User's Manual

78
Logs
When the microscope and its software are in use, all the actions taken by the user and the software are stored
into files called log files or audit trail. In case of problems with the microscope or its software these files give
crucial information about the origin of the issues.
In case of problems, we recommend changing the logging level to Trace and to try to reproduce the issue.
Detailed information about the microscope behavior will be recorded and can subsequently be sent to
support@ovizio.com for further analysis and diagnosis.
The log configuration screen lets the user choose the number of back-up files to keep and their maximum size.
We recommend keeping the default values unless the Ovizio support team gives you other instructions.
Audit trail
OsOne is recording all user actions inside a filed named "audittrail.log". The "audittrail.log" log file is part of the
OsOne log files.
User actions
User actions are actions initiated by a user or involving a user using OsOne:
from the microscope touch screen on any of the OsOne user interfaces (e.g., starting a monitoring via
the touch screen by pressing the "Start" button)
on disposables connected to the microscope (e.g., inserting the Bioconnect cartridge inside the
microscope)
remotely via OPC or via a Web service (e.g., starting a monitoring remotely via OPC)