User Manual
4848B Reactor Controller
Parr Instrument Company
12
the arrow keys to change the pattern number 
and press the set key to solidify the change.
5.  Press the return key once. The upper display will 
read [
SP00
], where the “00” represents the 
pattern and step number, respectively. The lower 
display will indicate the set point for indicated 
step. The set point can be adjusted using the 
arrow keys then press the set key to solidify the 
change.
Note: The fi rst SP_O used must be a higher 
temperature than the current temperature 
(PV) or the program will not execute properly.
6.  Press the return key once. The upper display will 
read [
TI00
], where the “00” represents the 
pattern number and the step number, respec-
tively. The lower display will show the time to 
execute the current set point in “hh:mm” format. 
The time setting can be changed by using the 
arrow keys then press the set key to solidify the 
change.
7.   Repeat steps 4 and 5 for “SP01” through “SP07” 
and “tI01” through “tI07” as required. Once you 
have entered all the steps you want, you can 
leave the remaining “SP” and “tI” variables at 
zero and press the return key to advance. 
8.  Press the return key to display [
PSY0
] on the 
upper display, where “0” represents the current 
pattern number. This parameter determines the 
last step to execute in the pattern. For example, 
if PSy0 has a value of 3, Steps SP00 through 
SP03 will execute. Use the arrow keys to change 
the setting and press the set key to solidify the 
change.
9.  Press the return key to display [
CYC0
], where 
“0” represents the pattern number. This pa-
rameter defi nes how many times you want the 
indicated pattern to repeat. If you would like the 
pattern to only execute once, set this value to 
zero. Use the arrow keys to change the setting 
and press the set key to solidify the change.
10. Press the return key to display [
LIN0
], where 
“0” represents the pattern number. This pa-
rameter allows you to link the indicated pat-
tern another. If you would like more than seven 
separate steps in the process, set this to the next 
pattern number you wish to link to. Use the ar-
row keys to change the setting and press the set 
key to solidify the change.
11.  Press set to return to the main screen. Note 
that the lower display of the main screen has 
changed to show the program step.
Note: The program will attempt to hold the 
temperature at the set point determined by 
the last step executed. So, if you wish for the 
reactor cool to room temperature once the 
program has fi nished, the last executed step 
must have a set point consistent with room 
temperature and the time for the last step 
can be set to 00:00.
Engaging a Program
1.  From the main screen, press the return key two 
times to display [
PTRN
] on the upper display. The 
lower display will indicate the pattern at which you 
wish to execute.
2.  Press the return key until the run/stop parameter 
[
R-S
] is on the upper display. The lower display 
will indicate the setting of the run/stop parameter. 
Change this value to [
RUN
] in order to start the 
program determined in step 1. Press set to return to 
the main screen. 
If you wish to stop the program completely, you 
may change this value to [
STOP
]. Changing the 
run/stop mode back to stop will restart the program 
from the beginning. 
If you wish to stop the program and continue con-
trolling the temperature at the current set point, you 
may change this value to program stop [
PSTP
]. 
Changing back to run mode from a program stop 
will restart the program from the beginning. 
If you wish to hold the program at the current 
temperature, you may change this to program hold 
[
PHOD
]. Changing back to run mode from a pro-
gram hold will continue the program from its cur-
rent step.
3.  The main screen will now show the current set 
point on the lower display. Pressing the arrows will 
allow you to change between: [
P-ST
] to display 
the current pattern and step, [
R-TC
] to display the 
remaining time in the current pattern step, or [
SP
] 
to display the set point for the current pattern step. 
Press set will solidify this display change.