User's Manual

43
Is it handled by OsOne?
If this error occurs during a monitoring, a software component running in the background called the “Restart
tool” will:
Close OsOne
Restart OsOne
Restart the monitoring
If the error is not automatically resolved after 3 consecutive tries of the “Restart tool”, monitoring is stopped.
How to fix it?
Check if any tubing is pinched
Check in the OsOne “Live” image if no aggregates are present in the field of view and if yes, try to flush
the fluidic system
Replace the BioConnect
How to prevent it?
Be careful not to pinch any tubing during the setup
Have the motor replaced during the maintenance if it appears that it is too old and makes noise during
normal operations.
Pump driver error
Description:
A hardware error related to the pump motor has been detected by OsOne
Cause(s):
There was an unexpected error with the pump:
When?
This alarm can be triggered anytime when OsOne uses the BioConnect
Type:
ERROR