English Français Deutsch Italiano Español Automatic Blood Pressure Monitor Model M6 Comfort Instruction Manual Nederlands Ɋɭɫɫɤɢɣ Türkçe IM-HEM-7321-E-01-08/2013 2298869-1A
1 2 b J N c d a O
A B G K C D E F J I H L H G P V Q W X Y Z R M N S AA T AB H AC H O U AD AE AF
Check following components! Vérifier les composants suivants ! Prüfen Sie folgende Teile des Lieferumfangs! Controllare i componenti indicati di seguito! ¡Compruebe los siguientes componentes! English Français Automa tic Blood Pre ssure Mon Model itor M6 Com fort Instruc tion Ma nual Deutsch Italiano Españo l Nederla nds Ɋɭɫɫɤɢɣ Türkçe Controleer de volgende onderdelen! ɉɪɨɜɟɪɶɬɟ ɫɥɟɞɭɸɳɢɟ ɤɨɦɩɨɧɟɧɬɵ! $úD÷ÕGDNL ELOHúHQOHUL NRQWURO HGLQ
Contents Thank you for purchasing the OMRON M6 Comfort Automatic Blood Pressure Monitor. The OMRON M6 Comfort is a compact, fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation without the need of pressure pre-setting or re-inflation the device uses its advanced “IntelliSense” technology.
Important Safety Information Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) Consult your physician before using the device in pregnancy including pre-eclampsia, or if diagnosed with arrhythmia or arteriosclerosis. Do not use the device on the injured arm or the arm under medical treatment. Do not apply the arm cuff on the arm while being on an intravenous drip or blood transfusion.
Important Safety Information If there are any abnormalities during the measurement, remove the arm cuff. Do not use this device on infants or persons who cannot express their intentions. Do not inflate the arm cuff more than necessary. Do not use the device for any purpose other than measuring blood pressure. Use only the approved arm cuff for this device. Use of other arm cuffs may result in incorrect measurement results.
Important Safety Information (Battery Usage) Do not insert the batteries with their polarities incorrectly aligned. Use only 4 “AA” alkaline or manganese batteries with this device. Do not use other types of batteries. Do not use new and used batteries together. Remove the batteries if the device will not be used for three months or more. General Precautions • Do not forcibly crease the arm cuff or the air tube excessively. • Do not press the air tube while taking a measurement.
1. Know Your Device Open the rear cover page to read the following: The alphabet in the rear cover page correspond to those in the body page.
1. Know Your Device Display Symbols Irregular Heartbeat Symbol ( ) Normal Heartbeat When the monitor detects an irregular rhythm two or more times during the Pulse measurement, the irregular heartbeat symbol ( ) will appear on the Blood pressure display with the measurement values. An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure.
1. Know Your Device Blood Pressure Colour Indicator If your systolic or diastolic pressure is above the standard range (135 mmHg for the systolic blood pressure and/or 85 mmHg for the diastolic blood pressure), the blood pressure colour indicator will light in “orange” when the measurement result is displayed. If they are within the standard range, the blood pressure colour indicator will light in “green”. The JNC7* Guideline recommends the following guideline.
2. Preparation 2.1 Battery Installation 1. Remove the battery cover. 2. Insert 4 “AA” batteries as EN indicated in the battery compartment. 3. Replace the battery cover. Notes: • When the low battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same time. Long life alkaline batteries are recommended. • The measurement values continue to be stored in memory even after the batteries are replaced. • The supplied batteries may have a shorter life.
2. Preparation Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page. 2.2 Setting the Date and Time Set the monitor to the correct date and time before taking a measurement for the first time. 1. Press the H Date/Time setting ( ) button. The year flashes on the V Date/Time display. 2. Push or button to change the year. Push button to confirm the year and then the month flashes.
3. Using the Device Open both the front and rear covers to read the following: The letter identifiers on the cover pages correspond to those in the body of this page. 3.1 Applying the Arm Cuff Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm. Do not place the arm cuff over thick clothes. EN 1. Insert N the air plug into J the air jack securely. 2. Apply the arm cuff to your left upper arm. The bottom edge of the arm cuff should be a 1 to 2 cm above the elbow.
3. Using the Device 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement. • • • • 10 Sit on a chair with your feet flat on the floor. Sit upright with your back straight. --- c Sit with your back and arm being supported. The arm cuff should be placed on your arm at the same level as your heart.
3. Using the Device 3.3 Taking a Measurement Notes: • To cancel a measurement, press the START/STOP button to release the air in the arm cuff. • Remain still while taking a measurement. The monitor is designed to take measurements and store the measurement values in the memory for 2 people using USER ID 1 and USER ID 2. EN 1. Select your USER ID (1 or 2). 2. Press the START/STOP button. The arm cuff starts to inflate automatically.
3. Using the Device 3. Remove the arm cuff. 4. Press the START/STOP button to turn the monitor off. The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes. Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement. Using the Guest Mode The monitor stores measurement values for 2 users in the memory.
3. Using the Device 3.4 Using the Memory Function The monitor automatically stores the results up to 100 sets for each user (1 and 2). It can also calculate an average value based on the last 3 measurement values taken within 10 minutes. Notes: • If there are only 2 measurement values in the memory for that period, the average will be based on these 2 values. • If there is 1 measurement value in the memory for that period, this is displayed as the average.
3. Using the Device To View the Average Value 1. Select your USER ID (1 or 2). 2. Press and hold the button for more than 3 seconds. Notes: • If the previous measurement was taken without setting the date and time, the average value is not calculated. • If there are no measurement values stored in the memory, the screen to the right is displayed.
3. Using the Device To View the Weekly Average Value The monitor calculates and displays weekly averages for the measurements taken in the morning and evening within 8 weeks for each user. Note: The week begins Sunday at 2:00. 1. Select your USER ID (1 or 2). 2. Press the button. EN The morning average for “THIS WEEK” appears on the display. Note: The morning hypertension symbol ( ) appears if the morning weekly average is above 135/85 mmHg.
3. Using the Device Note: If there are no measurement values stored in the memory, the screen to the right is displayed. Morning Weekly Average This is the average for the measurements taken during the morning (4:00 - 11:59) between Sunday and Saturday. An average for each day is calculated for up to three measurements taken within 10 minutes of the first measurement of the morning.
3. Using the Device To Delete All the Values Stored in Memory The values stored in the memory are deleted by USER ID. 1. Select your USER ID (1 or 2). 2. Press the Memory button, while the memory symbol ( appears. ) EN 3. While holding the button down, press the START/STOP button for more than 3 seconds. Note: You cannot partially delete the values stored in the memory. All values for the user you select will be deleted.
4. Error Messages and Troubleshooting 4.1 Error Messages Display 18 Cause Solution Irregular heartbeats are detected. Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Repeat the steps in section 3.3. If this error continues to appear, contact your physician. Movement during measurement. Carefully read and repeat the steps in section 3.3. Arm cuff is applied too loosely. Apply the arm cuff tighter. Refer to section 3.1. The batteries are low.
4. Error Messages and Troubleshooting Display Cause Solution Air plug is disconnected. Insert the air plug securely. Refer to section 3.1. Arm cuff is applied too loosely. Apply the arm cuff tighter. Refer to section 3.1. Air is leaking from the arm cuff. Replace the arm cuff with a new one. Refer to section 5.3. Movement during measurement and the arm cuff has not been inflated sufficiently. EN Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3.
4. Error Messages and Troubleshooting 4.2 Troubleshooting Problem Cause Solution Arm cuff is applied too loosely. Apply the arm cuff tighter. Refer to section 3.1. Movement or talking during measurement. Remain still and do not talk during measurement. Refer to section 3.3. Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff. Refer to section 3.1. The air connector is not securely connected into the air jack.
4. Error Messages and Troubleshooting Problem Cause The batteries are empty. Replace the batteries with new ones. Refer to section 2.1. The batteries have been inserted incorrectly. Insert the batteries with the correct (+/-) polarity. Refer to section 2.1. Nothing happens when you press the buttons. Other problems. Solution EN • Press the START/STOP button and repeat measurement. • Replace the batteries with new ones. If the problem continues, contact your OMRON retail outlet or distributor.
5. Maintenance and Storage 5.1 Maintenance To protect your device from damage, please observe the following: • • • • Store the device and the components in a clean, safe location. Do not use any abrasive or volatile cleaners. Do not wash the device and any components or immerse them in water. Do not use petrol, thinners or similar solvents to clean the device. • Use a soft and dry cloth, or a soft and moistened cloth and neutral soap to clean on the monitor and the arm cuff.
5. Maintenance and Storage 5.2 Storage Keep the device in its storage case when not in use. 1. Unplug the air plug from the air jack. 2. Gently fold the air tube into the EN arm cuff. Note: Do not bend or crease the air tube excessively. 3. Place the monitor and the arm cuff in the storage case. Do not store the device in the following situations: • If the device is wet. • Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapours.
5. Maintenance and Storage 5.
5. Maintenance and Storage Using the Optional AC Adapter 1. Insert the AC adapter plug into the AC adapter jack on the rear side of the monitor. EN 2. Plug the AC adapter into an electrical outlet. To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the monitor.
6. Specifications Product description Model Display Measurement method Measurement range Accuracy Inflation Deflation Memory Rating Power source Battery life Applied part Automatic Blood Pressure Monitor OMRON M6 Comfort (HEM-7321-E) LCD digital display Oscillometric method Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats / min.
6. Specifications Notes: • These specifications are subject to change without notice. • In the clinical validation study, the 5th phase was used on 85 subjects for determination of diastolic blood pressure. • This device has not been validated for use on pregnant patients. EN • This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive).
6. Specifications Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.
7. Warranty Thank you for buying an OMRON product. This product is constructed of high quality materials and great care has been taken in its manufacturing. It is designed to give you every satisfaction, provided that it is properly operated and maintained as described in the instruction manual. This product is guaranteed by OMRON for a period of 3 years after the date of purchase. The proper construction, workmanship and materials of this product is guaranteed by OMRON.
8. Some Useful Information about Blood Pressure What is Blood Pressure? Blood pressure is a measure of the force of blood flowing against the walls of the arteries. Arterial blood pressure is constantly changing during the course of the heart’s cycle. The highest pressure in the cycle is called the Systolic Blood Pressure; the lowest is the Diastolic Blood Pressure. Both pressures, the Systolic and Diastolic, are necessary to enable a physician to evaluate the status of a patient’s blood pressure.
Manufacturer OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www.omron-healthcare.com Production facility Site de production Produktionsstätte Stabilimento di produzione Planta de producción OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD.