User's Manual
42
Introduction
Various events may occur during the setup of a bioreactor run or during the run itself such as bad quality images
leading to measurement issues. In all cases the alarms or notifications are displayed with a banner on top of the
live screen.
The next sections describe each of them in detail.
BioConnect Alarms
Pump motor stalled
Description:
OsOne informs the user that the pump motor has stalled.
Cause(s):
• Pinched tubing during the setup in the lab.
• Tube welding not popped open
• BioConnect assembly defect (internal tubing pinched).
• Aggregates/Clogging present in the cell culture.
• Unexpected pinch valve action (software bug).
• Wrong pump profile setting.
• The pump motor could be too old and not working as expected (not powerful anymore).
When?
This alarm can be triggered anytime (before or during a monitoring) when the pump is executing one of the
following sequences:
• Priming
• Flushing
• Backup pumping (pumping mode when not in an active monitoring session)
Type:
ERROR
Does it prevent from starting a monitoring?
YES