HEM-7000-E(V)_main.book Page 2 Friday, December 11, 2009 10:03 AM Contents Introduction ..............................................................................3 Important Safety Information ..................................................4 1. Overview ..............................................................................5 2. Preparation ..........................................................................7 2.1 Installing/Replacing the Batteries.....................................7 2.
HEM-7000-E(V)_main.book Page 3 Friday, December 11, 2009 10:03 AM Introduction The OMRON M6 Comfort is a compact, fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation without the need of pressure pre-setting or re-inflation the devices uses its advanced “IntelliSense” technology.
HEM-7000-E(V)_main.book Page 4 Friday, December 11, 2009 10:03 AM Important Safety Information • Blood pressure measurement is not suitable in cases of serious arteriosclerosis (hardening of the arteries). • The pulse display is not suitable for monitoring the frequency of cardiac pacemakers. • Pregnant women should only measure their own blood pressure in consultation with their doctor, since the readings can be changed by pregnancy.
HEM-7000-E(V)_main.book Page 5 Friday, December 11, 2009 10:03 AM 1.Overview 1. Overview Monitor A E B D C A. B. C. D. F G Display E. Battery compartment O/I START button F. Air jack Memory ( ) buttons G.
HEM-7000-E(V)_main.book Page 6 Friday, December 11, 2009 10:03 AM 1.Overview Display H I J L O K Q M R N H. I. J. K. L. M. N. P Systolic blood pressure Diastolic blood pressure Memory symbol Average value symbol Movement error symbol Irregular heartbeat symbol Battery low symbol O. Heartbeat symbol 1. Flashes during measurement 2. If flashing after measurement complete indicates blood pressure out of recommended range P. Pulse display Q. Deflation symbol R.
HEM-7000-E(V)_main.book Page 7 Friday, December 11, 2009 10:03 AM 2.Preparation 2. Preparation 2.1 Installing/Replacing the Batteries 1. Remove the battery cover. 2. Insert four “AA” batteries as indicated in the battery compartment and then replace the battery cover. Notes: • If the battery low symbol ( ) appears on the display, replace all four batteries at the same time. - When the battery low symbol ( ) starts to blink, you will still be able to use the unit for a short while.
HEM-7000-E(V)_main.book Page 8 Friday, December 11, 2009 10:03 AM 2.Preparation 2.2 Setting the Date and Time Your blood pressure monitor automatically stores up to 90 measurements values in its memory and calculates an average reading based on the measurements from the last three readings taken within 10 minutes. • Set the monitor to the correct date and time before taking a measurement for the first time.
HEM-7000-E(V)_main.book Page 9 Friday, December 11, 2009 10:03 AM 2.Preparation 4. Repeat steps 2 and 3 to set the month and date (day). or or 5. Repeat steps 2 and 3 to set the hour and minutes for the time. or EN or 6. Press the O/I START button when you have finished setting the date and time.
HEM-7000-E(V)_main.book Page 10 Friday, December 11, 2009 10:03 AM 3.Using the Unit 3. Using the Unit 3.1 How to Sit Correctly When Taking a Measurement You can take a measurement on either your left or right arm. Notes: • Measurements should be taken in a quiet place and you should be in a relaxed, seated position. Make sure that the room is not too hot or cold. • Avoid eating, smoking, or exercising for at least 30 minutes before taking a measurement. • Do not move or talk during measurement.
HEM-7000-E(V)_main.book Page 11 Friday, December 11, 2009 10:03 AM 3.Using the Unit Notes: • Correct posture during measurement is necessary to get accurate results. • You should also try to measure your blood pressure at the same time each day. (Within 1 hour after waking up is recommended.
HEM-7000-E(V)_main.book Page 12 Friday, December 11, 2009 10:03 AM 3.Using the Unit 3.2 Applying the Arm Cuff You can wrap the cuff either on your right or left arm. • Remove tight-fitting clothing from your upper arm. • Do not place the cuff over thick clothes and do not roll up your sleeve if it is too tight. 1. Insert the air plug into the air jack (on the left side of the device). Air jack Air plug The cuff must be fully deflated when it is inserted into the air jack. 2.
HEM-7000-E(V)_main.book Page 13 Friday, December 11, 2009 10:03 AM 3.Using the Unit 1) Hold the grip on the cuff securely with your hand. Grip Fabric fastener 2) Turn the palm of your hand upward. 3. Turn the palm of your hand upward. Bottom edge 1 to 2cm Blue strip Air tube Apply the cuff to your upper arm so that the blue strip is centred on the middle of your inner arm and points down the inside of the arm.
HEM-7000-E(V)_main.book Page 14 Friday, December 11, 2009 10:03 AM 3.Using the Unit 4. When the cuff is positioned correctly, close the fabric fastener FIRMLY. Notes: • Make certain the cuff fits snugly around your arm. • The cuff should make good contact with your skin. You should be able to fit your index finger between the cuff and your arm easily, so you can pull the cuff off and on. • Make sure that there are no kinks in the air tubing.
HEM-7000-E(V)_main.book Page 15 Friday, December 11, 2009 10:03 AM 3.Using the Unit 3.3 Taking a Reading 1. Press the O/I START button and remain still, the cuff will start to inflate automatically. As the cuff begins to inflate, the monitor automatically determines your ideal inflation level. Remain still and do not talk until measurement is completed. Note: Do not inflate the cuff if it is not wrapped around your arm. 2. Inflation stops automatically and measurement is started.
HEM-7000-E(V)_main.book Page 16 Friday, December 11, 2009 10:03 AM 3.Using the Unit 4. Press the O/I START button to turn the monitor off. Note: If you forget to turn the monitor off, it will automatically shut itself off after five minutes. Important: • Wait 2-3 minutes before taking another blood pressure measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.
HEM-7000-E(V)_main.book Page 17 Friday, December 11, 2009 10:03 AM 3.Using the Unit heart- or circulation-disease) a lower limit of blood pressure might be necessary and that it is recommended to consult the doctor in such cases.
HEM-7000-E(V)_main.book Page 18 Friday, December 11, 2009 10:03 AM 3.Using the Unit 3.4 Instructions for Special Conditions If your systolic pressure is known to be more than 220 mmHg, after the cuff starts to inflate, press and hold the O/I START button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure. Important: • The monitor will not inflate above 300 mmHg. • Do not apply more pressure than necessary.
HEM-7000-E(V)_main.book Page 19 Friday, December 11, 2009 10:03 AM 3.Using the Unit 3.5 Using the Memory Function This monitor has a memory capable of storing 90 sets of readings. Every time you complete the measurement, the monitor automatically stores blood pressure and pulse rate with the date and time. The monitor also calculates an average reading based on the measurements from the last three readings taken within 10 minutes.
HEM-7000-E(V)_main.book Page 20 Friday, December 11, 2009 10:03 AM 3.Using the Unit To View Previous Readings Stored in Memory 1. Press the button, while the average reading is displayed, to view reading stored in memory from the most recent to the oldest. Alternating display 2.
HEM-7000-E(V)_main.book Page 21 Friday, December 11, 2009 10:03 AM 4.Quick Reference Guide 4. Quick Reference Guide To help ensure a reliable reading, avoid eating, smoking, or exercising for at least 30 minutes before taking a measurement. Note: Remove any tight-fitting clothing from your upper arm. 1. Sit on a chair with your feet flat on the floor and place your arm on a table so that the arm cuff will be at the same level as your heart. 2.
HEM-7000-E(V)_main.book Page 22 Friday, December 11, 2009 10:03 AM 5.Handling Errors and Problems 5. Handling Errors and Problems 5.1 Error Messages Error Display Cause Remedy Cuff is under inflated. Movement during measurement Carefully read and repeat the steps listed under section 3.3. Cuff over inflated. Blinks or appears continuously. 22 This icon indicates irregular or weak pulses are detected, but result can be considered reliable. Remove the arm cuff.
HEM-7000-E(V)_main.book Page 23 Friday, December 11, 2009 10:03 AM 5.Handling Errors and Problems 5.2 Troubleshooting Problem No display appears when the O/I START button is pressed. Cannot measure or readings are too high. The blood pressure is different each time. The reading is extremely low (or high). Cause Remedy The batteries are empty. Replace the batteries. Is the cuff wrapped firmly around the arm? Attach the arm cuff correctly.
HEM-7000-E(V)_main.book Page 24 Friday, December 11, 2009 10:03 AM 6.Maintenance and Storage 6. Maintenance and Storage To protect your monitor from damage, please observe the following: • Do not subject your monitor and cuff to extreme temperatures, humidity, moisture or direct sunlight. • Do not fold the cuff or tubing tightly. • Do not disassemble the monitor. • Do not carry out repairs of any kind yourself.
HEM-7000-E(V)_main.book Page 25 Friday, December 11, 2009 10:03 AM 7.Optional Parts 7. Optional Parts AC Adapter (Q-adapter) Arm Cuff Order No. 1098336-8 Order No. 9956685-4 Using the Optional AC Adapter Insert batteries into the battery compartment even when using the AC adapter. Notes: • Never unplug the power cord with wet hands. • Use only the original AC adapter designed for this unit. Use of unsupported adapters may damage the unit.
HEM-7000-E(V)_main.book Page 26 Friday, December 11, 2009 10:03 AM 8.Technical Data 8.
HEM-7000-E(V)_main.book Page 27 Friday, December 11, 2009 10:03 AM 8.Technical Data Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC.s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation.
HEM-7000-E(V)_main.book Page 28 Friday, December 11, 2009 10:03 AM 9.Some Useful Information about Blood Pressure 9. Some Useful Information about Blood Pressure What is Blood Pressure? Blood pressure is a measure of the force of blood flowing against the walls of the arteries. Arterial blood pressure is constantly changing during the course of the heart’s cycle. The highest pressure in the cycle is called the Systolic Blood Pressure; the lowest is the Diastolic Blood Pressure.
HEM-7000-E(V)_main.book Page 29 Friday, December 11, 2009 10:03 AM 9.
HEM-7000-E(V)_main.book Page 30 Friday, December 11, 2009 10:03 AM 9.Some Useful Information about Blood Pressure Manufacturer OMRON HEALTHCARE CO., LTD. 24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto, 615-0084 Japan EUrepresentative OMRON HEALTHCARE EUROPE BV Kruisweg 577, 2132 NA Hoofddorp, The Netherlands Production facility www.omron-healthcare.com EC REP OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD.