HEM-7221-E_B_M02_100928.pdf Check following components! Manufacturer OMRON HEALTHCARE Co., Ltd. 24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto 615-0084 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Kruisweg 577, 2132 NA Hoofddorp, THE NETHERLANDS www.omron-healthcare.
Contents 1 2 b F Intended Use This product is designed to measure the blood pressure and pulse rate of people within the range of the designated arm cuff, following the instructions in this instruction manual. It is mainly designed for general household use. Please read the Important Safety Information in this instruction manual before using the unit. K c L d a Important Safety Information ........................1 1. Overview.................................................................3 2.
HEM-7221-E_A_M_main.book Page 1 Thursday, October 7, 2010 2:12 PM Important Safety Information Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis. Please read this section carefully before using the unit. Warning: • Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) • Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.
HEM-7221-E_A_M_main.book Page 2 Thursday, October 7, 2010 2:12 PM Important Safety Information (AC Adapter Usage) • Use only the original AC adapter (optional) designed for this unit. Use of unsupported adapters may damage and/or may be hazardous to the unit. • Plug the AC adapter into the appropriate voltage outlet. Do not use a multiple-tap. • Do not use the AC adapter if the unit or the power cord is damaged. Turn off the power and unplug the power cord immediately.
HEM-7221-E_A_M_main.book Page 3 Thursday, October 7, 2010 2:12 PM 1. Overview Open the rear cover page to read following: The alphabet in the rear cover page correspond to those in the body page.
HEM-7221-E_A_M_main.book Page 4 Thursday, October 7, 2010 2:12 PM 2. Preparation 2.1 Installing/Replacing the Batteries 1. Remove the battery cover. 2. Insert four “AA” batteries as indicated in the battery compartment and then replace the battery cover. Notes: • If the battery low symbol ( ) appears on the display, turn the unit off then replace all batteries at the same time. • The measurement values continue to be stored in memory even after the batteries are replaced.
HEM-7221-E_A_M_main.book Page 5 Thursday, October 7, 2010 2:12 PM 2. Preparation 2.2 Setting the Date and Time 1. Press the button. 2. Set the monitor to the correct date and time before taking a measurement for the first time. EN 1) Push or change. 2) Push the buttons to button to confirm. Next setting appears. Year Month Day Hour Minute 3. Press the O/I START button to store the setting.
HEM-7221-E_A_M_main.book Page 6 Thursday, October 7, 2010 2:12 PM 3. Using the Unit Open both the front and rear covers to read following: The alphabet and number in the cover page correspond to those in the body page. 3.1 Applying the Arm Cuff Remove tight-fitting clothing or tight rolled up sleeve from your upper arm. Do not place the cuff over thick clothes. 1. Insert K the air plug into F the air jack. 2. Hold L the grip on the cuff securely with your hand. 3. Apply the cuff to your upper arm.
HEM-7221-E_A_M_main.book Page 7 Thursday, October 7, 2010 2:12 PM 3. Using the Unit 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. No bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement. • Sit on a chair with your feet flat on the floor. • Sit upright with your back straight. --- c • The cuff should be at the same level as your heart.
HEM-7221-E_A_M_main.book Page 8 Thursday, October 7, 2010 2:12 PM 3. Using the Unit 3.3 Taking a Reading Notes: • To cancel a measurement, press the O/I START button to turn the unit off and to release the air in the arm cuff. • Remain still while taking a measurement. 1. Press the O/I START button. The cuff will start to inflate automatically.
HEM-7221-E_A_M_main.book Page 9 Thursday, October 7, 2010 2:12 PM 3. Using the Unit Dual Check System Press the O/I START button, then the Dual Check System lamp lights and the unit starts detecting a device error. If the unit is accurate and functioning correctly, the Dual Check System lamp remains lit during the measurement. If an error is detected, the Dual Check System lamp flashes and “Er” will appear on the display.
HEM-7221-E_A_M_main.book Page 10 Thursday, October 7, 2010 2:12 PM 3. Using the Unit • Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats can influence the results of the measurement. The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated. If the measurement results are affected by irregular heartbeats but the result is valid, the result is shown together with the irregular heartbeat symbol ( ).
HEM-7221-E_A_M_main.book Page 11 Thursday, October 7, 2010 2:12 PM 3. Using the Unit 3.4 Using the Memory Function The monitor automatically stores the result up to 90 sets. It can also calculate an average reading based on the measurements from the last three readings taken within 10 minutes. If there are only two readings in memory for that period, the average will be based on two readings. If there is one reading in memory for that period, the average will be based on one reading.
HEM-7221-E_A_M_main.book Page 12 Thursday, October 7, 2010 2:12 PM 3. Using the Unit To View the Readings Stored in Memory 1. Press the button, while the average value is displayed. The Memory number appears for a second before the pulse rate is displayed. The newest set is numbered “1”. 2. Press the or Alternating date/ time display button to view the readings stored in memory.
HEM-7221-E_A_M_main.book Page 13 Thursday, October 7, 2010 2:12 PM 4. Troubleshooting and Maintenance 4.1 The Icons and Error Messages Error Display Cause Remedy Irregular heartbeats are detected. Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Repeat the steps in section 3.3. If this error continues to appear, contact your doctor. Movement during measurement. Carefully read and repeat the steps in section 3.3. Cuff is not applied correctly. Apply the arm cuff correctly.
HEM-7221-E_A_M_main.book Page 14 Thursday, October 7, 2010 2:12 PM 4. Troubleshooting and Maintenance Error Display Cause Remedy Air plug disconnected. Insert the plug securely. Refer to section 3.1. Arm cuff not applied correctly. Apply the arm cuff correctly. Refer to section 3.1. Air is leaking from the arm cuff. Replace the cuff with the new one. Refer to Chapter 5. Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3. 14 Cuff over inflated.
HEM-7221-E_A_M_main.book Page 15 Thursday, October 7, 2010 2:12 PM 4. Troubleshooting and Maintenance 4.2 Troubleshooting Problem Cause Remedy Arm cuff not applied correctly. Apply the arm cuff correctly. Refer to section 3.1. Movement or talking during measurement. Remain still and do not talk during measurement. Refer to section 3.3. Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff. Refer to section 3.1.
HEM-7221-E_A_M_main.book Page 16 Thursday, October 7, 2010 2:12 PM 4. Troubleshooting and Maintenance Problem Nothing happens when you press the buttons. Other problems. 16 Cause Remedy The batteries are empty. Replace the batteries with new ones. Refer to section 2.1. The batteries have been inserted incorrectly. Insert the batteries with the correct (+/ -) polarity. Refer to section 2.1. • Press the O/I START button and repeat measurement.
HEM-7221-E_A_M_main.book Page 17 Thursday, October 7, 2010 2:12 PM 4. Troubleshooting and Maintenance 4.3 Maintenance To protect your unit from damage, please observe the following: • • • • • • • • Do not subject the main unit and the cuff to extreme temperatures, humidity, moisture or direct sunlight. Do not fold the cuff or tubing tightly. Do not disassemble the unit. Do not subject the unit to strong shocks or vibrations (for example, dropping the unit on the floor).
HEM-7221-E_A_M_main.book Page 18 Thursday, October 7, 2010 2:12 PM 4. Troubleshooting and Maintenance 4.4 Storage Keep the unit in its storage case when not in use. 1. Unplug the air plug from the air jack. 2. Gently fold the air tube into the arm cuff. Note: Do not bend the air tube excessively. 3. Place the arm cuff and main unit in the storage case. Do not store the unit in the following situations: • If the unit is wet.
HEM-7221-E_A_M_main.book Page 19 Thursday, October 7, 2010 2:12 PM 5. Optional Parts Arm Cuff AC Adapter S Arm circumference 22 - 42 cm Comfort cuff-9956685-4 Adapter S-9515336-9 EN Using the Optional AC Adapter 1. Insert the AC adapter plug into the AC adapter jack on the rear side of the main unit. 2. Plug the AC adapter into an electrical outlet. To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the main unit.
HEM-7221-E_A_M_main.book Page 20 Thursday, October 7, 2010 2:12 PM 6. Technical Data Product Description Model Display Measurement Method Measurement Range Accuracy Inflation Deflation Memory Rating Power Source Battery Life Applied Part Protection Against Electric Shock Automatic Blood Pressure Monitor OMRON M6 Comfort (HEM-7221-E / HEM-7221-E8) LCD Digital Display Oscillometric method Pressure: 0 to 299 mmHg Pulse: 40 to 180/min.
HEM-7221-E_A_M_main.book Page 21 Thursday, October 7, 2010 2:12 PM 6. Technical Data • This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive). • This blood pressure monitor is designed according to the European Standard EN1060, Noninvasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems. • This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co. Ltd.
HEM-7221-E_A_M_main.book Page 22 Thursday, October 7, 2010 2:12 PM 6. Technical Data Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation.
HEM-7221-E_A_M_main.book Page 23 Thursday, October 7, 2010 2:12 PM 7. Some Useful Information about Blood Pressure What is Blood Pressure? Blood pressure is a measure of the force of blood flowing against the walls of the arteries. Arterial blood pressure is constantly changing during the course of the heart’s cycle. The highest pressure in the cycle is called the Systolic Blood Pressure; the lowest is the Diastolic Blood Pressure.
HEM-7221-E_A_M_main.book Page 24 Thursday, October 7, 2010 2:12 PM 7. Some Useful Information about Blood Pressure Why is it a Good Thing to measure Blood Pressure at Home? 24 Upper curve: systolic blood pressure Lower curve: diastolic blood pressure 200 150 mmHg Many factors such as physical activity, anxiety, or the time of day, can influence your blood pressure. A single measurement may not be sufficient for an accurate diagnosis.
HEM-7221-E_A_M_main.book Page 25 Thursday, October 7, 2010 2:12 PM 7. Some Useful Information about Blood Pressure Classification of Blood Pressure by the World Health Organization The World Health Organization (WHO) and the International Society of Hypertension (ISH) developed the Blood Pressure Classification shown in this figure. This classification is based on the blood pressure values measured on people in a sitting position in outpatient departments of hospitals.