User's Manual

10
Safety Warnings
General
The manual contains information and warnings, which have to be followed by the user to ensure safe
installation, operation and to retain the equipment in safe condition. If the equipment is used in a
manner not specified in this manual, the protection provided by the equipment may be impaired.
This equipment is designed for monitoring bioreactors and must not be used for other purposes
Wear personal protective equipment adapted to lab use of system
This equipment must be used by trained personnel only
If the equipment is integrated in a larger system, the installation in that system and the safety of the
overall system is the responsibility of the integrator of the system.
The power supply unit must be connected to a power socket providing connection to earth.
The equipment must be installed such that it does not obstruct the way to the power supply connecting
cable.
Risk of electrical shock hazard
Any interruption of the electrical wiring outside the iLine F Pro will make the device hazardous.
Intentional interruption is prohibited.
Capacitors inside the iLine F Pro may still be charged, even if the iLine F Pro has been disconnected
from all voltage sources.
Any adjustment, maintenance and repair of the opened device under voltage must be avoided as far
as possible and, if inevitable, the activities must only be carried out by qualified personnel.
Only use the equipment with the dedicated power supply, as described in the specifications
Risk of damaging the equipment
In most cases, a pump is installed with the iLine F Pro. The membrane part of the pump is disposable
and may get damaged after sterilization, resulting in fluid leaking into the pump engine. Make sure
that there are no visible damage to the pump membrane ( See part 15 in the Connect > Get Started >
Description section) such as tears, before connecting to the pump engine.
Do not use the front pump module for other purposes than connecting the BioConnect disposable
The equipment needs cooling via the metal housing and the heat sink. Make sure the housing is not
covered and the sides and the top have a free air flow passing.
High power LED
The instrument operates high power LEDs that shall not be viewed with optical instruments, but are otherwise
safe for the user.
Other risks
The equipment housing may become hot during normal operation.
LED RADIATION
DO NOT VIEW DIRECTLY WITH OPTICAL INSTRUMENTS
CLASS 1M LED PRODUCT