2) Press the Fluo. button on the Temperature and Process Cycle screens to switch to the Fluorescent Graph
screen during the test†.
Fluorescence graph does not appear in “Stage 1” during PCR stage.
3) Press the> Stop button to terminate the test in the middle of run. The intermediate result of the stopped test
will be deleted.
Test Result Test Result Screen
Figure 18. Test result Screen
1) When the test is completed, the test result is automatically displayed on the screen (Scatter chart screen in Figure
18).
2) On the left side of the screen, start and end date and time of test are shown. On the top of the screen, sample ID,
concentration information is displayed.
3) In the middle of the screen, a well image is displayed. Yellow means negative well (-), green means green channel
positive, red channel negative (+-), red means red channel positive, green channel negative, and blue means both
channel positive wells.
4) Scatter chart shows the result of distribution of fluorescence values detected in each well in two dimensions. The
displayed distribution is based on a real-time curve, which is the most important advantage of Dr. PCR, and requires
no threshold adjustment.
5) Press the play button ( ) to see the scatter chart distribution move from cycle 1 to the last cycle set. You can stop
at the desired cycle by pressing the stop button.
6) Press Graph button to see real-time PCR curve of each channel (Figure 19, Figure 20).
Figure 19. CH1 Digital PCR Result Screen