Calibration Alarms
Calibration file missing
Description:
OsOne informs the user that the calibration file is missing.
Cause(s):
OsOne did not find a file named CalibrationImage.h5 inside the “C:\Users\<OsOne
user>\AppData\Roaming\OsOne\Calibration” folder
When?
This alarm is triggered at the start of OsOne.
Type:
ERROR
Does it prevent from starting a monitoring?
YES
Does it interrupt the monitoring?
NO
Is it handled by OsOne?
YES, if the BioConnect cartridge is not inserted at the start of OsOne: in that case, OsOne will automatically take
a new background image.
NO, if the BioConnect cartridge is inserted inside the microscope at the start of OsOne.
How to fix it?
By removing the BioConnect cartridge from the microscope the time that OsOne can take a new background
image.
How to prevent it?
By never deleting the CalibrationImage.h5 file inside the “C:\Users\<OsOne
user>\AppData\Roaming\OsOne\Calibration” folder.