Instruction Manual M24/7 Ambulatory Blood Pressure Monitor EN www.omron-healthcare.
Introduction Thank you for purchasing the OMRON M24/7 ambulatory blood pressure monitor. The OMRON M24/7 is a silent, compact and very reliable ambulatory blood pressure monitor. A large LCD screen insures easy readability and its lightweight design provides for maximum patient comfort. The OMRON M24/7 operates with two 1.5V AA batteries (included in the package) or two 1.2V AA rechargeable batteries.
Content Introduction........................................................................................................ 2 Important Safety Information ........................................................................... 4 Recommended use of ambulatory blood pressure monitoring .................... 6 1. Overview ..................................................................................................... 7 1.1. 1.2. 1.3. 1.4. Main Unit...............................................
Important Safety Information Important Safety Information This symbol on the Omron M24/7 monitor is a warning that you should read the accompanying documentation (this manual). Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Caution: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. (General Usage) No user serviceable parts inside. The OMRON M24/7 monitor contains high complexity electronic and ne mechanical components. If you have any problems, please refer your monitor to quali ed service personnel.
Recommended use of ambulatory blood pressure monitoring The European Society of Hypertension has published recommendations for ambulatory blood pressure measurement (O’Brien, E. et al. on Behalf of the European Society of Hypertension Working Group on Blood Pressure Monitoring, European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement, Journal of Hypertension 2003, 21: 821-848).
1. Overview 1.1. Main Unit Front side Back side LCD START button EVENT button DAY/NIGHT button Interface connection CUFF connection CUFF connection 1.2. Arm Cuffs The set contains 2 adult cuffs: normal size (24-32cm) and large size (32-42cm) A. Arm Cuff B. Air Plug C.
1.3. Display Time Time display display Heart symbol Heart symbol for for pulse indication pulse indication Night mode: mode: moon moon sign is islit lit Crossed battery Crossed battery symbol warns symbol warns about low battery about Communication Communication PC with PC Batteryvoltage voltage Battery display display For details of different display options see p. 11 1.4.
2. Operating Instructions 2.1. Working with the OMRON M24/7 Unit The OMRON M24/7 device is a compact, lightweight monitoring unit typically worn by the patient for 24 hours. The OMRON M24/7 operates with two AA size batteries. The memory is capable of storing more than 600 measurements. Stepwise pressure de ation ensures quality measurements, even if disturbing environmental factors occur. You may nd the battery compartment and the rating label on the backside housing.
Switching the device off Press and hold the START button for more than 10 seconds then release it when two horizontal segments appear on the LCD, this way the OMRON M24/7 will be switched off. If you do not release the button in 2 seconds after the two horizontal segments appeared, the monitor will return to normal operation. This feature helps to avoid unintended power-off. The device can only be switched on manually.
2.3. Overview Display The OMRON M24/7 shows important status information, the processes and results of individual readings on its LCD. The most important displays are listed here. In addition to these, a lot of extraordinary situations and errors have their own code displayed on the LCD. These codes are stored together with recorded data and listed in the OMRON BP Tracker Software. This helps service personnel in case of support issues. Normal status: time is displayed.
The monitor must be programmed with the OMRON BP Tracker Software installed on the computer (for details see Software Manual). Once the preprogrammed time is reached, the monitor will start operating automatically and perform blood pressure measurements based on the monitoring plan. To obtain reliable BP readings, certain rules must be observed. 2.4. Rules of Monitoring 1. Inform the patient about the goal and expected results of the monitoring. Provide an event/patient diary and rules to observe.
A successful monitoring session consists of the following steps: 1. Connecting the OMRON M24/7 monitor to the PC (First connection only – not required to repeat with each session) 1. Connect the optical interface, which was delivered with the device, to the USB port of the PC. 2. The other side of the optical USB interface cable, which is a two-point plastic connection, should be connected to the socket of the OMRON M24/7, in a way that the red plastic ring is directed to the lower side of the device. 3.
2.6. Manual Programming The OMRON M24/7 can be programmed with using its buttons, without using a PC. Programming Options There are three different measurement plans which may be selected in case of manual programming. These plans are stored in the device’s inside memory and they cannot be changed. The rst plan which is also the default one in the OMRON BP Tracker Software, measurements every 15 minutes during daytime and 30 minutes during night.
Software The OMROM M24/7 programmed manually can be used with the OMRON BP Tracker Software version 1.14.6 or newer. Manual programming LCD displays: Normal status: time is displayed 10 second delay state: time is displayed but with no double-dot 1. measurement plan: 15/30 minute day/night intervals 2. measurement plan: 20/40 minute day/night intervals 3. measurement plan: 30/30 minute day/night intervals Programming in progress: blinking signal 2.7.
In order to change batteries, take the monitor out of the holder carry pouch and remove the battery compartment cover on the back-side. Place two new, longlife AA alkaline batteries or two properly charged, high capacity AA rechargeable batteries into the compartment as shown in the polarity drawing. Close the compartment. 2.8. Cuffs and their application The OMRON M24/7 comes with 2 cuffs, normal and large adult size (see below). The bladders are made of neoprene and are latex-free.
Note: It is recommended that the cuff be applied as tightly as acceptable for the patient. A too loose cuff will cause much longer blood pressure measurement times and possibly aborted measurements. With an overly loose cuff, the monitor must pump to tighten the cuff on the arm and then it must reach the pressure necessary for measurement. This causes considerable inconvenience for the patient and results in less data for evaluation.
E 3 battery rundown E 4 batteries changed E 8 pressure limit exceeded E 9 temporary disturbance the AA batteries exhausted during measurement the AA batteries were replaced during the measurement – not shown on LCD the pressure in the pneumatic system exceeded the preset pressure limit external electric signals (e.g.
3.3. Maintenance and Storage Conditions Temperature: -20 - 50 °C Humidity: 10 - 95 %, non condensing Regular checks, warranty, service Veri cation of pressure measurement accuracy is recommended biannually. OMRON M24/7 monitors are covered by a two-year warranty. This warranty does not cover any malfunction or defects arising from improper use, the use of inadequate accessories, accident, theft, or use of the device outside operating environmental speci cations or intended measurement range.
4. Technical Data 4.1. Technical parameters Power supply: 2 AA rechargeable NiCd or NiMH batteries or 2 AA alkaline batteries Display: liquid-crystal Data storage: internal solid state memory Data transmission: USB optical cable PC interface: USB interface Operating environment: Temperature: 10 - 45 °C Humidity: 10 - 95 %, non condensing Atmospheric pressure: 70 - 106 kPa Storage conditions: Temperature: -20 - 50 °C Humidity: 10 - 95 %, non condensing Size (H x W x D): 70 x 99 x 30 mm Weight: app.
4.2. Safety Concerns Electric shock hazard protection OMRON M24/7 monitors meet relevant shock hazard protection standards. OMRON M24/7 monitors operate with two 1.5V AA batteries or two 1.2V AA rechargeable batteries. This excludes all electric shock hazards, even in the unlikely case of multiple device errors. Use only standard long-life (alkaline) batteries, or standard NiCd or NiMH rechargeable batteries of the proper size. Do not use lithium batteries.
4.3. EMC Information Medical electrical equipment should be used with precautions according to EMC, and must be installed according to the EMC notices disclosed in this manual as mobile RF transceivers could adversely affect it. Directive and declaration of manufacturer – Electromagnetic Emission The OMRON M24/7 is suitable for use in the speci ed electromagnetic environment.
Directive and declaration of manufacturer – Electromagnetic immunity OMRON M24/7 is suitable for use in the speci ed electromagnetic environment. The purchaser or user of OMRON M24/7 should assure that it is used in an electromagnetic environment as described below.
Directive and declaration of manufacturer – Electromagnetic immunity OMRON M24/7 is suitable for use in the speci ed electromagnetic environment. The purchaser or user of OMRON M24/7 should assure that it is used in an electromagnetic environment as described below.
Recommended separation distance OMRON M24/7 is intended to be used in electromagnetic environment with controlled RF disturbances. The purchaser or user of M24/7may help to reduce electromagnetic disturbances by de ning the separation distance between the transportable or mobile RF telecommunication equipment (transmitters) and OMRON M24/7, depending on the highest output power of the telecommunication equipment.
AMBULATORY BLOOD PRESSURE MONITORING PATIENT DIARY Name: Date of birth: Monitoring date: 8:00-8:30 8:30-9:00 9:00-9:30 9:30-10:00 10:00-10:30 10:30-11:00 11:00-11:30 11:30-12:00 12:00-12:30 12:30-13:00 13:00-13:30 13:30-14:00 14:00-14:30 14:30-15:00 15:00-15:30 15:30-16:00 16:00-16:30 16:30-17:00 17:00-17:30 17:30-18:00 18:00-18:30 18:30-19:00 19:00-19:30 19:30-20:00 20:00-20:30 20:30-21:00 21:00-21:30 21:30-22:00 22:00-22:30 22:30-23:00 23:00-23:30 23:30-24:00 0:00-0:30 0:30-1:00 1:00-1:30 1:30-2:00 2:00-
4.4. Product Warranty Information (a) MONITOR WARRANTY. The main monitor unit will be free from defects in materials and workmanship under normal use and service for a period of two (2) years from the date of receipt. This warranty covers the monitor unit only. This warranty does not cover any accessories that might come with the monitor unit. (b) ACCESSORIES WARRANTY.
Remarks 2010. All rights reserved Made in Hungary Meditech Ltd.