User's Manual

47
Cause(s):
OsOne lost the connection with the BioConnect cartridge during a monitoring and then stop it.
When?
This alarm can be triggered anytime during a monitoring.
Type:
ERROR
Does it prevent from starting a monitoring?
NO
Does it interrupt the monitoring?
YES
Is it handled by OsOne?
NO
How to fix it?
Insert the BioConnect cartridge back into the iLine F Pro and restart the monitoring.
How to prevent it?
By educating the OsOne users not to remove a BioConnect cartridge during a monitoring.
Backup pumping disabled
Description:
OsOne informs the user that the backup pumping has been disabled.
Cause(s):
An authenticated user pressed the “Stop pumping” button.
When?
When the monitoring is not running
When the “Stop pumping” button is enabled