9
Back side
4838 Reactor Controller 230V
Thermocouple burnout protection is included in 
each of these controllers. If the thermocouple 
should burn out, or if there is a break in the thermo-
couple circuit, power to the heater will be shut off. 
In most cases this will cause the controller to show 
a “NO CONN” reading. A similar high reading will 
be displayed if the controller is turned on without an 
attached thermocouple. In either of these situations, 
the high limit reset will activate. The thermocouple 
break must then be corrected and the High Limit Re-
set must be reset in order to activate the system. 
High Limit Reset
Each Series 4838 Controller has a high temperature 
limit control which will trip the High Limit Relay 
and shut off the heater if the temperature reaches 
a preset level. If tripped, the lockout relay will pop 
out, thereby preventing the system from reheating 
the reaction when the temperature falls back below 
the limit setting. This must be manually reset before 
heating can occur again.
On Model 4838 Controllers the high temperature 
limit is preset at the factory to 375 °C for most sys-
tems. On these controllers an alarm condition is 
indicated by an illuminated red indicator labeled as 
"ALM". If the temperature exceeds the limit, the con-
troller alarm condition will appear, the High Limit 
Reset will be tripped, and power to the heater will 
be shut off. 
Access to the Controller Cabinet
To open the controller cabinet, unplug the power 
cord and remove the two screws in the upper 
corners of the back panel. Then lift the top plate 
upwards, allowing the hinged top and front panel to 
swing forward, providing full access to all internal 
components.
Caution!
Inside of controller cabinet should be ac-
cessed by qualifi ed personnel only.
Fuses
The 4838 Controller has a number of fuses which 
are intended to be fi eld serviceable. 
Fuse Rating Part number
Instrument 
Fuse
Fast acting, 1 
Amp, 250VAC
139E6
Line Fuse(s) Fast acting, 20 
Amp, 250VAC
2074E
Caution!
Unplug unit before servicing. For continued 
protection against possible hazard, replace 
fuses with same type and rating of fuse.