User's Manual

Performing Advanced Functions
SC Dig6 Dig5 Dig4 Dig3 Dig2 Dig1 Dig0
b7 b0
SC-Battery Charge (0-off, 1-on) Dig n – Digital I/O n
Fig. 16: The Digibyte
The remote version is limited to a single frame. An example of such a command is
given below:
9999 DATA 9999 30/9/1999 14:50:00
9999 DATA
30 9 1999 14 54 55 21 37 255 255 77 0 0 0 0 89 156 126 20 0 0
0 0 0 0 0 0 0 3197 0
#
Notice that if you need to get data that is not the last (newest) slot remotely from a
device, the ID must be supplied twice. If you need to get the last slot stored, you
can ignore the ID and the date/time parameters:
9999 DATA
9999 DATA
13 9 1999 19 26 36 21 37 255 255 79 0 0 0 0 87 148 149 15 0 0
0 0 0 0 0 0 0 3148 0
#
IMME
AVAILABLE FOR
A731 A733A732 A733GSM
DESCRIPTION Samples all inputs and immediately returns the sampled data.
P
ARAMETER The ID of the requested subsystem; default is the standard A/D subsystem of the
A733 (ID=0).
Note: Currently only the default subsystem is implemented on the A73x.
R
ETURNS A data block (see also “DATA” on page 29).
REMARKS GET only. The command needs a certain delay to execute (for example, for the
standard subsystem this delay amounts to two seconds. The delay is necessary to
allow for the sensors to settle after applying power to them.
REMOTE No.
E
XAMPLE IMME
15695 IMME 26 5 2003 17 18 28 21 37 0 0 127
0 554 0 0 84
0 0 0 1016 2048 3072 0 0 0 0 0 0
9056 0
#
FDEV
AVAILABLE FOR
A731 A733A732 A733GSM
DESCRIPTION Formats the internal memory (destroys all the data).
31