10. With the VERASENSE device and shim attached, physician should manually compress / apply load to the device and
verify the response on the User Interface prior to placing VERASENSE device into the tibial tray.
11. Place VERASENSE within tibial tray.
12. Confirm that the VERASENSE device with shim is fully seated when placed in the tibial tray.
13. Flex the joint throughout its full range of motion to ensure appropriate response on the User Interface.
14. Proceed with TKA process per physician / hospital protocol.
Note: If maximum allowable load of 70 lbf is reached in either compartment, the VERASENSE device must be
removed from the knee joint and “re-zeroed” by holding VERASENSE with superior side (articulating surface)
facing the floor for three (3) seconds, Re-Zero enabled will appear on the VERASENSE Software Application,
followed by Re-Zero Complete indicating that VERASENSE has been reset to zero; or Re-Zero button from the
VERASENSE Software Application by Pressing the Re-Zero button.
a. Compartment Re-Zero allows for a re-zero of the device without removing it from the joint. With
VERASENSE in the joint, apply a varus or valgus thrust to the knee to off-load the desired medial or lateral
compartment. When the VERASENSE surface is visually unloaded, select the laterality indicator of the
unloaded side to initiate the Re-Zero. Wait for Re-Zero Enabled and Re-Zero Completed messages to
display. Repeat the steps on the opposite compartment to complete the Re-Zero.
Note: Compartment Re-Zero is only available with VERASENSE devices that communicate with the
LinkStation MINI via Bluetooth. Check label on the front of the sensor box for the Bluetooth symbol .
15. In VSA 5.3.0.64 or greater, when a VERASENSE for Zimmer Biomet Persona device is activated with the magnet, the
user must acknowledge a description of the recommended workflow to handle temperature conditioning of the device
prior to use.
a. After device is inserted into the knee, pressing the CONDITION or
button on the VSA.
b. After conditioning is complete, follow the instructions to unload and immediately re-zero sensor.
c. Repeat this Step 15 if the VERASENSE for Zimmer Biomet Persona device is removed from the knee.
16. Upon completion of the procedure, deactivate the VERASENSE Software Application by pressing the Exit Button on
the User Interface.
17. Dispose of the VERASENSE device per institutional guidelines for biohazardous medical waste.
Note: VERASENSE for Zimmer Biomet Persona contains lithium batteries, thus special disposal instructions
should be taken in the state of California, USA. The device cannot be incinerated.
Model
Measurement Load Range
Load Accuracy
VERASENSE for Biomet Vanguard
5-40 lbf per compartment
3.5 lbf
VERASENSE for Stryker Triathlon
VERASENSE for Zimmer NexGen
VERASENSE for Smith & Nephew Legion
VERASENSE for Smith & Nephew Journey II
VERASENSE for Zimmer Biomet Persona