Outdated calibration file
Description:
OsOne informs the user that the calibration image is outdated.
Cause(s):
The calibration image has been taken more than 30 days ago.
When?
This alarm is triggered at the start of OsOne when OsOne checks when the calibration image has been taken.
Type:
ERROR
Does it prevent from starting a monitoring?
YES
Does it interrupt the monitoring?
NO
Is it handled by OsOne?
YES, if the BioConnect cartridge is not inserted at the start of OsOne: in that case, OsOne will automatically take
a new background image.
YES, if the BioConnect cartridge is inserted and a monitoring session is started. In that case OsOne acquires a
background image when cells are flowing and hence blurred, leading to an equivalent background image as
above
NO, if the BioConnect cartridge is inserted inside the microscope at the start of OsOne and cells are not flowing
How to fix it?
By removing the BioConnect cartridge from the microscope the time that OsOne takes a new background image
if a monitoring session is not active.
How to prevent it?
By occasionally removing the BioConnect cartridge (when OsOne is running) so that OsOne can automatically
take a new background image.