NE-U17-E_E_en_M03_101027.pdf OMRON ULTRASONIC NEBULISER Model NE-U17 Instruction Manual Table of Contents Be sure to Read This Section ■ Thank you very much for purchasing this OMRON Ultrasonic Nebuliser. ■ Be sure to read this instruction manual carefully before using the unit so that you can use it safely and correctly. ■ Please keep this instruction manual always at hand for future reference. Exemptions from Liabilities......... 1 Intended Use .............................. 2 Notes on Safety ......
Exemptions from Liabilities Please understand that OMRON is not liable for the following: 1. Occurrence of trouble and/or damage caused by the maintenance and/or repair conducted by other than OMRON or OMRON’s authorized dealer. 2. Trouble and/or damage of OMRON’s product caused by the product of other manufacturer not delivered by OMRON. 3. Trouble and/or damage caused by the maintenance and/or repair using the repair part(s) not authorized by OMRON. 4.
Intended Use Medical Purpose This product is intended to be used for inhaling medication for respiratory disorders or for humidification in combination with oxygen therapy. Intended User • Legally certified medical experts, such as doctor, nurse and therapist, or healthcare personnel or patient under the guidance of qualified medical experts. • The user should also be capable of understanding general machine operation and the content of instruction manual.
Notes on Safety ■ The warning signs and the sample icons shown here are listed for you to use this product safely and correctly as well as to prevent the risk of injury to you and others. ■ The icons and meanings are as follows. Warning sign Contents Warning Caution Indicates matters in which the possibility of death or severe bodily injury may arise as a result of incorrect handling. Indicates matters in which bodily injury or material damage* may arise as a result of incorrect handling.
Notes on Safety Caution If the nebulisation parts have already become dirty before using the unit, clean them first before use. • If the medication remains in the nebulisation unit, your physical conditions may change for the worse. Wash off the remaining medication in the parts thoroughly with water after cleaning these parts with disinfectant. • If you inhale with the remaining disinfectant in the parts, your physical conditions may change for the worse.
Notes on Safety General Advice Do not use the unit for the purpose other than for inhalation. Do not use any parts other than the accessories or the optional parts listed on page 22. Do not fill the medication cup with more than 150 ml medication or less than 5 ml. • The unit may not nebulise. Do not move the unit while medication remains in the unit. Do not block the air ventilation holes at the rear side of the nebuliser. Place the nebuliser on a plain and stable surface during operation.
Features of the Product Aerosol= Nebulised medication About the principle of nebulisation in the NE-U17 ➀ The NE-U17 transmits the energy of ultrasonic vibration from the vibrator located at the bottom of the water tank into water. ➁ The ultrasonic vibration is transmitted to the medication in the medication cup through the water in the water tank. ➂ The ultrasonic vibration emits medication from the medication cup (like a fountain) and this is dispersed as aerosol.
Composition of the Product Ultrasoni c Nebu liser Inhalation hose (M) with a cuff, 70 cm NE-U17 AIR-FLO W VOL UME WATER LEVEL MEDICAT ION CUP COVER FAN COVE R TIMER NEBULIS ATION VOL UME CONTINU OUS TIMER( MIN) CONTIN UOUS POWER Mouthpiece set START/ STOP Two medication cups Main unit Power cord – As the unit is a Class I electroshock protection type, be sure to use only the supplied power cord with an “earth”.
Names of the Parts * indicates replaceable supplies. Medication cup cover Air duct cover Washer Fan cover Mouthpiece set Air filter case Air filter* Handle Inhalation hose M Display of air volume Medication cup* Display Ultrasonic Nebuliser (Refer to the “Description of the LCD Display” on Page 7.) Air volume adjustment knob Display of nebulisation rate (Refer to the “Description of the LCD Display” on Page 7.
How to Prepare the Unit before Inhalation – Clean the parts before starting the inhalation. [Parts that can be disinfected by boiling] (For the disinfecting method, refer to Page 17.) Submerge these parts in a container filled with sufficient water and boil for 10 minutes. Medication cup cover 1 Filling the cooling water. 1) Remove the nebulisation section by rotating the medication cup cover-fixing lever clockwise.
How to Prepare the Unit before Inhalation 2 Filling medication into the medication cup. 1) Remove the medication cup cover by rotating it counterclockwise. 2 3 Assembling and attaching the nebulisation unit to the main unit 1) Fix the nebulisation unit by rotating the medication cup cover-fixing lever counterclockwise. 1 2) Attach the inhalation hose and the mouthpiece set. 2) Fill medication into the medication cup. Maximum of 150 ml Hose holder Minimum of 5 ml. 3) Turn on the Power Button ( ).
How to Inhale 1 2 Setting the inhalation time. Setting the inhalation time by rotating the timer knob clockwise. Push the START/STOP Button ( ). The fan will rotate and the unit starts to nebulise. POWE R 1) To set the timer The timer can be adjusted by 1 minute when setting the time up to 15 minutes. When the timer is used: The display counts down by 1 minute. 16 m in . Please note that the timer has intervals of 1 or 2 minutes between 15 and 30 minute timer settings. 18 m in . in.
How to Inhale 3 Inhaling the aerosol. 4 Finishing the treatment. When the timer is used: The buzzer sound notifies the completion of time set by the timer and the unit stops operation. When the operation ends, “0” on the display illuminates and the display changes to the initial set time. Sample display when the timer is set to 10 minutes. In approximately 3 seconds TIMER CONTINUOUS TIMER When the timer comes to end.
How to Inhale (Using Optional Parts) 1 Using the unit for a long time By using the optional “long time nebulisation set (4997217-7)” and the “supply medication set (4997206-1)”, the unit can nebulise continuously for a long time. ➀ Remove the medication cup cover of the main unit. Attach an optional long time nebulisation set in place of the medication cup cover and assemble the unit according to the procedure shown on Page 10. At this time, check that the air duct is not blocked by the air duct cap.
How to Inhale (Using Optional Parts) Oxygen supply pipe 3 Using a bacterial filter Having passed through a bacterial filter, the filtered air can be supplied to the inside of the long time nebulisation set. ➀ Remove the medication cup cover and assemble the long time nebulisation set with the main unit. To assemble, use the long time nebulisation set in place of the medication cup cover by following the procedure shown on Page 10. At this time, close the air duct with the air duct cap.
How to Take Care of the Unit How to Take Care of the Parts 1 Turn off the power by pushing the ON/ OFF Button ( ), and unplug the power plug from the electric outlet. 2 Remove the mouthpiece set, or inhalation mask. 3 Rotate the medication cup cover-fixing lever clockwise. 4 Remove the nebulisation unit for disassembling. 5 Discard the remaining medication. Discard the cooling water in the water tank by using the draining hose. 6 Clean and disinfect the parts.
How to Take Care of the Unit How to Take Care of the Air Filter 4 3 1 Replace the air filter when it becomes dirty. ➀ Remove the nebulisation unit from the main device. ➁ Remove the fan cover by lifting it up. ➂ Remove the air filter case from inside the fan cover. ➃ Remove the air filter from the air filter case to replace with a new one. Assemble the above mentioned parts in the reverse order. Nebulisation section 2 How to Take Care of the Fan Fan cover Open the fan cover and remove the fan.
How to Disinfect the Unit How to Disinfect the Unit Disinfection by boiling cannot be done to some parts. They may become deformed. [Parts that can be disinfected by boiling] Medication cup cover Washer Inhalation hose (M) with a cuff, 70 cm [Parts that cannot be disinfected by boiling] Air filter case Air duct cover Fan Fan cover Medication cup Air filter Medication cup holder Mouthpiece set Disinfection by disinfectant Different disinfectants have different anti-bacterial actions.
How to Disinfect and Store the Unit Sterilization with ethylene oxide gas (E.O.G.) Observe the Instruction Manual of the sterilizer and consult with the manufacturer of the E.O.G. sterilizer for the details. Some of the rubber and plastic parts of the nebuliser may have residual ethylene oxide gas. After sterilization, consult with the manufacturer of the sterilizer and provide sufficient aeration until the safe level to human beings can be obtained. The equipment and the parts to be sterilized by E.O.G.
Troubleshooting Where to Inspect Trouble The power lamp does not illuminate. Is the power plug securely plugged into the electric outlet? Plug the power plug in the electric outlet correctly. (Refer to Page 10.) Is ➀ of ERROR display flashing due to insufficient amount of water in the operating chamber? Is the medication cup correctly set to the main unit? Fill the water up to the specified water level mark in the operating chamber. (Refer to Page 9.) Set the medication cup correctly.
Specifications Technical data Product name Model Power source Power consumption Fuse Ultrasonic frequency Particle size Nebulisation rate Sound Air volume Amount of cooling water Capacity of medication cup External dimensions Weight of the main unit Protection class Operating condition Operating temperature / humidity Storage temperature / humidity / air pressure Accessories included : Omron Ultrasonic Nebuliser : NE-U17 (NE-U17-E) : 230 VAC, 50 Hz : Approx. 80 VA : T2A L250V : Approx. 1.7 MHz : * MMAD 4.
Specifications Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.
List of Optional Parts Medication cup (4997211-8) Bacterial filter (4997213-4) Air filter set (4997214-2) Air filter (4997215-0) Used to transmit the ultrasonic vibration. (If it has any damage or deformation, the unit may not be able to nebulise. Handle this part carefully.) Material: PP Packaging unit: 5 cups Used for inhalation such as for tracheal lavage. Materials: filter; PP, connector; PMP Packaging unit: 1 filter + 1 adapter Protects the unit from the dust in air.
Manufacturer OMRON HEALTHCARE Co., Ltd. 24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto, 615-0084 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Kruisweg 577, 2132 NA Hoofddorp, THE NETHERLANDS www.omron-healthcare.com Production facility OMRON MATSUSAKA Co., Ltd. 1855-370, Kubo-cho, Matsusaka-city, Mie-prefecture 515-8503, Japan OMRON HEALTHCARE UK LIMITED Opal Drive, Fox Milne Milton Keynes, MK15 0DG U.K. Subsidiary OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH John-Deere-Str.