User's Manual

57
Type:
ERROR
Does it prevent from starting a monitoring?
NO
Does it interrupt the monitoring?
NO
Is it handled by OsOne?
If OsOne detects out of focus cells in the images, it will try to correct the focus automatically without stopping
the monitoring. If cells are still out of focus after this focus correction, OsOne will trigger the “Out of focus cells”
alarm.
How to fix it?
It is recommended to stop the monitoring and restart it so that a long-range auto focus is performed by the
monitoring wizard. If the restart of the monitoring does not solve the issue, it is recommended to:
Stop the monitoring
Activate the “Start monitoring with manual focus” in the monitoring configuration screen.
Set the focus manually
Start the monitoring
How to prevent it?
It is recommended to look at the first images (just after having started a monitoring) and to check if the cells
are in focus when OsOne is recording captures.
No cell detected at startup of the monitoring
Description:
OsOne informs the user that no cells have been detected when starting the monitoring, and that the monitoring
has not started
Cause(s):
No cells have been detected in the image at the start of the monitoring.
When?
This alarm is triggered at the end of the monitoring wizard.
Type:
WARNING