User's Manual

60
Hardware failure
Description:
OsOne informs the user that it could not connect with one (or many) of the hardware components at the start
of OsOne.
Cause(s):
OsOne could not connect with one (or many) of the hardware components at startup.
When?
This alarm is triggered at the start of OsOne
Type:
ERROR
Does it prevent from starting a monitoring?
YES
Does it interrupt the monitoring?
YES, if it happens during an automatic restart performed by the “Restart tool”.
Is it handled by OsOne?
NO
How to fix it?
If restarting OsOne does not solve the issue, please try to power off the microscope, wait a minute and power
the microscope on.
If the issue still occurs after having power off/on the microscope, please contact the Ovizio support
(support@ovizio.com).
How to prevent it?
There is no way to prevent a hardware failure at the start of OsOne.