User's Manual

89
Maintenance
Cleaning
The instrument can be cleaned with a soft cloth, wetted with IPA (iso propyl alcohol), if required
The instrument does not contain user serviceable parts.
Servicing can only be performed by a qualified technician (see chapter “Servicing”). A yearly maintenance is
advised.
Decontamination
The instrument can be cleaned with a soft cloth, wetted with IPA (iso propyl alcohol), Spor-Klenz
®
(STERIS Life
Sciences) or Vesta Syde SQ Disinfectant (STERIS Life Sciences) if required.
Do not spray Spor-Klenz
®
(STERIS Life Sciences) or Vesta Syde SQ Disinfectant (STERIS Life Sciences)
directly on the black heatsink at the back of the instrument.
Do not use strong acid or base or abrasive cleaning agents.
The black seal in the front housing can leave marks on the cloth you’re using when wiping with
cleaning agents. That is a known behavior and not a sign of degradation of the seal.
The disposable is delivered sterile (by gamma irradiation) and should be trashed after used. It is not
autoclavable.
Servicing
Safety precautions
Mains
This instrument can only be used with the dedicated power supply as supplied by Ovizio (see
Specifications chapter at the end of this manual). Replacement by other power supply can
jeopardize safe operation of the instrument.