User's Manual

52
Error during calibration
Description:
OsOne informs the user that an error occurred while taking a new background image.
Cause(s):
A hardware error occurred while OsOne tried to capture a background image.
The user inserted the BioConnect cartridge while OsOne was capturing a background image and a
monitoring session is not active.
When?
This alarm is triggered while OsOne is capturing a background image (outside of a monitoring).
Type:
ERROR
Does it prevent from starting a monitoring?
YES
Does it interrupt the monitoring?
NO
Is it handled by OsOne?
NO.
How to fix it?
By removing the BioConnect cartridge and restarting OsOne.
If restarting OsOne does not solve the issue, try to power off the microscope, then power on.
If powering off and on the microscope does not solve the issue, please contact Ovizio support
(support@ovizio.com).
How to prevent it?
There is no way to prevent that kind of error.