User Manual

7
Standard IEC61010-1
CAT III-600V
Pollution degree II, CE Approved
6. Operation
AC/DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure AC/ DC voltages if a motor on the
circuit is being switched ON or OFF. Large voltage surges may
occur that can damage the meter.
1. Insert the black test lead into the negative COM terminal and
the red test lead into the positive V terminal.
2. Set the function switch to VAC or VDC position.
3.
Connect the test leads in parallel to the circuit under test.
4.
Read the voltage measurement on the LCD display.
AC/DC CURRENT MEASUREMENTS
1. Set the function switch to the µA/mA position.
2. Insert the black test lead into the negative COM terminal and
the red test lead into the positive µA/mA terminal.
3. For current measurements up to 2000µA DC/AC, set the
function switch to the mA position
4. Press the MODE button to indicate “DC” / “AC” on the
display.
5. Remove power from the circuit under test, then open up the
circuit at the point where you wish to measure current.
6. Touch the black test probe tip to the negative side of the
circuit.
Touch the red test probe tip to the positive side of the circuit.
7. Apply power to the circuit.
8. Read the current in the display