Datasheet
17
30
1.181
30
1.181
20
0.787
1.5
0.059
4.5
0.177
ø8.3
ø0.327
30
1.181
20
0.787
25.5
1.004
M5 female thread,
4 0.157 deep
M3 female thread,
4 0.157 deep
Connector
PRECAUTIONS FOR PROPER USE
Flat attachment MS-DP1-F□
• Make sure to mount MS-DP1-F□ with the sensor properly.
If it is not mounted properly, air leakage may occur.
• Take care that the excessive mounting and dismounting
of this product may cause deterioration of the O-ring.
• If you touch the O-ring of MS-DP1-F□, or any scratch or
dust, etc. is attached to it, air leakage may occur and the
sensing performance may deteriorate.
Take sufcient care when using and storing MS-DP1-F□.
• If connecting a commercially-available joint to the
pressure port of the sensor, hold the main unit in your
hand to steady it, and tighten to a torque of 1 N·m or less.
If it is tightened to an excessive torque, the joint or the
main unit may become damaged.
• If connecting a commercially-available joint to the
pressure port of the MS-DP1-7 conversion bushing,
tighten to a torque of 9.8 N·m or less.
Piping
14 mm 0.551 in spanner
• The tightening torque should be 1 N·m or less when
connecting a coupling to the pressure port of MS-DP1-FM
at attachment.
Coupling (Purchase separately.)
Pressure port
• When connecting the coupling to the pressure port of
MS-DP1-FR/FE/FN at attachment, hold the pressure
port with a 14 mm 0.551 in spanner and make sure that
the tightening torque is 9.8 N·m or less.
In addition, in order to prevent any leakage, wind a
sealing tape on the coupling when connecting.
14 mm 0.551 in spanner
Note: Do not tighten the pressure port by holding the product with the spanner.
It may cause the product breakage.
Others
• This product has been developed / produced for industrial
use only.
• Use within the rated pressure range.
• Do not apply pressure exceeding the pressure
withstandability value. The diaphragm will get damaged
and correct operation shall not be maintained.
• Do not use during the initial transient time (0.5 sec.
approx.) after the power supply is switched on.
• Avoid dust, dirt, and steam.
• Take care that the sensor does not come in direct contact
with water, oil, grease, or organic solvents, such as,
thinner, etc.
• Do not insert wires, etc., into the pressure port. The
diaphragm will get damaged and correct operation shall
not be maintained.
• Do not operate the keys with pointed or sharp objects.
DP-10□ZL3-M-P(-C)
Sensor
DIMENSIONS (Unit: mm in)
The CAD data can be downloaded from the website.
Dual Display Digital Pressure Sensor DP-100L SERIES
DP-100L
SERIES