R INSTRUCTION MANUAL PORTABLE NEBULIZER MODEL NE-U03V
TABLE OF CONTENTS Introduction………………………………………………………………………2 Important Safeguards ……………………………………………………………3 Know Your Unit …………………………………………………………………4 Cautions-Warnings ……………………………………………………………5, 6 Unit Assembly …………………………………………………………………7 Operating Instructions …………………………………………………………8 Unit Disassembly & Maintenance ………………………………………………9 Cleaning ………………………………………………………………………10 Mesh Cleaning …………………………………………………………………11 Storage …………………………………………………………………………12 Battery Operation ………………………………………………………………13 Troubleshooting
IMPORTANT SAFEGUARDS When using electrical products, especially when children are present, basic safety precautions should always be followed, including: READ ALL INSTRUCTIONS BEFORE USING DANGER - To reduce the risk of electrocution: 1. Always unplug this product immediately after using. 2. Do not use this product while bathing. 3. Do not place or store this product in a place where it can fall or be pulled into a tub or sink. 4. Do not place or drop into water or any liquid. 5.
KNOW YOUR UNIT Mouthpiece Mask Mesh Cap (Extra Cap Included) Vibration Horn Unit Cover Medication Cup On/Off Button MicroAir Unit Rinse Bottle AC Adapter* Battery Lid AC Jack * The shape of the plug may vary from country to country.
CAUTIONS - WARNINGS 1. WARNING The most delicate part of the unit is the mesh cap.NEVER touch the mesh cap directly. Follow the cleaning instructions exactly to avoid damaging the mesh cap. (See page 10 for cleaning instructions.) 2. WARNING Before using the MicroAir for the first time, or after storing it for an extended period, be sure to clean and sterilize all necessary parts as explained on page 10. 3. WARNING NEVER submerge or wet the AC power unit or cord. An electrical shock may result. 4.
CAUTIONS - WARNINGS 9. WARNING Do not submerge the unit in water to prevent damage. 10. WARNING Fill the medication cup before using the unit. Do not shake the unit during use. 11. CAUTION Assure that the proper adapter is used for connection to any electrical outlet to prevent damage to the unit or injury. 12. CAUTION Check the medication cup and mouthpiece weekly. Replace as necessary. 13. CAUTION Handle the mess cap with care to prevent breakage. 14. CAUTION The medication cup holds 5 ml.
UNIT ASSEMBLY How to Assemble A. Clean and sterilize all unit parts before the first use or after a long storage period. B. Fill the medication cup before plugging in the unit and turning on the power. 2. Attach the medication cup to the back of the unit. Push the medication cup vertically until a click is heard. 1. Remove the cover by gently pushing in an upward direction. 3. The mesh cap is attached as follows: A. Place the cap position indicator onto the unit turning counterclockwise to 10 o’clock B.
OPERATING INSTRUCTIONS 1. Before Operating Make sure all parts are mounted properly. Check to fill the medication up to correct level of 5 ml. 2. Plug In Connect the adapter into a standard AC 120V outlet. Four (4) "AA" alkaline batteries may also be used to run the unit. 3. Nebulizing Start the unit by pressing the ON/OFF button. Continuous aerosol is controlled by holding the button down for four (4) seconds. Press the ON/OFF button again to turn off the unit. NOTES: A.
UNIT DISASSEMBLY & MAINTENANCE NOTE: The mesh cap may be damaged by hardened medication that is left in the vibration horn effecting how the unit nebulizes. The following are maintenance recommendations after each use to keep your unit working at its best level. 1. NOTE: Be sure that the unit is turned OFF. Remove the medication cup and dispose of any remaining medications. 2. Be sure to nebulize all medication from the unit.
CLEANING Be sure unit is OFF and the power cord is disconnected before cleaning. Nebulizer Parts and Casing (EXCEPT MESH CAP) 1. There are several ways to clean the medication cup and mouthpiece. A. Soak parts in a mixture of one part vinegar and three parts distilled water for 30 minutes B. Use a commercially available disinfectant C. Parts may also be boiled for 10 minutes. D. All parts must be rinsed thoroughly and air dried. 2.
MESH CLEANING NOTE: The mesh cap is the most delicate part of the MicroAir. It is important not to directly touch the mesh cap. Follow the cleaning instructions to avoid damaging the mesh cap and keep it in good condition. A. After Every Cleaning 1. Remove the medication cup and rinse with water after turning the unit OFF. 2. Nebulize all medication remaining in the mesh cap and transducer after removing the medication cup. Turn the unit off when the aerosol stops. 3.
STORAGE 1. Completely dry all cleaned and sterilized parts. Mesh Cap Attach the mesh cap to the unit as describe in the operating instructions. Handle the mesh cap with care as it is very fragile and may be easily damaged. Medication Cup Attach the medication cup to the unit as described in the operating instructions. Unit Cover Attach cover to the unit by sliding the cover down until it clicks into place. Mouthpiece or Mask The mouthpiece or mask can be stored in the carrying case or a plastic bag.
BATTERY OPERATION Using Batteries 1. Turn the unit upside down. The battery lid will lift off by pressing down towards the arrows. 2. Insert four (4) "AA" alkaline batteries matching the polarities to the marks( + or - ) inside the unit. 3. Replace the battery lid by pressing downward until it clicks into place as shown on the right. NOTE: The batteries will need to be replaced if either of these conditions occur: A. When the green operation light blinks during continuous use B.
TROUBLE SHOOTING GUIDE (check before sending for service*) Check the following if your unit should fail during operation. You can also refer to the pages of this manual for complete instructions. Symptom Continuous operation button does not light up. • With AC adapter in use Continuous operation button does not light up. • With battery No nebulization. Check Remedy ➀ Is the AC adapter plugged into 120V electrical outlet? Insert plug. ➁ Is the AC adapter connected to the unit? Connect correctly.
AVAILABLE ACCESSORIES A full range of accessories make the MicroAir adjustable to virtually every requirement.
SPECIFICATIONS* Power Source: AC 120v 60 Hz Power Consumption: Approximately 1.2 w Vibration Frequency: Approximately 65 KHz Nebulization Rate: 0.25 ml/min ±0.05 Particle Size Range: 6 ~ 9.5 microns Medication Cup: Maximum capacity 5 ml Dimensions: 67 x 32 x 165 mm Weight: 148 grams, approximately 5.2 oz Battery: 4 “AA” alkaline (not included) Battery Life: Approximately 300 Minutes * Specifications subject to change without notice.
COMMON QUESTIONS ABOUT THE MICROAIR Q. A. Can I use any type of double “AA” batteries in the MicroAir? We recommend that alkaline batteries be used for best performance. Q. A. How long will a set of batteries operate the MicroAir? The expected battery life is 300 minutes of operation. Q. A. What other sources of power can the MicroAir use? The MicroAir can operate from either AC power or battery. Q. A. How often should I clean the MicroAir? You must clean the MicroAir after each use.
COMMON QUESTIONS ABOUT THE MICROAIR Q. A. Will the mesh break easily? The mesh is like all small delicate parts that can be broken if not gently handled. It is important that after cleaning the mesh that it be placed back on the unit. This will protect the mesh from damage while the unit is not being used. Be sure to replace the unit cover after use for added protection. Q.
TWO YEAR LIMITED WARRANTY Your Omron MicroAir Portable Nebulizer is warranted to the original retail purchaser to be free from defects in material and workmanship for a period of 2 years from the date of purchase. This warranty does not cover failures resulting from misuse, abuse, alteration of the Omron Nebulizer, use of improper components, or failure to comply with the operating instructions. This warranty does not include the disposable accessories.
R Distributed by OMRON HEALTHCARE, INC. 300 Lakeview Parkway Vernon Hills, Illinois 60061 Phone: 1-800-634-4350 www.omronhealthcare.com Copyright © OMRON HEALTHCARE, INC. NE-U03 INST Rev B OMRON uses recycled paper to help preserve our natural resources.