Panasonic* Operating Instructions Instrucciones de Uso Wrist Blood Pressure Monitor Monitor de presión arteriai de muñeca Modei No7№ de Modeio EW-BW10 English Español S1 Before operating this device, piease read these instructions compieteiy and save this manuai for future use. Antes de utilizar este dispositivo, lea completamente estas instrucciones y conserve este manual para futura referencia.
Panasonic Oscillometric Automatic Wrist Blood Pressure Monitor Model EW-BW10 Is a device Intended to measure systolic and diastolic blood pressure and pulse rate of an adult individual by using a pressurized cuff on the wrist. The device Is not Intended for use on neonatal, Infants and children. The device Is designed for home use only, not for ambulatory measurement (measurement recorded continuously during the day). Specifications of this device are listed in page 22.
I Table of Contents Introduction............................................................................................. 3 Basics of Biood Pressure ..................................................................... 3 Important Instructions Before Use....................................................... 4 Precautions to Ensure Safe, Reiiabie Operation ................................ 6 Easiiy Check Your Biood Pressure Readings Against the JNC 7* Ciassification.......................................
I Introduction Thank you for purchasing the Panasonic Automatic Wrist Biood Pressur Monitor EW-BW10. Measuring your own biood pressure is an important way of monitoring your health. High blood pressure (hypertension) is a major health proble which can be treated effectively once detected. Measuring your blood pressure between doctor visits on a regular basis in the comfort of your home, and keeping a record of the measurements, will help you monitor any significant changes in your blood pressure.
I Important Instructions Before Use 2. Do not confuse self-monitoring with self-diagnosis. Blood pressure measurements should only be interpreted by a health care professional who is familiar with your medical history. If you are taking medication, consult with your physician to determin the most appropriate time to measure your blood pressure. NEVER change a prescribed medication without first consulting with your physician. 3.
this is not possible, however, try to take measurements at a specifie time prior to breakfast, and before you have become active. You should relax for about 5 minutes before taking the measurement. The following situations may cause substantial variations in blood pressure readings and should therefore be avoided at least 30 minutes prior to taking your blood pressure.
g. Do not short-circuit batteries. h. Do not attempt to recharge the batteries inciuded with the unit. 12. Piease remove the cuff if abnormai operation, such as proionged overinfiation, is observed.
Easily Check Your Blood Pressure Readings Against the JNC 7* Classification Blood Pressure Categories ^A 0. (mmHg) OQ Q (0 0) £ 0. ■a o o 100 OQ o Stage 2 Hypertension e.g.: SBP165 DBP105 Stage 1 Hypertension e.g.: SBP 150 DBP 95 90 Prehypertension e.g.: SBP 130 DBP 85 80 Normal e.g.: SBP 115 DBP 75 120 140 160 (mmHg) Systolic Blood Pressure (SBP) If the two blood pressure measurements (systolic, diastolic) fall into separate categories, your level is classified in the higher of the two categories.
I Diagram of the Device * Only use alkaline batteries. Body Display This product includes: • Main unit • Storage case • Batteries • Operating Instructions • Guarantee card 'Specifications of the product or parts may be changed without prior notice.
I Inserting/Replacing Batteries *Always use alkaline batteries. (2 AAA-size LR03 alkaline batteries) • When replacing batteries, always use 2 fresh alkaline batteries of the same type from the same manufacturer, and replace them simultaneously. 0 0 1. Remove the device from the storage case. Insert the batteries. (1) Slide the battery cover open as shown In the diagram. (2) Insert the two batteries making sure that they are aligned correctly (+/-). Slide the battery cover shut until it clicks.
For Batteries • The frequency of use for batteries is approximately 300 measurements (3 times per day) with fresh Panasonic alkaline batteries (AAA-size LR0 alkaline batteries). [Measurement conditions: Room temperature 73.4 ° (23 °C); Pressurized to 170 mmHg; Wrist circumference: 6-3/4" (17 cm) • If batteries other than alkaline batteries are used or batteries are used in a cold room, and when measuring a person with high blood pressure battery life may be dramatically shorter. *Use alkaline batteries.
I Fitting the Pressure Cuff (cont.) 2. Firmly wrap and fasten using the hoop and loop fastener. • A loosely applied pressure cuff will result in high blood readings or incapable measurement. • Press fastener softly to avoid the peel off. Do not remove the hoop and loop fastener by twisting it off. Be sure to peel it in an horizontal direction in alignment with the cuff. Fit as shown in the figure. The cuff can be used on either wrist.
I Obtaining Accurate Measurements Always take your blood pressure at the same time while sitting In the same position and using the same wrist. It is recommended to check your blood pressure at least twice a day. It is best In the morning betöre breaklast and In the evening after work. In the following situations, accurate measurements cannot be taken.
I Obtaining Accurate Measurements (cont.) Do not take measurements in the positions shown below as they wili be inaccurate. Do not bend your wrist inward. If the device is placed at a height lower than the heart then the resulting readings may be higher than your actual blood pressure. Do not clench your fist. If the device is placed at a height higher than the heart then the resulting readings may be lower than your actual blood pressure.
I Measuring Your Blood Pressure FiSi Dis; ifisr aaam dgw HR/f1R^SiS8 1. Place the arm to which the device is attached on top of the storage case with the wrist positioned at the level of the heart. (Please see “Obtaining Accurate Measurements” on pages 12 and 13.) 2. Press the Off/On Start button. • The display will be lit up for approx. 2 seconds. • Inflation will then begin automatically. As measurements are being taken during pressurization, operating sounds and the rate of pressurization may change.
I Measuring Your Blood Pressure (cont.) Reading the blood pressure display Blood pressure values for both systolic and diastolic readings are determined as being within normal or hypertensive ranges based on definitions and classification of blood pressure levels by the JNC 7. Whe blood pressure Is In the high range, the reading displayed on screen flashes on and off to alert user. According to JNC 7 Classification, values consistently In excess of 140 mmHg (Sys.) and/or 90 mmHg (DIa.
Advice: In general, blood pressure measurements are lower on the second attempt because you are calmer. For people who easily get tense, measuring twice each time Is advised It Is recommended to store all values when performing two measurements or more. When taking two measurements, always stay quiet for 4 to 5 minutes after the first measurement before taking the second measurement.
I Movement Detection Function If movement is detected while measurement is being carried out and mo pressure is applied by the pressure cuff, the rnark will be displayed. • Please carry out blood pressure measurement again if the mark flashes or is lit. • If the hand or wrist to which the pressure cuff is attached is moved whi measurement is being carried out (e.g. wrist is bent suddenly), in some cases correct measurements may not be obtained.
I Irregular Pulse Function If the pulse rate varies by more than ±25 % from the average while a measurement is being taken, when measurement is completed the i mark will be displayed. • The mark will lit after approximately 6 seconds of blinking when the measurement is completed. • If the LJiiWB mark is lit, please take the measurement again. • In some cases, if the pulse rate fluctuates greatly while a measurement is being taken, it may be impossible to obtain an accurate reading.
I Saving and Recalling Data Saving Readings • All data Is retained even when the batteries go flat or are changed. 1. The M mark will flash after measurement has been completed. m Press the Memory/Recall button. IMI /min. a? OFF; ST/ ¡38I*** BP /min. JO___ 11 19 IJBIHa 2. Saving complete. • Up to 90 sets of readings can be stored In memory and each Is assigned a number. When the limit Is reached, any new readings are recorded over existing readings starting with the oldest ones.
Recalling Stored Data Readings can be recalled after being saved without having to first turn th device on. 1. Press the Memory/Recall button. 1 OFh 1 STA • The average of all recorded readings will be displayed. 'W • The A/g. mark will be displayed. SYS. imH, DIA. imH: • If, for example, 90 sets of readings have been saved In the memory then the average of those 90 sets of readings will be displayed. If only one Pulfe Aig. /min. set of readings has been saved then the mark will not be displayed.
I After Use Storage 1. Fold the end of the pressure cuff in the direction of the arrow. 2. Next, fold the pressure cuff in on itself and secure it on loop fastener. • Place the device in the storage case. .
I Specifications Power source DC 3 V (2 /\/WLR03 size alkaline batteries) Display Digital LCD Method of measurementOscillometric system Measurement range Pressure: 0 mmHg to 280 mmHg Pulse rate: 30 to 160 beats per minute Accuracy Pressure: Within ±3 mmHg Pulse rate: Within ±5 % Operation temperature/ 50 °F to 104 -F (10 °C to 40 °C), 30 % to humidity range 85 % RH Storage temperature/ -4 °F to 140 °F (-20 °C to 60 °C), 10 % to humidity range 95 % RH Measurable wrist Approx. 5" to 8-3/4" (12.5 cm to 22.
I Care and Maintenance Do not attempt to disassemble, repair or modify the unit. • Doing so may cause fire or cause the unit to malfunction. It may also lead to injury. Do not apply excessive force to the unit or drop the unit. • Doing so may cause damage. Do not insert dust or foreign objects into the unit. • Doing so may result in damage. If the unit is stored at temperatures below the freezing point, do not use immediately. Leave it in a warm place for at least an hour before use.
I Troubleshooting Display Status before error Cause and solution The indication disappears in Batteries are exhausted. UOiis the middle of the operation, displayed. (the ■! 3 mark is displayed.) (Please see page 9.) Pressurization to more than Were measurements taken 280 mmHg was performed. according to the proper (Cuff pressurizes several procedure and in the correct times.) posture? (Please see pages 12 and 13.) Pressure decreased suddenly. Is the cuff wrapped properly? displayed.
I Troubleshooting (cont.) Symptom Blood pressure is abnormally high or low. Measurements are different from those taken by a doctor. Measurements are different each time. Possible cause • The posture of the person is different each time measurements are carried out. (Please see pages 12 and 13.) • The person is slightly tense when with the healthcare professional, causing the measurements to be different. • Relax for ten minutes and then carry out the measurement again.
For questions or assistance with your biood pressure monitor, cail us at 1-800-338-0552. Panasonic Corporation of North America One Panasonic Way, 1H-1 Secaucus, NJ 07094 1 -800-338-0552 (USA only) for questions and comments W9030BW10204 4 Revision date: 2013-06-01 B No.4 N0.