Operating Instruction Manual 9 No.
6200 Calorimeter Operating Instruction Manual TABLE OF CONTENTS Preface I Scope ............................................................................................................ I Explanation of Symbols................................................................................. II Safety Information ......................................................................................... II Intended Usage......................................................................................
6200 Calorimeter Operating Instruction Manual TABLE OF CONTENTS Operating Instructions 5-1 Operating the 1108 Oxygen Bomb.............................................................. 5-1 Operating the Filling Connection................................................................. 5-1 Operating the Calorimeter........................................................................... 5-2 Samples and Sample Holders.....................................................................
6200 Calorimeter Operating Instruction Manual TABLE OF CONTENTS Appendix B – Calculations B-1 Calculating the Heat of Combustion............................................................ B-1 General Calculations................................................................................... B-1 Thermochemical Corrections ...................................................................... B-2 ASTM and ISO Methods Differ....................................................................
6200 Calorimeter Operating Instruction Manual TABLE OF CONTENTS Figures 2-1 2-2 2-3 2-4 2-5 4-1 5-7 F-1 F-2 F-3 F-4 F-5 F-6 F-7 F-8 F-9 6200 Calorimeter Back Panel ........................................................... 2-4 Closed Loop Configuration with 6510 ............................................... 2-5 Closed Loop Configuration with 1563 ............................................... 2-5 Open Loop Configuration with Tap Water.........................................
200 Calorimeter Operating Instruction Manual PREFACE This manual contains instructions for installing and operating the Parr 6200 Calorimeter. For ease of use, the manual is divided into nine chapters. Concept of Operation Installation Instrument Description No.
6200 Calorimeter Operating Instruction Manual PREFACE Explanation of Symbols I On position O Off position ~ Alternating Current (AC) This CAUTION symbol may be present on the Product Instrumentation and literature. If present on the product, the user must consult the appropriate part of the accompanying product literature for more information. ATTENTION, Electrostatic Discharge (ESD) hazards. Observe precautions for handling electrostatic sensitive devices. Protective Earth (PE) terminal.
6200 Calorimeter Operating Instruction Manual PREFACE Safety Information To avoid electrical shock, always: 1. Use a properly grounded electrical outlet of correct voltage and current handling capability. 2. Ensure that the equipment is connected to electrical service according to local national electrical codes. Failure to properly connect may create a fire or shock hazard. 3. For continued protection against possible hazard, replace fuses with same type and rating of fuse. 4.
6200 Calorimeter Operating Instruction Manual PREFACE Getting Started These steps are offered to help the user become familiar with, install, operate and develop the full capabilities of the Parr 6200 Calorimeter. 1. Review the Concept of Operations, Chapter 1, to get an understanding of the overall capabilities of the calorimeter and microprocessor control. 2. Unpack and install the calorimeter in accordance with the Installation Instructions, Chapter 2.
6200 Calorimeter Operating Instruction Manual 1 CONCEPT OF OPERATION The 6200 Calorimeter has been designed to provide the user with: • A traditional design calorimeter with removable oxygen bomb and bucket. • A moderately priced calorimeter which uses real time temperature measurements to determine heat leaks using a controlled calorimeter jacket. • A high precision calorimeter capable of exceeding the repeatability and reproducibility requirements of all international standard test methods.
6200 Calorimeter Operating Instruction Manual 1 Dynamic Operation In its Dynamic Operating Mode, the calorimeter uses a sophisticated curve matching technique to compare the temperature rise with a known thermal curve to extrapolate the final temperature rise without actually waiting for it to develop. Full Microprocessor Based Process Control A. B. C. D. E.
6200 Calorimeter Operating Instruction Manual 2 INSTALLATION The 6200 Calorimeter is completely assembled and given a thorough test before it is shipped from the factory. If the user follows these instructions, installation of the calorimeter should be completed with little or no difficulty. If the factory settings are not disturbed, only minor adjustments will be needed to adapt the calorimeter to operating conditions in the user’s laboratory. This apparatus is to be used indoors.
6200 Calorimeter Operating Instruction Manual 2 Connect the cooling water outlet on the back of the calorimeter to a drain using either nylon (HJ0025TB035) or copper ¼” tubing. A shut off valve in tap water supply line is also a good idea if the calorimeter will not be used for an extended period. Jacket Cooling Water Connection Continued Oxygen The 6200 Calorimeter is Filling equipped with an automatic bomb Connection oxygen filling system.
6200 Calorimeter Operating Instruction Manual 2 1. 2. 3. 4. When Swagelok Tube Fittings are used, the instructions for installation are: Simply insert the tubing into the Swagelok Tube Fitting. Make sure that the tubing rests firmly on the shoulder of the fitting and that the nut is finger-tight. Before tightening the Swagelok nut, scribe the nut at the 6 o’clock position. While holding the fitting body steady with a back-up wrench, tighten the nut 1-1/4 turns.
6200 Calorimeter Operating Instruction Manual 2 Figure 2-1 6200 Calorimeter Back Panel 2-4 16
6200 Calorimeter Operating Instruction Manual 2 Line 1 & 2 – Maximum length of 10 feet -¼” OD, Polyurethane (Part Number HJ0025TB035) Line 3 - Maximum length of 25 feet - 1/8” OD, Nylon (Part Number HX0012TB024) 6200 Calorimeter External Plumbing Closed Loop Configurations Figure 2-2 Figure 2-3 2-5 17
6200 Calorimeter Operating Instruction Manual 2 Open Loop Configurations Figure 2-4 Figure 2-5 2-6 18
6200 Calorimeter Operating Instruction Manual 2 The calorimeter must be accurately standardized prior to actually performing calorimetric tests on sample materials. Review Appendix C - Standardization, in order to become familiar with the general procedure and calculations. The user should configure the calorimeter at this time to accommodate the desired sample weight entry mode. The calorimeter can be placed into standardization mode on the Calorimeter Operation Page, with the operating mode key.
6200 Calorimeter Operating Instruction Manual 2 This page intentionally left blank.
6200 Calorimeter Operating Instruction Manual 3 INSTRUMENT DESCRIPTION All calorimeter configurations and operations are handled by a menu-driven system operated from the bright touch screen display. The settings and controls are organized into nine main sections or pages which comprise the MAIN MENU. Types of Controls Note: Keys with a “double box” in the upper left hand corner lead to sub-menus. The controls that change the data field information in the menus will be one of the following: 1. Toggles.
6200 Calorimeter Operating Instruction Manual 3 This page intentionally left blank.
6200 Calorimeter Operating Instruction Manual 4 PROGRAM INSTALLATION & CONTROL Software The program in the 6200 Installation Calorimeter can be extensively modified to tailor the unit to a wide variety of operating conditions, reporting units, laboratory techniques, available accessories and communication modes. In addition, the calculations, thermochemical corrections and reporting modes can be modified to conform to a number of standard test methods and procedures.
6200 Calorimeter Operating Instruction Manual 4 The default parameters of the 6200 Calorimeter can be changed to guarantee that the 6200 Calorimeter, when cold restarted, will always be in the desired configuration before beginning a series of tests. Users who wish to permanently revise their default settings may do so using the following procedure: • 4-2 24 Establish the operating parameters to be stored as the user default settings.
6200 Calorimeter Operating Instruction Manual 4 Factory Default Settings Figure 4- 1 6200 Factory Default Settings Calorimeter Operation Operating Mode Determination Bomb Installed/EE 1/2400.0 Heater and Pump OFF Operating Controls Method of Operation Reporting Units Use Spiking Correction “OTHER” Multiplier Calibrate Touchscreen LCD Backlight Timeout(s) LCD Contrast Print Error Messages Language Dynamic Btu/lb OFF 4.
6200 Calorimeter Operating Instruction Manual 4 Data Entry Controls Prompt for Bomb ID ON Weight Entry Mode Touch Screen Acid Entry Mode Touch Screen Hydrogen Entry Mode Touch Screen Auto Sample ID Controls ON Sample Weight Warning above 2.
6200 Calorimeter Operating Instruction Manual 5 OPERATING INSTRUCTIONS Operating the Detailed Instructions for 1108 Oxygen preparing the sample and charging Bomb the 1108 Oxygen Bomb are given in Operating Instructions No. Operating the Filling Connection To fill the bomb, connect the hose to the bomb inlet valve and push the O2 button on the calorimeter control panel.
6200 Calorimeter Operating Instruction Manual 5 All operations required to standardize the 6200 Calorimeter, or test an unknown sample, should proceed step-wise in the following manner: 1. Turn on the calorimeter and activate the pump and heater using Calorimeter Operations. Allow at least 20 minutes for the calorimeter to warm up and the jacket temperature to stabilize. Once the jacket temperature comes within 0.
6200 Calorimeter Operating Instruction Manual 5 Operating the Calorimeter Continued 4. Set the bucket in the calorimeter. Attach the lifting handle (421A) to the two holes in the side of the screw cap and partially lower the bomb in the water. Handle the bomb carefully during this operation so that the sample will not be disturbed. Push the two ignition lead wires into the terminal sockets on the bomb head.
6200 Calorimeter Operating Instruction Manual 5 10. Remove all unburned pieces of fuse wire from the bomb electrodes; straighten them and measure their combined length in centimeters. Subtract this length from the initial length of 10 centimeters and multiply this burned length by 2.3 calories per cm (for Parr 45C10 Fuse Wire) to obtain the fuse correction. The scale on the fuse wire card can be used to obtain this value directly. 11.
6200 Calorimeter Operating Instruction Manual 5 Samples and Sample Particle Size and Moisture Holders Content. Solid samples burn best in an oxygen bomb when reduced to 60 mesh, or smaller, and compressed into a pellet with a 2811 Parr Pellet Press. Large particles may not burn completely and small particles are easily swept out of the capsule by turbulent gases during rapid combustion. Note: Particle size is important because it influences the reaction rate.
6200 Calorimeter Operating Instruction Manual 5 Non-volatile samples to be tested in Parr oxygen bombs are weighed and burned in shallow capsules measuring approximately 1" diameter and 7/16" deep. These are available in stainless steel, fused silica and platinum alloyed with 3-1/2% rhodium. Stainless steel capsules (43AS) are furnished with each calorimeter. When combusting samples that contain metal particles such as aluminum or magnesium, the non-metallic (fused silica) 43A3 Capsule is required.
6200 Calorimeter Operating Instruction Manual 5 Corrosive The 1108 Oxygen Bomb is made Samples of a corrosion resistant alloy designed to withstand the corrosive mixture of sulfuric and nitric acids produced in normal fuel testing operations. Samples containing chlorine and particular samples containing more than 20 mg of chlorine samples with high sulfur contents will greatly accelerate corrosion of the bomb.
6200 Calorimeter Operating Instruction Manual 5 Use the following procedure when filling and handling any of these tape-sealed sample holders: • Weigh the empty cup or capsule; then cover the top with tape, trim with a knife and press the trimmed edge firmly against the metal rim. Also cut and attach a small flag to the disc (see Figure 5-7). • Puncture the tape at a point below the flag, then re-weigh the empty cup with its tape cover.
6200 Calorimeter Operating Instruction Manual 6 CORRECTIONS & FINAL REPORTS Entering Corrections and Obtaining the Final Report Final reports for each test can be obtained whenever the operator is prepared to enter any required corrections for fuse, acid and sulfur. When entering corrections, the user can choose either of two methods.
6200 Calorimeter Operating Instruction Manual 6 This page intentionally left blank.
6200 Calorimeter Operating Instruction Manual 7 REPORTING INSTRUCTIONS The 6200 Calorimeter can transmit its stored test data in either of two ways. The REPORT DESTINATION key on the Reporting Controls Page toggles the report destination between the display and an There are two kinds of calorimeter reports: Preliminary and Final. Preliminary Reports are generated at the conclusion of a test. They will not contain the thermochemical corrections for sulfur, fuse, or acid.
6200 Calorimeter Operating Instruction Manual 7 Report Generation Continued The displayed files can be sorted by sample ID number, by type, by status or by date of test by simply touching the appropriate column. Individual files can be chosen by highlighting them using the up and down arrow keys to move the cursor. Press the SELECT key to actually enter the selection. Once selected the highlight will turn from dark blue to light blue.
6200 Calorimeter Operating Instruction Manual 8 FILE MANAGEMENT The 6200 Calorimeter will hold data for 1000 tests in its memory. These tests may be pre weights, preliminary or final reports for either standardization or determination runs. Once the memory of the controller is filled, the controller will not start a new analysis until the user clears some of the memory. The FILE MANAGEMENT key on the main menu leads to the file management sub-menu.
6200 Calorimeter Operating Instruction Manual 8 This page intentionally left blank.
6200 Calorimeter Operating Instruction Manual 9 MAINTENANCE & TROUBLESHOOTING Oxygen Bomb Under normal usage the 1108 Parr Oxygen Bomb will give long service if handled with reasonable care. However, the user must remember these bombs are continually subjected to high temperatures and pressures that apply heavy stresses to the sealing mechanism. The mechanical condition of the bomb must therefore be watched carefully and any signs of weakness or deterioration should be replaced before they fail.
6200 Calorimeter Operating Instruction Manual 9 Fuses The replacement of protective fuses for the 6200 Calorimeter should be performed by qualified personnel. All fuses except Parr part # 139E23 are located on the A1794E I/O board located inside the instrument. To gain access to the inside, first disconnect the power cord from the rear of the unit. Remove the two screws at the upper rear of the unit that secure the front cover. Pry up on the back lip of the cover and carefully hinge it forward.
6200 Calorimeter Operating Instruction Manual 9 A / D Initialization Failed. Shortly after power is applied to the calorimeter controller and the operating system has started, the CPU attempts to read the unique IO board calibration information from the IO board. If the IO board is not connected to the CPU, or the information on the board is not valid, this error will be issued. Bomb ID – Has Been Fired – Times Which Exceeds The Bomb Service Interval.
6200 Calorimeter Operating Instruction Manual 9 This page intentionally left blank.
6200 Calorimeter Operating Instruction Manual APPENDIX A MENU OPERATING INSTRUCTIONS The settings and controls are organized into nine main sections or pages which comprise the Main Menu. The calorimeter will normally be Calorimeter operated from the Calorimeter Operation Operation Page, although tests can Menu always be started from any menu page. This appendix describes all pages of the menu-based operating system of the 6200 Calorimeter.
6200 Calorimeter Operating Instruction Manual APPENDIX A Setup. Press this key to access the Temperature Plot Setup Menu, which has many keys that permit the user to fully customize both the x (time) axis and the scaling of the y axis. Bucket Plot Symbol. Toggles between: • • • • • • • Enable Bucket. Toggles ON/OFF. Bucket Autoscale. Toggles ON/OFF. Enable Jacket. Toggles ON/OFF. Temperature vs.
6200 Calorimeter Operating Instruction Manual APPENDIX A Temperature Plot Setup Menu Continued Time Units. Toggles between minutes and seconds. Bucket Plot Color. Toggles between: • Red • Green • Yellow • Blue • Magenta • Cyan • White • Black Bucket Max Value. Press this key to access its numeric dialog box to set a maximum bucket value. Jacket Plot Color: (same as Bucket Plot Color, above). Jacket Max Value. Press this key to access its numeric dialog box to set a maximum jacket value. Time Minimum.
6200 Calorimeter Operating Instruction Manual APPENDIX A Spiking Correction. Accesses sub-menu, Operating Spiking Controls. Spiking is the material addition, such as benzoic acid or mineral oil, Controls Menu to samples which are difficult to burn in order to drive the combustion to completion. Use Spiking. When set to ON, the calorimeter will prompt for the weight of the spike added and will compensate for the heat of combustion in the calculations. Heat of Combustion of Spike.
6200 Calorimeter Operating Instruction Manual APPENDIX A Operating Controls Menu Continued Calibrate Touchscreen. This key prompts the user to touch the screen at predefined points in order to facilitate touchscreen calibration. LCD Backlight Time-out. The unit is equipped with an automatic circuit to shut-off the backlight when it is not being used. The back light will shut-off if there is no keyboard activity for the number of seconds entered.
6200 Calorimeter Operating Instruction Manual APPENDIX A Save User Default Settings. Used to record the setup to the memory once the user has configured the instrument to their operating requirements. Program available for the calorimeter. When the user chooses a bomb, the instrument Information must be re-booted to load the correct and Control version of the software.
6200 Calorimeter Operating Instruction Manual APPENDIX A EE Max Std Deviation. Displays the maximum relative standard deviation in percent that will be permitted for any EE value calculated by the calorimeter and accesses the submenu on which this limit is set. If this value is exceeded, the user will be warned to take corrective action before proceeding with testing. This calorimeter is capable of achieving a value of 0.15 or better for 15 tests. A setting of zero disables this check.
6200 Calorimeter Operating Instruction Manual APPENDIX A Print Standardization Runs. Will print all of the tests that have been incorporated into the calculated EE value. This will be helpful in evaluating a series of tests which fail to produce a satisfactory EE value and relative standard deviation. Reset Bomb Fire Count. After bomb service, press this button to reset the fire count to zero Standardization Correction Bomb 2. Accesses sub-menu, Bomb 2. Provides the same controls as described for Bomb 1.
6200 Calorimeter Operating Instruction Manual APPENDIX A Thermochemical Corrections Menu Continued Calculation Factors. Accesses submenu, Calculation Factors which sets a number of options for the way the thermochemical corrections are applied. Fuse Multiplier. The fuse corrections represent the number of calories liberated by the burning thread used to ignite the sample. If another measurement is used, the correction factor must be entered here. Use Offset Correction (ISO).
6200 Calorimeter Operating Instruction Manual APPENDIX A Auto Sample ID Increment. Establishes the increment between sample numbers. Data Entry Controls Menu Next Auto Sample ID Number. Establishes the initial sample number for a series of tests and then shows the next sample ID which will be assigned. Prompt for Bomb ID. In the ON position the controller will prompt for a Bomb ID(1-4) when a test is started. Weight Entry Mode.
6200 Calorimeter Operating Instruction Manual APPENDIX A Data Entry Controls Menu Continued Auto Preweigh ID Controls. Accesses sub-menu, used to automatically assign Sample ID numbers when a series of samples are preweighed ahead of the time they are actually tested. Automatic Preweigh ID. ON/OFF toggle for this feature. Automatic Preweigh ID Prefix. An entry here will be used as a prefix for all pre-weigh sample IDs. Next Automatic Preweigh ID Number.
6200 Calorimeter Operating Instruction Manual APPENDIX A Number of Data Bits. Standard options for data bits. Toggles between 7 and 8. Communication Controls Menu Parity. Standard options for parity. Choose from None, Odd or Even. Number of Stop Bits. Standard options for stop bits. Toggles between 1 and 2. Handshaking. Standard options for handshaking. Choose from Xon/Xoff, RTS/CTS and None. Communication Controls. Accesses submenus which set the communications protocols for the printer and balances.
6200 Calorimeter Operating Instruction Manual APPENDIX A Communication Controls Menu Continued Balance Port (RS232). Accesses sub-menu, Balance Port Communications. Balance Type. This key toggles through the available balance templates. Customize Balance Settings. This key leads to a submenu that allows the balance template and related port frame parameters to be altered as required. Baud Rate. Standard options for baud rate. Choose from 19.2K , 9600, 4800, 2400, 1800, 1200, 600, 300, 150, 134.
6200 Calorimeter Operating Instruction Manual APPENDIX A Run Data File Manager - This key activates the File Manager. The File Manager is used to delete or rename test report files. It is also used to convert file types. See the Report Generation section in chapter 7. Format the CompactFlash – This key provides access to a function that will format the user installed CF card in a manner that is compatible with the CPU Boot loader.
6200 Calorimeter Operating Instruction Manual APPENDIX A Diagnostics Menu Allows the user to test many of the components and subsystems of the calorimeter. These capabilities must be used in conjunction with the Maintenance Instructions to obtain the maximum benefits from these capabilities. View System Log - This key is used to display the contents of /flash/log/messages. This file is used primarily to log application program debug messages.
6200 Calorimeter Operating Instruction Manual APPENDIX A This page intentionally left blank.
6200 Calorimeter Operating Instruction Manual APPENDIX B CALCULATIONS Calculating the Heat of Combustion The 6200 Calorimeter will automatically make all of the calculations necessary to produce a gross heat of combustion for the sample. However, it is important that the user understand these calculations to ensure the instrument is set up so the calculations match the procedures and the units are consistent throughout the process.
6200 Calorimeter Operating Instruction Manual APPENDIX B Nitric Acid Correction. In the high pressure oxygen environment within the oxygen bomb, nitrogen that was present as part of the air trapped in the bomb is burned to nitric oxide which combines with water vapor to form nitric acid. All of this heat is artificial since it is not a result of the sample burning. The nitric acid correction removes this excess heat from the calculation. Thermochemical Corrections Sulfur Correction.
00 Calorimeter Operating Instruction Manual APPENDIX B Current ASTM, ISO, and British Standard Methods differ on their treatment of the nitric and sulfuric acid thermochemical corrections. ASTM Methods call for titrating the bomb washings to determine the total acid present. This is assumed to be all nitric acid with a heat of combustion of -14.1 Kcal per mole. The amount of sulfur is then determined and converted to equivalents of sulfuric acid.
6200 Calorimeter Operating Instruction Manual APPENDIX B Acid and Sulfur Corrections • • • • • • • • • • Total acid is the amount of base required to titrate the bomb washings (milliliters). Nitric acid is that portion of the total acid in the bomb washings that result when the nitrogen in the air that is trapped in the bomb is burned at high pressure.
6200 Calorimeter Operating Instruction Manual APPENDIX B Users may find it convenient to enter a fixed value for the acid correction and avoid the need to determine this correction for each test. Use of a fixed value for the acid correction is highly recommended. Fixed acid corrections can be entered when Fixed Acid - Thermochemical Corrections, is set to ON. A correction of 10 calories is a good number for the fixed nitric acid value.
6200 Calorimeter Operating Instruction Manual APPENDIX B In the ASTM treatment, the correction for acid formation assumes that all the acid titrated is nitric acid. Obviously, if sulfur is present in the sample, which in turn produces sulfuric acid, part of the correction for the sulfuric acid formed is already included in the ASTM nitric acid correction (e1). This is adjusted by a separate computation based upon the sulfur content of the sample. An additional correction of 1.
6200 Calorimeter Operating Instruction Manual APPENDIX B It is sometimes necessary to add a spiking material to samples which are very small, have a low heat of combustion, or have a high moisture content to add sufficient heat to drive the combustion to completion. Benzoic acid is an excellent material for spiking for all of the same reasons it is a good standard material. White oil is also an excellent material, particularly for liquid samples.
6200 Calorimeter Operating Instruction Manual APPENDIX B This page intentionally left blank.
6200 Calorimeter Operating Instruction Manual STANDARDIZATION Standardizing the The Energy Equivalent Factor. Calorimeter The term “standardization”, as used here, denotes the operation of the calorimeter on a standard sample from which the energy equivalent or effective heat capacity of the system can be determined. The energy equivalent, W or EE of the calorimeter, is the energy required to raise the temperature one degree, usually expressed as calories per degree Celsius.
6200 Calorimeter Operating Instruction Manual APPENDIX C The 6200 Calorimeter includes a provision for calculating and using a mean energy equivalent for each of up to 4 separate bomb and bucket combinations. ASTM procedures recommend that the energy equivalent be determined by averaging ten tests. The 6200 Calorimeter automatically determines and uses up to 10 tests in its memory and will update the EE Value as additional standardizations are run.
6200 Calorimeter Operating Instruction Manual APPENDIX D Printer Port COMMUNICATION INTERFACES RS232C Connections The 6200 Calorimeter is also equipped with an RS232C port for connection to either a 40 or 80 column printer and/or a balance. Before making either of these connections the data transmission rate of the calorimeter and the printer or balance must be matched. Generally the baud rates on either device can The 6200 Calorimeter supports input from multiple balance types indicated below.
6200 Calorimeter Operating Instruction Manual APPENDIX D Field polarity space data space stability CR LF Length 1 1 8 1 2 1 1 The polarity field must contain either a “+” or a space. Leading zeros in the data field are blanked, except for the one to the left of the decimal point. The stability field must contain “g_” for the calorimeter to accept a mass.
6200 Calorimeter Operating Instruction Manual APPENDIX D Generic Interface Field data CR Length 9 1 The data field should consist of 9 numeric characters (0 through 9, +, - and space) terminated with a carriage return (CR). Leading zeros may be blanked as spaces and are counted. Non-numeric characters are ignored and will reset the input buffer if the data field has not been filled. Any characters received after filling the data field and before the carriage return are ignored.
6200 Calorimeter Operating Instruction Manual APPENDIX D Test data files are named with the following convention. Test type Preliminary Standardization Final Standardization Preliminary Determination Final Determination Pre-weigh Filename .std.plim.csv .std.finl.csv .det.plim.csv .det.finl.csv .---.pwgh.
6200 Calorimeter Operating Instruction Manual APPENDIX D Ethernet Interface Calorimeter test data can be transferred to an Ethernet network connected computer using the FTP File Transfer Protocol. First, you must know the IP address of the network-connected calorimeter. The network DHCP (Dynamic Host Configuration Protocol) server provides this address shortly after the calorimeter is turned on. The address can be seen on the “software and hardware info” page, under “program information and control”.
6200 Calorimeter Operating Instruction Manual APPENDIX D The following screenshot illustrates the contents of the calorimeter data directory as presented by a web browser. You can drag and drop or copy and paste test data files (with the csv suffix) from the web browser window to any convenient folder or directory on the PC.
6200 Calorimeter Operating Instruction Manual APPENDIX D The calorimeter offers a web server service. Test reports can be viewed with a web browser using a URL of the following form. http://192.168.0.125 Where 192.168.0.125 is the IP address of the calorimeter. The following screenshot illustrates the calorimeter home page.
6200 Calorimeter Operating Instruction Manual APPENDIX D Clicking on the Sample Data tab displays a list of reports currently in the instrument memory.
6200 Calorimeter Operating Instruction Manual APPENDIX D Clicking on any given report will provide a display as follows.
6200 Calorimeter Operating Instruction Manual APPENDIX D The use of barcodes in the laboratory has become a highly accurate, rapid and inexpensive way to identify samples. When purchasing this feature, the user must supply Parr with the MAC address of the calorimeter (found in the Software & Hardware Info menu screen). This allows Parr to activate the feature key. In order to enable the calorimeter to use the bar code feature, the feature key needs to be entered into the instrument.
6200 Calorimeter Operating Instruction Manual APPENDIX E Should you need assistance in the operation or service of your instrument, please contact the Technical Service Department. Telephone: (309) 762-7716 Toll Free: 1-800-872-7720 Fax: (309) 762-9453 Email: parr@parrinst.com Any correspondence must include the following basic information: 1. 2. 3. 4. The model and serial # of the instrument. Date purchased. Software version(s) shown on the “Software and Hardware Information” page. Help system revision.
6200 Calorimeter Operating Instruction Manual APPENDIX E This page intentionally left blank.
6200 Calorimeter Operating Instruction Manual APPENDIX F 6200 CALORIMETER PARTS LIST Principal Assemblies in Calorimeter Item Description Oxygen Combustion Vessel Oval Bucket Regulator Assembly, Oxygen Controller Assembly Pump Assembly, Circulating, 115V Pump Assembly, Circulating, 230V Motor Assembly, Pump, 12V Cold Water Solenoid Stirrer Hub Assembly Lead Wire Oxygen Solenoid, w/o fittings Power Supply Capacitor, 40V, 81000 uF Lid Seal Thermistor Bucket Male Connector 1/8 NPTM-T-BT Nylon Gas Spring Fuse
6200 Calorimeter Operating Instruction Manual APPENDIX F Parts List for Temperature Control Assembly Item 1281DD 1417E 538VB 252HWHJ 280HWHJ A1275DDEB A1275DDEE 535VB 337VB 283VB A1276DD 117HW3 Description Manifold, Temperature Control Thermistor, Jacket Male Connector, 1/8 NPTM-T-BT Nylon Elbow, Hose Barb, 1/2 x 3/8 M Elbow, Hose Barb, 1/4 T X3/8 M Cartridge, Heater Assembly 120V Cartridge, Heater Assembly 240V Male Connector, 1/4T x 3/8 NPTM Male Connector, 1/4T x 1/8 NPTM Adapter, Male 1/4T x 1/8 NPTM C
6200 Calorimeter Operating Instruction Manual APPENDIX F Parts List for Water Tank Assembly Item 1301DD 386VB 413VB 387VB 1020DD 271HWHJ Description Water Tank, 6200 Nipple, ½ NPT, Nylon Cap, ½ NPT, Nylon Elbow,1/2” NPT, Plastic Plug Cap Hose Barb, Male, 1/2T – 1/2NPTM Parts List for Cooling Water Supply Item 328VB 196VB 343VB A1276DD Description Union, Bulkhead ¼ Tube Valve, Brass, ¼ Tube Port Connector, ¼ Tube, Brass Cold Water Solenoid Assembly w / connectors Parts List for Oxygen Filling System Item
6200 Calorimeter Operating Instruction Manual APPENDIX F Figure F-1 6200 Oxygen Bomb Calorimeter Cutaway Front F-4 86
6200 Calorimeter Operating Instruction Manual APPENDIX F Figure F-2 6200 Oxygen Bomb Calorimeter Cutaway Rear F-587
6200 Calorimeter Operating Instruction Manual APPENDIX F Figure F-3 A1279DD Control Schematic F-6 88
6200 Calorimeter Operating Instruction Manual APPENDIX F Figure F-4 Oxygen Solenoid Assembly & Fittings F-789
6200 Calorimeter Operating Instruction Manual APPENDIX F Figure F-5 Water Tank Assembly F-8 90
6200 Calorimeter Operating Instruction Manual APPENDIX F Figure F-6 A1311DD Circulating Pump Assembly with Fittings F-991
6200 Calorimeter Operating Instruction Manual APPENDIX F Figure F-7 Temperature Control Assembly with Fittings F-10 92
6200 Calorimeter Operating Instruction Manual APPENDIX F Figure F-8 A1284DD2 Stirrer Hub Assembly F-11 93
6200 Calorimeter Operating Instruction Manual APPENDIX F Figure F-9 Stirrer Motor Assembly F-12 F-12 94
PARR INSTRUMENT COMPANY 211 Fifty - Third Street Moline, Illinois 61265 USA 309 / 762-7716 800 / 872-7720 Fax 309 / 762-9453 http://www.parrinst.com parr@parrinst.