHOOTING If the unit malfunctions during use, please check the following: Trouble The unit inflates to abnormally high (low) pressure. The monitor cannot measure blood pressure. What to inspect How to correct Is the cuff wrapped correctly? Wrap the cuff correctly, and measure again. (Refer to Page 13 and 14.) Is the patient moving during inflation? Ask the patient not to move during measurement, and measure again.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 29 Specifications NOTES CAUTION ON SAFETY CAUTION: Changes or modifications not expressly approved by Omron Healthcare, Inc. could void the user’s authority to operate this product. NOTE: POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules.
HEM-907-Z2_E_M.qxd 09.8.
HEM-907-Z2_E_M.qxd 09.8.
HEM-907-Z2_E_M.qxd 09.8.28 11:45 AM Page 32 For Customer Service Call Toll Free 1-877-216-1336 Made in Japan Manufactured by OMRON HEALTHCARE Distributed by: OMRON HEALTHCARE, INC. 1200 Lakeside Drive Bannockburn, Illinois 60015 www.omronhealthcare.com Copyright © 2009 OMRON HEALTHCARE, INC.