Use Instructions
Orpyx
®
SI – Instructions for Use
Version 1.0 – August 2019
©2019 Orpyx
®
Medical Technologies Inc. - All rights reserved. All trademarks and registered trademarks are the property of their respective owners.
Page 13 of 26
4. Using Your Orpyx
®
SI Sensory Insoles System
A. STARTING YOUR ORPYX SI SENSORY INSOLE SYSTEM
STEP 1
Start the Mobile Application on your smartphone by clicking the Orpyx SI icon.
STEP 2
Your Orpyx SI sensory insoles are made to fit inside your diabetic or walking shoes. First, remove the
existing insole or sock liner from your shoes. With standard shoes, the sock liners may need to be
removed.
Orpyx SI Sensor Insoles may come in contact with a patient during normal use. It is
recommended the user wears sock with their Orpyx SI Sensor Insoles
Failure to remove an existing insole or sock liner from your shoe before inserting the Orpyx
SI sensory insole may result in a persistent high pressure state.
Place the appropriate left and right Orpyx SI sensory insoles into each shoe and run your hand over
the entire insole to ensure that the insole is lying flat. It is recommended that you wear your insole
system with the shoes that have been fitted by a professional.
Do not modify the Orpyx SI Sensory Insole, if required a professional orthotist or podiatrist
can modify insole outside the indicated margin to ensure proper fit.
When the insoles have been stored at a temperature below 5C, or above 40C, please allow 30 minutes
for the insoles to acclimate before using the system.
STEP 3
To start using your system, put on your shoes and apply pressure by standing and walking. This will
“wake up” the system and re-establish a wireless connection to your mobile application.
If your device is not connected, ensure that your device is fully charged and pair your device. Please
refer to section 2 in the Instructions for Use Manual or press “Care” on your mobile application for
“Setting up your Sensory Insoles”.