User Manual
Table Of Contents
1.Safety Information 
3 
Measurement Category   
The multimeter has a safety rating of 1000 V, CAT III.   
Measurement category definition   
Measurement CAT I  applies to  measurements performed on circuits not directly 
connected to the AC mains. Examples are measurements on circuits not derived from the 
AC mains and specially protected (internal) mains- derived circuits. 
Measurement CAT II  applies to protect against transients from  energy-consuming 
equipment supplied from the  fixed installation, such as TVs, PCs, portable tools,  and 
other household circuits. 
Measurement CAT III  applies to protect against  transients in equipment in fixed 
equipment  installations, such as distribution panels,  feeders and short branch circuits, 
and lighting systems in large buildings.   
Measurement CAT IV  applies to measurements performed at the source of the low- 
voltage installation. Examples are electricity meters and measurements on primary over 
current protection devices and ripple control units. 