User's Manual

45
Type:
ERROR
Does it prevent from starting a monitoring?
YES
Does it interrupt the monitoring?
YES
Is it handled by OsOne?
YES
If this error occurs during a monitoring, the “Restart tool” will:
Close OsOne
Restart OsOne
Restart the monitoring
If the error is not automatically resolved after 3 consecutive tries of the “Restart tool”, monitoring is stopped.
How to fix it?
Restart OsOne
Remove the BioConnect cartridge from the microscope and insert it again
Power off the microscope, wait a minute and power the microscope on.
If the above actions do not fix the issue, please contact Ovizio support (support@ovizio.com).
How to prevent it?
There is no way to prevent that kind of error.
Cannot start the pump error
Description:
OsOne informs the user that there was an error when OsOne tried to start a pump sequence.
Cause(s):
OsOne tried to start a sequence while the pump was still busy