User's Manual

54
How to fix it?
If the automatic flush wizard does not solve the issue, it is recommended to stop the monitoring and start a
manual flush (duration = 30 minutes).
How to prevent it?
There is no way to prevent that kind of error.
Fluidic issue Blurry images
Description:
OsOne informs the user that a potential fluidic issue could have occurred during the monitoring.
Cause(s):
If the tubing is not inserted inside the pinch valves during the monitoring, it is possible that the cells
moved during the acquisition, making the recorded images blurry.
Large aggregates in the images.
Bubbles in the imaging window
When?
This alarm is triggered during the monitoring, when OsOne completed the computation of the last recorded
images.
Type:
ERROR
Does it prevent from starting a monitoring?
NO
Does it interrupt the monitoring?
YES
Is it handled by OsOne?
YES. If a fluidic issue occurs during the monitoring, OsOne will automatically (automatic flush wizard):
Stop the monitoring
Start a flush sequence of 5 minutes
Restart the monitoring.
If this error occurs 5 times consecutively without being resolved, the above automatic actions won’t be
performed anymore.