User's Manual

42
Introduction
Various events may occur during the setup of a bioreactor run or during the run itself such as bad quality images
leading to measurement issues. In all cases the alarms or notifications are displayed with a banner on top of the
live screen.
The next sections describe each of them in detail.
BioConnect Alarms
Pump motor stalled
Description:
OsOne informs the user that the pump motor has stalled.
Cause(s):
Pinched tubing during the setup in the lab.
Tube welding not popped open
BioConnect assembly defect (internal tubing pinched).
Aggregates/Clogging present in the cell culture.
Unexpected pinch valve action (software bug).
Wrong pump profile setting.
The pump motor could be too old and not working as expected (not powerful anymore).
When?
This alarm can be triggered anytime (before or during a monitoring) when the pump is executing one of the
following sequences:
Priming
Flushing
Backup pumping (pumping mode when not in an active monitoring session)
Type:
ERROR
Does it prevent from starting a monitoring?
YES