T- DualScan T-DualScan
I T-DualScan T-DualScan is an innovative planar near-field system that offers the best compromise between accuracy, flexibility and measurement speed. The tower positioner can rotate 180° to switch easily from the single-probe set-up (0.8 - 110 GHz) to the multi-probe set-up (0.8 - 18 GHz). T-DualScan can also be offered as an upgrade to existing installations.
I T-DualScan System overview INSTRUMENTATION ROOM CHAMBER USB 5 Data Acquisition & Processing Platform 4 1 Real Time Controller 2 6 Triggers 3 Triggers N-PAC GPIB Primary Synthesizer Amplification Unit Mixer Unit Transfer Switching Unit RF Switch Auxiliary Synthesizer Rx Tx Positioner Controller Measurements can be performed in both continuous wave and pulsed mode.
I T-DualScan Standard system components Linear probe array The required scan area is calculated according to the following formula: Scan length = D + 2 L tg ( ) a Where: - is the relevant data angle in far-field - L, the distance between the probe and the AUT - and D, the antenna size.
I T-DualScan System specifications Multi-probe set-up* System specifications Single-probe set-up* 30 dBi AUT 30 dBi AUT PEAK GAIN ACCURACY PEAK GAIN ACCURACY 0.8 GHz - 1 GHz ± 0.5 dB 0.8 GHz - 1 GHz ± 0.5 dB 1 GHz - 6 GHz ± 0.5 dB 1 GHz - 6 GHz ± 0.5 dB 6 GHz - 18 GHz ± 0.5 dB 6 GHz - 18 GHz ± 0.5 dB Repeatability ± 0.3 dB 18 GHz - 40 GHz ± 0.5 dB Repeatability ± 0.3 dB - 20 dB SIDELOBES ACCURACY 0.8 GHz - 1 GHz ± 0.6 dB - 20 dB SIDELOBES ACCURACY 1 GHz - 6 GHz ± 0.6 dB 0.
I T-DualScan Probe positioning system Scanner horizontal slide - Horizontal translation of the scanner on the x axis Scanner vertical slide - Vertical translation of the measurement probe and the calibration arm Probe roll - Rotation of the measurement or calibration probe on the z axis Probe horizontal slide - Horizontal positioning of the probe on the z axis Probe array vertical slide - Vertical translation of the probe array for oversampling Azimuth rotation of the tower Mechanical characteristic
I T-DualScan Calibration process Probe array Number of modules Probe array length (cm) Scan area (cm) Number of probes 0.
I T-DualScan DUT positioning system (optional) H DUT positioning system with air cushion for easier movement. Single-probe The DUT positioner is composed of standard ORBIT/FR positioning equipment that can be adapted to the specific requirements of each customer. An innovative DUT positioner base allows for easy movement of the DUT positioning equipment, thanks to four air cushion pads.