Operator's Manual
Table Of Contents
- What is new in publication version 7.2?
- 1 Introduction
- 1.1 Before you start
- 1.2 Important safety instructions and additional information
- Important information regarding safety
- Disposal of the system
- Product safety
- General care
- Accessory box
- Meter
- Power off meter
- Automatic power-off
- Shut down meter
- Automatic shutdown
- Battery Pack
- Touchscreen
- Electromagnetic compatibility (EMC)
- Electrostatic discharge (ESD)
- Wireless connectivity
- Radiofrequency radiation exposure information
- Local Area Network: protection from unauthorized access
- Wired network connection
- 1.3 System components
- 1.4 Overview of the meter
- 1.5 Overview of the code key reader
- 1.6 Overview of the Accu-Chek Inform II Base Unit
- 1.7 Overview of the Accu-Chek Inform II Base Unit Hub
- 1.8 Overview of the accessory box
- 1.9 Reagents and consumables
- 1.10 Instructions for initial setup
- 2 Powering Up and Entering an Operator ID
- 3 Patient Glucose Testing
- 3.1 Information regarding blood glucose testing
- 3.2 Performing a patient glucose test
- Overview of test procedure
- Entering or selecting the patient ID
- Entering the patient ID manually
- Selecting the patient ID from a list
- Entering a patient ID with barcode scanner
- Confirming or selecting the test strip lot
- Patient identification information
- Inserting test strips
- Obtaining a blood sample
- Applying a blood sample
- Results screen
- Adding comments
- Add barcode content to a result
- Additional Patient Test
- 4 Glucose Control Testing
- 5 Review Results
- 6 Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
- 7 Linearity Testing
- 8 Proficiency Testing
- 9 Initial Startup
- 9.1 Connecting the base unit
- 9.2 Installing or replacing the battery pack
- 9.3 Docking the meter
- 9.4 Initial setup on the meter
- 9.5 Opening the Setup Menu
- 9.6 Date and time format
- 9.7 Display options and optional tests
- 9.8 Options for test strips
- 9.9 Options for glucose control tests
- 9.10 Value ranges (normal, critical, reportable)
- 9.11 Options for Operator ID Entry
- 9.12 Patient ID options
- 9.13 Creating a setup password
- 9.14 Setting the date and time
- 9.15 Beeper options
- 10 Maintenance and Care
- 11 Troubleshooting
- 12 General Product Information
- A Appendix
- B Appendix
- C Supplement for Other Test Entry
- D Supplement for Observed Test Sequence
- E Supplement for Enhanced Workflows
- F Appendix for Accu-Chek Inform II Base Unit (legacy version)
- G Limited Warranty
- Index
Troubleshooting • 11
175
08424705001 (04) 2021-06 USA • Accu-Chek Inform II Operator’s Manual Version 7.2
Warning Messages
Code Text
W-HBG QC Lockout: Warning! Glucose control is due. Required Controls must be run in order to proceed.
W-HBK Invalid Patient Result: This patient result is invalid due to unacceptable control bar result. Please
repeat test on new card.
W-HBL Invalid Patient Result: The patient result is invalid due to unacceptable control result. Please
repeat test on new card.
W-HBM The entered lot * has expired. Do not use an expired lot.
<This is only for Other Test Entry, and not for blood glucose test strip lots. * is the lot number>
W-HCB Out of Reportable Range: *
<* is the institution defined reportable range>
W-HCC Out of Critical Range: *
<* is the institution defined critical range>
W-HDK Invalid Control Result: This control result is invalid due to unacceptable control bar result. Please
repeat test on new card.
W-NBC Battery Low: The battery is getting low. Return this meter to the base unit for charging as soon as
possible.
W-SBB Download Warning: The meter needs to be downloaded soon. Last download occurred at *.
Please dock meter. If the problem persists, see your System Administrator.
<* is the date and time>
W-SBD Loss of Date/Time: D/T settings lost due to meter reset or loss of power. Before performing further
tests: Enter D/T manually and synchronize meter. OR: Ask System Admin. for help.
W-YBB Unexpected Warning: Unexpected warning. See log file for details. If the problem persists, see
your System Administrator.
Error Messages
Code Text
E-CBC Invalid Operator ID: Operator ID is invalid. ID entered was not found in the list of valid operator
IDs. Please see your System Administrator.
E-CBD Invalid Operator ID: Operator ID length is invalid. ID must be between * characters long.
<* is the minimum and maximum character length defined at the institution >
E-CBF Invalid Operator ID: Operator ID length is invalid. ID must be * characters long.
<* is the minimum character length defined at the institution >
E-CBG Invalid Password: Invalid operator password. Please enter a valid password or see your System
Administrator.
E-CBH Password Format: Invalid operator password format. The operator password must be at least four
characters.